<strong>Release</strong> <strong>Notes</strong><br />

PS Server Software Service Pack 5.90.7, Hugs Application 6.1.5<br />

and Pedz Application 1.5.5<br />

Overview<br />

The PS Server Software service pack 5.90.7 corrects unexpected restarts of the PS Server Software by the software<br />

Watchdog. It is a cumulative release, and incorporates the changes implemented in service packs 5.90.4 (LSASS memory<br />

leak) and 5.90.6 (XMLNS memory leak), as well as other minor improvements introduced in 5.90.2 and 5.90.5 (which<br />

were not general releases).<br />

In addition to supporting the improvements outlined above, the Hugs Application version 6.1.5 and Pedz Application<br />

version 1.5.5 improve the connection time of the client application software to the PS Server.<br />

PSS 5.90.7 - Key Technical Changes<br />

The following key technical changes have been implemented in the PSS 5.90.7:<br />

• Corrected unexpected Watchdog restarts of the service due to a corrupted request<br />

• Corrected XMLNS memory leak – carried forward from 5.90.6<br />

• Multiple changes to limit LSASS memory – carried forward from 5.90.4 and 5.90.5<br />

• Improved client to server connection time – carried forward from 5.90.5<br />

• Improved handling of LON supervision – carried forward from 5.90.5<br />

Hugs 6.1.5/Pedz 1.5.5 - Key Technical Changes<br />

The following key technical changes have been implemented in Hugs 6.1.5 and Pedz 1.5.5:<br />

• Improved client to server connection time<br />

• Bug fix – no longer able to manually admit a tag number already in the Census list without a warning<br />

• Bug fix – now able to filter tag supervision timeout alarms by PC<br />

• Updated to latest PC proximity card reader dll (pcProxAPI.dll – version 6.5.5.0)<br />

• Bug fix - GDI DotNetBar issue – related to number of User Defined Fields (UDFs) – that could prevent clients<br />

from connecting to the server<br />

• Improvements to contain GDI handle leak – prevent areas of the map not being drawn correctly on the screen<br />

• Improvements to correct proximity card reading errors when running two client applications on the same PC<br />

• Enhancement – limit number of UDFs to 24<br />

• Enhancement – add prompt when creating a new User Defined Field – prevents accidental adding of UDFs<br />

309 Legget Drive, Ottawa, ON K2K 3A3, Canada Toll free: 866-559-6275 Tel: +1 (613) 592-6997 Fax: (613) 592-4296 Web: www.stanleyhealthcare.com<br />

© 2010 XMARK Corporation. All Rights Reserved. Printed in Canada. July 2010. 970-806056-000 Rev 01.


Who should upgrade to PSS 5.90.7 and Hugs 6.1.5/Pedz 1.5.5<br />

The following guidelines should be used to determine if a site should be upgraded to PSS 5.90.7 and Hugs 6.1.5/<br />

Pedz 1.5.5:<br />

• All new installations<br />

• Any site running PSS 5.5 or higher that is experiencing unexpected restarts of the PS Server by the software<br />

Watchdog (see section below to determine if a site is having Watchdog restarts)<br />

• Any site running PSS 5.5 or higher that is experiencing an LSASS memory leak that is not already running 5.90.4<br />

• Any site running PSS 5.5 or higher that is experiencing an XMLNS memory leak that is not already running 5.90.6<br />

