General Half Pin Application<br />
(Cont.)<br />
Step Six<br />
Use the second pin clamp as a template for pin<br />
placement. Insert the two half pins through the pin<br />
clamp on the opposite side of the fracture.<br />
Step Seven<br />
Tighten the second pin clamp using the 6mm Allen<br />
Wrench. Make final adjustments using ball joints and<br />
<strong>Central</strong> Body translation. Usually leave 2cm between<br />
the skin and fixator for swelling.<br />
Step Eight<br />
Tighten all bolts using the 6mm Allen Wrench. Ball<br />
joints are sufficiently tightened when the gap in the<br />
housing is <strong>com</strong>pletely closed.<br />
Step Nine<br />
The JET-X CENTRAL Unilateral External Fixator is<br />
designed to facilitate obtaining anatomical reduction.<br />
Manual traction and closed manipulation of the<br />
fracture to achieve the best possible reduction prior<br />
to placing half pins is re<strong>com</strong>mended. Adjustments<br />
can be made in multiple planes:<br />
(a) 360° of reduction capability by rotating the<br />
ball joints.<br />
(b) Gross distraction by loosening the locking bolt in<br />
the telescoping <strong>Central</strong> Body.<br />
(c) Gradual <strong>com</strong>pression/distraction by using the<br />
snap-on Compression/Distraction Device.<br />
(See Page 9 for diagram.)<br />