External materials of construction<br />
Glass reinforced polyester<br />
Temperature<br />
Operating 15°C to 30°C<br />
Storage -15°C to 50°C<br />
Relative humidity<br />
Operating<br />
Storage<br />
Power requirement<br />
Operating position<br />
Dimensions (WxHxD)<br />
Open<br />
Closed<br />
Shipping carton<br />
Weight<br />
Maximum H 2<br />
O 2<br />
liquid volume<br />
Waste liquid bottle size<br />
H 2<br />
O 2<br />
liquid concentration<br />
Gas hoses<br />
Gas hose connections<br />
Control system<br />
Airflow rate<br />
Gassing<br />
Conditioning/aeration<br />
Temperature of supply gas at outlet 60°C<br />
H 2<br />
O 2<br />
liquid injection rate<br />
Double HEPA filtration from room inlet<br />
Noise level<br />
80% RH maximum<br />
80% RH at 38°C maximum<br />
AC 230V, single phase, 50Hz, 6 amps<br />
AC 120V, single phase, 60Hz, 11.5 amps<br />
AC 200V, three phase, 50/60Hz, 9.2 amps<br />
Upright only<br />
680 x 1260 x 900mm<br />
680 x 1060 x 900mm<br />
780 x 1350 x 1100mm<br />
140 kg, Gross shipping weight 170 kg<br />
500 ml or <strong>external</strong> 4 litre option<br />
500 ml, or <strong>external</strong> run-off, or <strong>external</strong> 4 litre option<br />
30-35% w/w<br />
38mm I/D x 2m long (optional extra)<br />
38mm (1.5 inch) male/female Camlock fittings<br />
Siemens ® S7 PLC<br />
(software compliant with GAMP)<br />
Greater than 30 m 3 /hr<br />
Greater than 30 m 3 /hr<br />
Approximate 1 – 8 g/min<br />
99.997% efficiency<br />
Less than 60dBA at 1 metre<br />
Enclosure <strong>decon</strong>tamination capacity For enclosures and rooms up to 350m 3<br />
(subject to configuration and loading)<br />
Data printer<br />
I/O communication port<br />
Locking brake<br />
Standard instrumentation<br />
• Wide range H 2<br />
O 2<br />
sensor (0-1000ppm)<br />
• Auxiliary test port for Dräger tube /<br />
hand pump sampler<br />
• Relative humidity sensor<br />
• Return and delivery gas temperature<br />
sensors (PT100)<br />
• Pre-heater temperature and safety cut-out<br />
• Re-heater temperature and safety cut-out<br />
• Vaporiser temperature and safety trip<br />
Thermal print record<br />
RS-485 D-type<br />
Integral foot pedal on/off<br />
• Cooling coil gas outlet temperature<br />
sensor (PT100)<br />
• Dosing system pressure / weight transducer<br />
• Air flow measurement transducer<br />
• Waste bottle full sensor<br />
• Bottle position sensors (internal bottles only)<br />
• Enclosure low/high pressure safety switches<br />
• Enclosure pressure transducer<br />
• Dual circuit valve position sensors<br />
Features in brief<br />
• Dual circuit technology for optimum cycle times<br />
• Closed circuit operation – no <strong>external</strong><br />
vent required<br />
• GAMP compliant software<br />
• Compact and highly portable design<br />
• Easily cleanable <strong>external</strong> surface<br />
• Volt-free contacts for <strong>external</strong> signals<br />
• Integral thermal paper printer<br />
• Remote start/stop<br />
• Simple connection, flexible supply/return hoses<br />
and pressure monitoring<br />
• Cycle edit and store facility for new applications<br />
• Effective against a wide range of micro-organisms<br />
• Utilises 30-35% w/w <strong>Bioquell</strong> hydrogen<br />
peroxide solution<br />
• Enclosures up to 350m 3<br />
084/10