How do I determine if a site is having unexpected Watchdog restarts<br />

Depending on system size, these unexpected restarts may not be immediately apparent, since the system may only be<br />

down for a short period. Unless a user has either seen a restart, or attempted to use the system during a restart, the<br />

site may not be aware of the issue.<br />

To check if a site has experienced a series of restarts:<br />

1. Go to Control Panel/Administrative Tools/Event Viewer.<br />

2. Open the Patient Security event log (or Xmark if on 5.54).<br />

3. If the log contains two events a few seconds apart with a yellow triangle beside them, then the system has<br />

had Watchdog restarts as shown here:<br />

2<br />

4. If there are any Watchdog restarts then you should proceed with the upgrade.<br />

Upgrading from installed PSS Software to 5.90.7<br />

Important Note: 5.90.7 is a service pack and must be applied on top of 5.90.6. In turn, 5.90.6 is also a service pack<br />

and must be applied on top of 5.90.4.<br />

1. Ensure the system to be upgraded is running 5.90.6.<br />

2. Copy the Patient Server 5.90.7 folder from the ftp site to a USB drive or other media.<br />

3. Stop all the Client Applications.<br />

4. Shut down the system by stopping the psService in Services.<br />

5. Insert the USB drive and click on startup.exe in the Patient Server 5.90.7 folder.<br />

6. Reboot the Server PC when prompted.<br />

7. Verify that all the devices are online before starting the clients.<br />

PS Server Software Service Pack 5.90.7, Hugs Application 6.1.5 and Pedz Application 1.5.5


Upgrading from Hugs Software 6.1.x to 6.1.5<br />

Hugs 6.1.5 is a full install* and can be installed on top of any previous version of Hugs 6.x.<br />

*Note: In order to make it easier to download from the ftp site, the installation folders do not contain the .NET<br />

file, which is very large (207 megabytes). If you wish to install the Client application on a computer that does not<br />

already have .NET 3 or .NET 3 SP1 installed you will need to download the file separately from the ftp site.<br />

1. Ensure the client PC has .NET installed.<br />

2. Download the Hugs 6.1.5 folder from the ftp site.<br />

3. Copy the file to a USB drive or other media.<br />

4. Close the Hugs application if it is running.<br />

5. Run the Setup.exe and complete the upgrade.<br />

6. Launch Hugs.<br />

Upgrading from Pedz Software 1.5.x to 1.5.5<br />

Pedz 1.1.5 is a full install* and can be installed on top of any previous version of Pedz 1.x.<br />

*Note: In order to make it easier to download from the ftp site, the installation folders do not contain the .NET<br />

file, which is very large (207 megabytes). If you wish to install the Client application on a computer that does not<br />

already have .NET 3 or .NET 3 SP1 installed you will need to download the file separately from the ftp site.<br />

1. Ensure the client PC has .NET installed.<br />

2. Download the Pedz 1.5.5 folder from the ftp site.<br />

3. Copy the file to a USB drive or other media.<br />

4. Close the Pedz application if it is running.<br />

5. Run the Setup.exe and complete the upgrade.<br />

6. Launch Pedz.<br />

Known issues with PSS 5.90.7<br />

The following are known issues with PSS 5.90.7 at the time of release:<br />

1. We continue to investigate the root causes of the LSASS memory growth. This release resolves or makes<br />

manageable the issue in the vast majority of sites, but there are a few sites that continue to experience<br />

problems. We urge you to continue to monitor your sites, and inform Technical Support if any of your<br />

customers are experiencing growth in the memory consumed by LSASS after applying the 5.90.4 or 5.90.7<br />

upgrades. Please provide complete site information and configuration to Technical Support.<br />

2. Changing the Timeout (hh:mm) value in the Users or User Groups areas under Settings in the System<br />

Manager (and also under the Admin tab > Users in the Hugs and Pedz applications) sets the timeout to<br />

00:00 when it is saved. Setting it again will save the correct timeout value. A bug in the latest version of<br />

Microsoft .NET has been defined as the cause of this issue. We are searching for a workaround for the<br />

next release of PSS software.<br />

3. Echelon LNS version 3.26.12 is not compatible with PSS 5.90.7. A few sites may have been upgraded to<br />

LNS 3.26.12 when troubleshooting LNS issues in 5.90.4. While 5.90.4 was compatible with LNS 3.26.12<br />

PSS 5.90.7 is not. Please verify the LNS version in the Control Panel if you have any concerns before<br />

doing the upgrade.<br />

3<br />