24.06.2015 Views

FACE® N-Linked Oligosaccharide Profiling Kit - ProZyme

FACE® N-Linked Oligosaccharide Profiling Kit - ProZyme

FACE® N-Linked Oligosaccharide Profiling Kit - ProZyme

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

FACE ® N-<strong>Linked</strong> <strong>Oligosaccharide</strong> <strong>Profiling</strong> <strong>Kit</strong><br />

GK90000


TOOLS FOR GLYCOBIOLOGY<br />

www.glyko.com<br />

Page 2


FACE N-<strong>Linked</strong> <strong>Oligosaccharide</strong> <strong>Profiling</strong> <strong>Kit</strong> (80 Reactions)<br />

NOTE: We want successful results for our customers, so please read this entire booklet before starting<br />

your experiment.<br />

Contents<br />

Page No<br />

<strong>Kit</strong> Contents 2<br />

Additional Reagents and Equipment Required 3<br />

Introduction 4<br />

Principle<br />

Description of FACE N-linked <strong>Profiling</strong> System<br />

Nature of FACE Oligo Standards and Controls<br />

Summary of Protocol<br />

Protocols<br />

Section 1: Enzymatic Release of <strong>Oligosaccharide</strong>s from Glycoproteins 9<br />

Section 2: Labeling of <strong>Oligosaccharide</strong>s from Glycoproteins 11<br />

Section 3: Preparation of Samples and Standards for Electrophoresis 12<br />

Section 4: FACE Electrophoresis 14<br />

I. Preparation of OLIGO Gel Running Buffer<br />

II. Set-up of FACE Electrophoresis Apparatus<br />

III. Loading and Running FACE N-<strong>Linked</strong> <strong>Profiling</strong> Gels<br />

Section 5: Processing of FACE N-<strong>Linked</strong> <strong>Profiling</strong> Gels 18<br />

I. Gel Imaging using UV Transilluminator (lightbox)<br />

II. Gel Imaging using FACE Imaging System<br />

III. Gel Handling<br />

IV. <strong>Oligosaccharide</strong> Preparation<br />

Troubleshooting Guide 21<br />

Figures 23<br />

Page 1


<strong>Kit</strong> Contents<br />

Catalog # Code Component Quantity Storage<br />

GK60000<br />

10 each<br />

Oligo <strong>Profiling</strong> Gels precast gels Room Temp<br />

WS0007 Oligo Gel Running Buffer 5 packs Room Temp on arrival<br />

GK50050<br />

4 packs<br />

GK50004<br />

2 packs<br />

4°C (when reconstituted)<br />

E1 2x Loading Buffer 1 tube Room Temp<br />

E2 Tracking Dye 1 tube Room Temp<br />

C1 N-Glycanase® 1 tube<br />

(100 mU)<br />

4°C<br />

C3 2x <strong>Profiling</strong> Enzyme Buffer 1 tube 4°C<br />

C11 2% SDS/1M β-mercaptoethanol 1 tube 4°C<br />

C12 15% NP-40 1 tube 4°C<br />

L1 Reducing Agent (NaBH 3 CN) 2 tubes Room Temp on arrival<br />

-70°C (when reconstituted)<br />

L2 Oligo Labeling Dye (ANTS) 2 tubes Room Temp on arrival<br />

-70°C (when reconstituted)<br />

L3 Oligo Labeling Diluent 2 tubes Room Temp<br />

L4 Labeling Solvent (DMSO) 2 tubes Room Temp<br />

GK50006 E3 Oligo Ladder Standard 1 tube -20°C<br />

E4 Oligo <strong>Profiling</strong> Control 1 tube -20°C<br />

E5 Oligo Quantitation Control 1 tube -20°C on arrival<br />

-70°C (when reconstituted)<br />

Page 2


Additional Reagents and Equipment Required<br />

• FACE Electrophoresis Gel Box (<strong>ProZyme</strong> Product Code GK40026)<br />

• A CAP Type I water supply (such as Milli-Q)<br />

• Assorted pipeting devices including a 0-10 µl capillary positive displacement pipette<br />

(e.g. Drummond Microdispenser, Fisher Cat# 21-176D)<br />

• Centrifugal vacuum evaporator<br />

• 45°C and 37°C oven or water bath<br />

• Cold ethanol (100 %, nondenatured, -20°C)<br />

• 1.5 ml microcentrifuge tubes<br />

• Microfuge<br />

• UV-protective eye wear or face shield<br />

• UV transilluminator or FACE Imager<br />

Page 3


INTRODUCTION<br />

Complex carbohydrates are important components of all living things. In addition to providing energy<br />

and structural supports for cells, increasing evidence has shown that the carbohydrate moieties of<br />

glycoconjugates are often important as recognition determinants in receptor-ligand or cell-cell<br />

