<strong>Operating</strong> conditions<br />
Your application is of prime importance.<br />
Please note that the application and system performance is dependent upon the<br />
temperature range, viscosity, and flow rate of the thermal fluid:<br />
- When choosing the thermal fluid, not only minimal and maximum temperatures<br />
have to be complied with, but also suitability regarding burn point, viscosity and<br />
/ or freezing. Furthermore, the thermal fluid has to be compatible with all the<br />
materials used in the unit.<br />
- Pressure changes with the length of hoses (keep as short as possible). Choose<br />
as large a diameter of hoses as possible (the width of the pump connections are<br />
considered as a point of reference) and may negatively affect temperature<br />
control results. <strong>Flow</strong> restrictions may occur if a too narrow connector is selected<br />
for corrugated hoses.<br />
- The use of unsuitable hoses or hose connections may cause thermal and toxic<br />
injury to personal and environment. Temperature control hoses and their<br />
connections have to be insulated / secured against contact / mechanical damage.<br />
- Do not kink the hoses.<br />
- Check hoses in regular intervals for material fatigue (e.g. cracks).<br />
Please refer to the sections on operating requirements and general safety precautions.<br />
Location<br />
Caution!<br />
- Transport the unit upright.<br />
- The unit should be mounted in an upright and secure position, on a solid, stable<br />
surface.<br />
- Place on a non flammable surface.<br />
- Keep the area around the unit clean, to avoid slip and trip hazards.<br />
- Set the brakes on the castors once the unit is in position.<br />
- Place suitable absorbent material under the unit to catch any condensate and<br />
thermal fluid spills.<br />
- Any spillage of thermal fluid should be immediately cleaned up.<br />
- For large units, check the weight / load capacity for the flooring<br />
9