PROBE LOCATION<br />

• When possible, locate <strong>probe</strong>s upstream from a hydraulic disturbance..<br />

• When possible, respect following minimum straight length :<br />

10 Ø upstream<br />

3 Ø downstream, Ø means diameter of the pipe .<br />

• In the case of an horiz<strong>on</strong>tal pipe, avoid installing the <strong>probe</strong>s<br />

<strong>on</strong> the top or the bottom part of the pipe because the<br />

risk of bubbles or deposits<br />

On a vertical pipe, there is no restricti<strong>on</strong> for <strong>probe</strong> locati<strong>on</strong>.<br />

• Requested accuracy <strong>on</strong> distance inter-<strong>probe</strong>s : +/- 0,5 mm<br />

• The c<strong>on</strong>verter calculates and displays this distance D.S to be adjusted between <strong>probe</strong>s<br />

around the central zero (see fig 1).<br />

PROBES INSTALLATION<br />

• Clean upthe pipe (rust, roughness). In most cases, the paint doesn’t matter .<br />

• Apply plenty of coupling agent <strong>on</strong> <strong>probe</strong>s face and <strong>on</strong> the pipe wall at the locati<strong>on</strong> .<br />

- Standard coupling gel for temperatures lower than 60°C (during a few hours).<br />

- High temperature special grease for temperatures up to 120°C or l<strong>on</strong>g term use .<br />

- Coupling elastomer, coated <strong>on</strong> both faces with special grease, for very l<strong>on</strong>g term use. It<br />

must be hardly tight.<br />

• The fixing is d<strong>on</strong>e by stretching the elastic around the pipe, or with stainless steel straps for<br />

fixed installati<strong>on</strong> (see fig 1).<br />

CONNECTIONS<br />

• The <strong>probe</strong> holder is equipped of two push-pull c<strong>on</strong>nectors for twinax shielded cables.<br />

• Use the cables re<strong>com</strong>mended by ULTRAFLUX.<br />

• C<strong>on</strong>nect upstream and downstream transducers to c<strong>on</strong>verter using corresp<strong>on</strong>ding inputs .<br />

COMMISSIONING<br />

• A good installati<strong>on</strong> in good site c<strong>on</strong>diti<strong>on</strong> must give a good echo signal and a reliable<br />

measurement result.<br />

• If the acoustic transmissi<strong>on</strong> result through the pipe is too poor (no echo or too high<br />

amplificati<strong>on</strong> gain), proceed as follows :<br />

- Check the distance D.S .between <strong>probe</strong>s .<br />

- Check the coupling agent between the <strong>probe</strong>s and the pipe .<br />

- If fault persists, try at another locati<strong>on</strong> (pipe internally corroded ?).<br />

- If no result, try with lower frequency transducers .<br />

- If liquid is flowing bubbles, try to improve these c<strong>on</strong>diti<strong>on</strong>s.<br />

FS210A GB1 04/10/04 page 2/2<br />

________________________________________________________________________________________________<br />

ULTRAFLUX Le Technoparc - 17, rue C. E. Jeanneret 78306 POISSY Cedex - France<br />

Tél. +33 (0)1 39 79 26 40 - Fax +33 (0)1 39 79 91 22 - e-mail : ultraflux@ultraflux.fr - http://www.ultraflux.<strong>com</strong>