interactions, in the modulation of immunogenicity and protein folding, and in the regulation of protein<br />

bioactivity. Changes in the biological activity of glycoproteins often result from alterations in protein<br />

glycosylation either through variable site occupancy or changes in the structure of the oligosaccharide<br />

occupying a particular site. Using the FACE Imaging System, individual oligosaccharides can be<br />

quantified to obtain molar ratios, amount of glycosylation and detect changes in the extent or nature of<br />

glycosylation.<br />

Principle<br />

The FACE or "Fluorophore-Assisted-Carbohydrate-Electrophoresis" System is based on the use of<br />

polyacrylamide gel electrophoresis to separate and quantify intact oligosaccharides released from<br />

glycoproteins. FACE analysis begins with the release of Asparagine-linked (N-linked) oligosaccharides<br />

from the glycoprotein using the enzyme N-Glycanase (Peptide N-glycosidase F). The released<br />

oligosaccharides are then labeled with a fluorophore at the reducing termini by reductive amination. The<br />

stoichiometry of labeling is such that only one molecule of fluorophore is attached to each molecule of<br />

oligosaccharide. When labeling 20 nmoles or less of total sugar using the reagents and labeling<br />

conditions described below, the fluorophore labeling efficiency is greater than 98%. Labeling more than<br />

20 nmoles of sugar in each reaction will result in reduced labeling efficiency. When labeling more than<br />

20 nmoles we recommend that you include the OLIGO Quantitation Control (E5) as an internal labeling<br />

control. The fluorescent labeled oligosaccharides are then separated and quantified on an OLIGO<br />

<strong>Profiling</strong> Gel using the FACE Imaging System. Figure 1 (page 23) shows the profile of N-linked<br />

oligosaccharides released from a number of glycoproteins.<br />

Page 4


Description of the FACE N-linked OLIGO <strong>Profiling</strong> System<br />

The FACE N-linked OLIGO <strong>Profiling</strong> <strong>Kit</strong> contains all reagents and buffers required to separate and<br />

quantify the N-linked oligosaccharides released from glycoproteins and glycopeptides. This kit contains<br />

all the components required to perform 80 enzymatic digests and 80 separate oligosaccharide labeling<br />

reactions. In addition, the kit contains electrophoresis size markers for determining the relative positions<br />

of oligosaccharide bands on the gel, and a control glycoprotein to verify the performance of the enzyme<br />

and reagents in the kit. The kit also contains 10 precast OLIGO <strong>Profiling</strong> Gels and 5 OLIGO Gel<br />

Running Buffer packs sufficient to prepare 7.5 liters of Electrophoresis Running Buffer.<br />

The FACE N-linked Oligo <strong>Profiling</strong> System involves four steps:<br />

Step I) Release of the N-linked oligosaccharides from the glycoprotein, performed enzymatically.<br />

Step II) Labeling of the mixture of released oligosaccharides with a fluorescent tag.<br />

Step III) Separation of the fluorophore labeled oligosaccharides by polyacrylamide gel electrophoresis.<br />

The resulting banding pattern represents the oligosaccharide profile of the glycoprotein. This pattern can<br />

be thought of as a glycosylation "fingerprint" with the bands representing the oligosaccharides released<br />

from the glycoprotein.<br />

Step IV) Imaging of the gel. Imaging can be performed either on a UV lightbox to obtain qualitative<br />

band conformation, or the gels can be imaged using the FACE Imaging System to determine the amount<br />

of oligosaccharide present in each band and the relative mobility of the bands. Once separated on the<br />

gel, individual oligosaccharide bands can be purified for further study.<br />

Page 5


Nature of FACE OLIGO Standards and Controls<br />

The OLIGO Ladder Standard (E3)<br />

The OLIGO Ladder Standard (E3) consists of a mixture of glucose polymers ranging from Glucose 1 to<br />

greater than Glucose 16 . An illustration of the OLIGO Ladder Standard as it should appear at the<br />

completion of electrophoresis is shown in Figure 4, Lane 1 (page 25). When the standard is prepared as<br />

directed below and the recommended amount is loaded onto the gel, the band representing Glucose 4<br />

contains 50 pmoles (25 pmoles for FACE imager users). The intensity of the Glucose 4 band in the<br />

standard mixture is less than the adjacent bands. This allows for easy identification of this band even if<br />

the lower bands in the standard have run off the gel. Quantitation of the oligosaccharide bands in the<br />

samples is achieved by comparing the intensity of the Glucose 4 band with the intensity of sample bands<br />

and it is therefore essential that the OLIGO Ladder Standard be present on each FACE OLIGO <strong>Profiling</strong><br />

Gel. If the FACE Imaging System is used, dilute the E3 standard so that the Glucose 4 band contains<br />

