2<br />
STOP-Warning<br />
The STOP-Warning is triggered when your pump switches from<br />
RUN mode to STOP mode, or when a battery is inserted. The<br />
pump emits two short beeps and vibrates once every minute.<br />
This warns you that insulin delivery has been interrupted.<br />
To turn off the STOP-Warning, press either a or s for three<br />
seconds. You then hear a melody and / or feel a vibration,<br />
confirming that the STOP-Warning has been turned off.<br />
If you press the key too briefly, the STOP-Warning will not be<br />
turned off.<br />
Note<br />
For your safety, the volume of the STOP-Warning cannot<br />
be changed. It is independent of the programmed beep<br />
volume.<br />
2.12 General safety information<br />
w Warning<br />
The cartridge and infusion set are sterile products intended<br />
for single use only. Sterility is guaranteed for unopened<br />
packaging up to the indicated expiry date. Do not use sterile<br />
products if the packaging is damaged.<br />
Do not re-use single-use materials because this might lead to<br />
a malfunction of your <strong>Accu</strong>-<strong>Chek</strong> <strong>Spirit</strong> <strong>Combo</strong> insulin pump,<br />
incorrect insulin delivery and / or infection. Always handle<br />
these items with clean hands. Avoid any contact of the<br />
infusion set and the connecting parts of your insulin pump<br />
with antiseptics, antibiotic creams, soaps, perfumes,<br />
deodorants, body lotions, or any other cosmetics. They may<br />
contaminate these parts.<br />
57