2<br />

STOP-Warning<br />

The STOP-Warning is triggered when your pump switches from<br />

RUN mode to STOP mode, or when a battery is inserted. The<br />

pump emits two short beeps and vibrates once every minute.<br />

This warns you that insulin delivery has been interrupted.<br />

To turn off the STOP-Warning, press either a or s for three<br />

seconds. You then hear a melody and / or feel a vibration,<br />

confirming that the STOP-Warning has been turned off.<br />

If you press the key too briefly, the STOP-Warning will not be<br />

turned off.<br />

Note<br />

For your safety, the volume of the STOP-Warning cannot<br />

be changed. It is independent of the programmed beep<br />

volume.<br />

2.12 General safety information<br />

w Warning<br />

The cartridge and infusion set are sterile products intended<br />

for single use only. Sterility is guaranteed for unopened<br />

packaging up to the indicated expiry date. Do not use sterile<br />

products if the packaging is damaged.<br />

Do not re-use single-use materials because this might lead to<br />

a malfunction of your <strong>Accu</strong>-<strong>Chek</strong> <strong>Spirit</strong> <strong>Combo</strong> insulin pump,<br />

incorrect insulin delivery and / or infection. Always handle<br />

these items with clean hands. Avoid any contact of the<br />

infusion set and the connecting parts of your insulin pump<br />

with antiseptics, antibiotic creams, soaps, perfumes,<br />

deodorants, body lotions, or any other cosmetics. They may<br />

contaminate these parts.<br />