25 pmoles. Refer to the FACE Software Manual for a detailed description of band finding, quantitation<br />

procedure, assigning a DP value, etc.<br />

OLIGO <strong>Profiling</strong> Control (E4)<br />

The OLIGO <strong>Profiling</strong> Control (E4) is trypsin inhibitor, used as a control for enzyme digestion and<br />

fluorophore labeling. The oligosaccharide pattern for trypsin inhibitor is shown in Figure 4, Lane 2<br />

(page 25). This control is included in the kit because:<br />

If you use this control when you use the kit for the first time it will help you become familiar<br />

with the kit procedures. If the profile you obtain looks like the pattern in Lane 2 of Figure 4 then<br />

you can be assured that the kit is working properly.<br />

If you have an unknown sample that may, or may not, contain N-linked oligosaccharides it is<br />

often useful to include a tube of this control along with your tubes containing samples.<br />

Following electrophoresis, if your samples do not show carbohydrate bands but you can see the<br />

normal pattern of oligosaccharides released from the control, this will assure you that the<br />

reagents are good and that the release and labeling procedures were performed properly.<br />

Page 6


OLIGO Quantitation Control (E5)<br />

The OLIGO Quantitation Control (E5) consists of 6 nmoles of unlabeled maltotetraose (Glucose 4 ).<br />

Once this control is reconstituted, 10 µl of E5 will contain 200 pmoles of maltotetraose.<br />

E5 may be used as an internal labeling control, provided that you verify that your sample does not<br />

contain an oligosaccharide that will co-migrate with maltotetraose (e.g. Man 3 GlcNAc 2 migrates very<br />

close to Glucose 4 ). As shown in Figure 3 (page 24), the labeling efficiency of your sample is greater<br />

than 95% when labeling less than 20 nmoles. With some unknown samples you may not know how<br />

much sugar you are labeling before you set up the digest. E5 is provided so that a "spike" of 200 pmoles<br />

(10 µl) may be added to the enzyme digest prior to the ethanol extraction step so that you can monitor<br />

labeling efficiency. The maltotetraose band from the "spike" will be clearly visible on the gel (it will comigrate<br />

with the Glucose 4 band in the OLIGO Ladder Standard (E3) as long as you load greater than<br />

1/40 th of the labeling reaction in the lane. After correcting for dilution of the labeling reactions you will<br />

be able to determine the labeling efficiency of your sample by comparing the quantity of the internal<br />

Glucose 4 band against the same band in a lane containing (E5) labeled in a separate tube, or the Glucose 4<br />

band in the lane containing the OLIGO Ladder Standard (E3). If you find that the quantity of your<br />

spiked internal Glucose 4 band (corrected for dilution) is only 190 pmoles (100% labeling efficiency<br />

would give you 200 pmoles), the labeling efficiency of the sample is about 95%.<br />

Page 7


Summary of <strong>Profiling</strong> Protocol<br />

Purified Glycoprotein<br />

N-Glycanase<br />

Digestion 2 hours to overnight<br />

Free <strong>Oligosaccharide</strong>s<br />

Label with<br />

fluorophore 3 hours to overnight<br />

Labeled <strong>Oligosaccharide</strong>s<br />

Load onto<br />

FACE Gel 30 minutes<br />

Electrophoresis 1 - 1 ½ hours<br />

FACE Imaging and Data Analysis<br />

Page 8


Protocols<br />

SECTION 1<br />

ENZYMATIC RELEASE OF OLIGOSACCHARIDES FROM GLYCOPROTEINS<br />

NOTE:<br />

FACE Electrophoresis Buffer should be prepared up to one day in advance of the<br />

electrophoresis run and stored at 4°C. See page 14 for instructions.<br />

1 Isolate the glycoprotein according to your usual procedures.<br />

NOTE: The purified glycoprotein should be prepared in a phosphate buffer containing a minimum<br />

amount of salt. Tris, acetate and citrate buffers will inhibit the labeling reaction.<br />

2 If the volume of the glycoprotein solution required is greater than 100 µl, dry the glycoprotein in a<br />

1.5 ml microcentrifuge tube. Generally 50-200 µg of glycoprotein is required for analysis.<br />

NOTE: The actual amount of glycoprotein required will depend on the size of the protein and the extent<br />

of glycosylation. In general, at least 1 nmole of glycoprotein should be digested for analytical work, a<br />

greater amount will be required for the preparation of purified oligosaccharides (see Page 19,<br />

"N−<strong>Linked</strong> <strong>Oligosaccharide</strong> Preparation from FACE <strong>Profiling</strong> Gels").<br />

3 Add an equal volume of 2x <strong>Profiling</strong> Enzyme Buffer (C3) to the glycoprotein in solution or dissolve<br />

the dried glycoprotein in 45 µl of 1x OLIGO <strong>Profiling</strong> Enzyme Buffer made from the 2x (C3) stock<br />

by adding an equal volume of water.<br />

4 Remove a 45 µl aliquot of the OLIGO <strong>Profiling</strong> Control (E4) and place in a 1.5-ml microfuge tube.<br />

