<strong>Operating</strong> conditions<br />

Your application is of prime importance.<br />

Please note that the application and system performance is dependent upon the<br />

temperature range, viscosity, and flow rate of the thermal fluid:<br />

- When choosing the thermal fluid, not only minimal and maximum temperatures<br />

have to be complied with, but also suitability regarding burn point, viscosity and<br />

/ or freezing. Furthermore, the thermal fluid has to be compatible with all the<br />

materials used in the unit.<br />

- Pressure changes with the length of hoses (keep as short as possible). Choose<br />

as large a diameter of hoses as possible (the width of the pump connections are<br />

considered as a point of reference) and may negatively affect temperature<br />

control results. <strong>Flow</strong> restrictions may occur if a too narrow connector is selected<br />

for corrugated hoses.<br />

- The use of unsuitable hoses or hose connections may cause thermal and toxic<br />

injury to personal and environment. Temperature control hoses and their<br />

connections have to be insulated / secured against contact / mechanical damage.<br />

- Do not kink the hoses.<br />

- Check hoses in regular intervals for material fatigue (e.g. cracks).<br />

Please refer to the sections on operating requirements and general safety precautions.<br />

Location<br />

Caution!<br />

- Transport the unit upright.<br />

- The unit should be mounted in an upright and secure position, on a solid, stable<br />

surface.<br />

- Place on a non flammable surface.<br />

- Keep the area around the unit clean, to avoid slip and trip hazards.<br />

- Set the brakes on the castors once the unit is in position.<br />

- Place suitable absorbent material under the unit to catch any condensate and<br />

thermal fluid spills.<br />

- Any spillage of thermal fluid should be immediately cleaned up.<br />

- For large units, check the weight / load capacity for the flooring<br />

9