D396-20-880 EXC100E-L turbo controller
D396-20-880 EXC100E-L turbo controller
D396-20-880 EXC100E-L turbo controller
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
1.6.2 Internal Timer<br />
The Internal Timer has two functions:<br />
Firstly, when the EXT pump is started by the Controller, the Internal Timer in the Controller also<br />
starts. If the EXT pump does not reach 50% of full rotational speed within the preset time<br />
measured by the timer, the Controller will trip into Fail condition. This function cannot be<br />
disabled.<br />
Secondly, you can configure the Controller to enable or disable the Internal Timer if the pump<br />
speed falls during pump operation:<br />
• If you disable the Internal Timer, the Controller will trip into Fail condition as soon as the<br />
pump speed falls to below 50% of full rotational speed.<br />
• If you enable the Internal Timer, the Internal Timer will start as soon as the pump speed falls<br />
to below 50% of full rotational speed; the Controller will trip into Fail condition if the pump<br />
speed is still below 50% of full rotational speed at the end of the preset time.<br />
The Controller is supplied with the Internal Timer enabled and adjusted for a preset time of<br />
eight minutes. You can adjust the timer for your application: refer to Section 3.11.<br />
EXC Turbomolecular Pump Controllers 7