External materials of construction<br />

Glass reinforced polyester<br />

Temperature<br />

Operating 15°C to 30°C<br />

Storage -15°C to 50°C<br />

Relative humidity<br />

Operating<br />

Storage<br />

Power requirement<br />

Operating position<br />

Dimensions (WxHxD)<br />

Open<br />

Closed<br />

Shipping carton<br />

Weight<br />

Maximum H 2<br />

O 2<br />

liquid volume<br />

Waste liquid bottle size<br />

H 2<br />

O 2<br />

liquid concentration<br />

Gas hoses<br />

Gas hose connections<br />

Control system<br />

Airflow rate<br />

Gassing<br />

Conditioning/aeration<br />

Temperature of supply gas at outlet 60°C<br />

H 2<br />

O 2<br />

liquid injection rate<br />

Double HEPA filtration from room inlet<br />

Noise level<br />

80% RH maximum<br />

80% RH at 38°C maximum<br />

AC 230V, single phase, 50Hz, 6 amps<br />

AC 120V, single phase, 60Hz, 11.5 amps<br />

AC 200V, three phase, 50/60Hz, 9.2 amps<br />

Upright only<br />

680 x 1260 x 900mm<br />

680 x 1060 x 900mm<br />

780 x 1350 x 1100mm<br />

140 kg, Gross shipping weight 170 kg<br />

500 ml or <strong>external</strong> 4 litre option<br />

500 ml, or <strong>external</strong> run-off, or <strong>external</strong> 4 litre option<br />

30-35% w/w<br />

38mm I/D x 2m long (optional extra)<br />

38mm (1.5 inch) male/female Camlock fittings<br />

Siemens ® S7 PLC<br />

(software compliant with GAMP)<br />

Greater than 30 m 3 /hr<br />

Greater than 30 m 3 /hr<br />

Approximate 1 – 8 g/min<br />

99.997% efficiency<br />

Less than 60dBA at 1 metre<br />

Enclosure <strong>decon</strong>tamination capacity For enclosures and rooms up to 350m 3<br />

(subject to configuration and loading)<br />

Data printer<br />

I/O communication port<br />

Locking brake<br />

Standard instrumentation<br />

• Wide range H 2<br />

O 2<br />

sensor (0-1000ppm)<br />

• Auxiliary test port for Dräger tube /<br />

hand pump sampler<br />

• Relative humidity sensor<br />

• Return and delivery gas temperature<br />

sensors (PT100)<br />

• Pre-heater temperature and safety cut-out<br />

• Re-heater temperature and safety cut-out<br />

• Vaporiser temperature and safety trip<br />

Thermal print record<br />

RS-485 D-type<br />

Integral foot pedal on/off<br />

• Cooling coil gas outlet temperature<br />

sensor (PT100)<br />

• Dosing system pressure / weight transducer<br />

• Air flow measurement transducer<br />

• Waste bottle full sensor<br />

• Bottle position sensors (internal bottles only)<br />

• Enclosure low/high pressure safety switches<br />

• Enclosure pressure transducer<br />

• Dual circuit valve position sensors<br />

Features in brief<br />

• Dual circuit technology for optimum cycle times<br />

• Closed circuit operation – no <strong>external</strong><br />

vent required<br />

• GAMP compliant software<br />

• Compact and highly portable design<br />

• Easily cleanable <strong>external</strong> surface<br />

• Volt-free contacts for <strong>external</strong> signals<br />

• Integral thermal paper printer<br />

• Remote start/stop<br />

• Simple connection, flexible supply/return hoses<br />

and pressure monitoring<br />

• Cycle edit and store facility for new applications<br />

• Effective against a wide range of micro-organisms<br />

• Utilises 30-35% w/w <strong>Bioquell</strong> hydrogen<br />

peroxide solution<br />

• Enclosures up to 350m 3<br />

084/10