1.6.2 Internal Timer<br />

The Internal Timer has two functions:<br />

Firstly, when the EXT pump is started by the Controller, the Internal Timer in the Controller also<br />

starts. If the EXT pump does not reach 50% of full rotational speed within the preset time<br />

measured by the timer, the Controller will trip into Fail condition. This function cannot be<br />

disabled.<br />

Secondly, you can configure the Controller to enable or disable the Internal Timer if the pump<br />

speed falls during pump operation:<br />

• If you disable the Internal Timer, the Controller will trip into Fail condition as soon as the<br />

pump speed falls to below 50% of full rotational speed.<br />

• If you enable the Internal Timer, the Internal Timer will start as soon as the pump speed falls<br />

to below 50% of full rotational speed; the Controller will trip into Fail condition if the pump<br />

speed is still below 50% of full rotational speed at the end of the preset time.<br />

The Controller is supplied with the Internal Timer enabled and adjusted for a preset time of<br />

eight minutes. You can adjust the timer for your application: refer to Section 3.11.<br />

EXC Turbomolecular Pump Controllers 7