Store the remaining glycoprotein at 4°C for future labeling controls.<br />

5 Optional: SDS is often required to completely denature the glycoprotein prior to enzymatic<br />

digestion. To denature add 2.5 µl of SDS/β-mercaptoethanol (C11), boil for 5 minutes and then add<br />

2.5 µl of NP-40 (C12) to both the samples and Oligo <strong>Profiling</strong> Control (E4) tube.<br />

NOTE: Some proteins will precipitate when boiled, e.g. immunoglobulins. This procedure should be<br />

used if your protein precipitates:<br />

a. Add SDS/β-ME at the recommended concentration. Incubate 5 minutes at room temperature.<br />

b. Add NP-40 according to directions.<br />

c. Add N-Glycanase and incubate overnight.<br />

Page 9


6 Add 2 µl of N-Glycanase (C1) to the glycoprotein sample and Oligo <strong>Profiling</strong> Control (E4). Mix<br />

with finger flicks and centrifuge for 5 seconds. Store any remaining enzyme at 4ºC.<br />

7 Incubate for 2 hours or overnight at 37°C.<br />

8 Prepare the OLIGO Quantitation Control (E5):<br />

a. Add 300 µl of distilled water to the tube labeled E5, which contains 6 nmoles of maltotetraose.<br />

b. Resuspend the lyophilized carbohydrate by vortexing.<br />

c. The concentration of the maltotetraose is now 200 pmoles/10 µl.<br />

9 Add 10 µl (200 pmoles) of reconstituted E5 to the glycoprotein digest as an internal labeling control,<br />

or label 10 µl in a separate tube to check your quantitation according to the guidelines on page 4.<br />

Store the unused reconstituted E5 at -20°C.<br />

10 Precipitate the protein by adding 3 volumes of cold ethanol. Keep samples on ice for 10 minutes.<br />

Spin samples in microcentrifuge for 5 minutes to pellet the protein.<br />

11 Remove the supernatant and transfer to a clean 1.5 ml microcentrifuge tube.<br />

DO NOT DISCARD THE SUPERNATANT; the released carbohydrates are present in the<br />

supernatant.<br />

12 If a large amount of protein was digested (>250 µg), 5-10% of the released oligosaccharides may<br />

remain in the pellet. The recovery of these oligosaccharides can be accomplished by drying the<br />

pellet completely in a centrifugal vacuum evaporator or lyophilizer. Add 50 µl water to resuspend,<br />

then 150 µl of cold ethanol and precipitate on ice. Centrifuge and combine the supernatants.<br />

13 Dry supernatants in a centrifugal vacuum evaporator or lyophilize to a translucent pellet.<br />

At this point samples may be stored at -20°C, or proceed with the fluorophore labeling procedure<br />

described in Section 2.<br />

Page 10


SECTION 2<br />

LABELING OF OLIGOSACCHARIDES FROM GLYCOPROTEINS<br />

NOTE: Sample Handling and Storage<br />

• Always avoid exposing labeled samples and dyes to light or excessive heat.<br />

• Labeled samples are stable for 3 months when stored at -70°C.<br />

• Unused solutions of the Labeling Dye and Labeling Reducing Agent may be stored for as long as<br />

2 weeks at -70°C. Thaw immediately before use.<br />

I. Preparation of fluorophore labeling reagents<br />

1 Add 125 µl of the Oligo Labeling Diluent (L3) to one vial of Oligo Labeling Dye (L2). Mix well by<br />

vortexing until dye is totally dissolved (warm at 37°C if necessary). Centrifuge for 1 second in a<br />

microcentrifuge. Reconstituted Labeling Dye solution may be stored for up to 2 weeks in the dark at<br />

-70°C.<br />

2 Add 125 µl of Labeling Solvent (L4) to one vial of Reducing Agent (L1). Mix well by vortexing<br />

until crystals are completely dissolved, warm briefly if necessary. Reconstituted Reducing Agent<br />

may be stored for 2 weeks at -70ºC.<br />

II. Labeling reactions<br />

Perform the following steps for the glycans released from your sample protein and the Oligo <strong>Profiling</strong><br />

Control (E4) (from Section 1 step 13). In addition, if you are using E5 as an external standard, take 10<br />

µl from the E5 tube, dry down in centrifugal vacuum evaporator or lyophilizer and follow these labeling<br />

instructions.<br />

1 Add 5 µl of reconstituted Oligo Labeling Dye (L2) to each dried oligosaccharide pellet. Mix well<br />

until the oligosaccharide pellet is dissolved.<br />

2 Add 5 µl of reconstituted Reducing Agent (L1). Mix well by vortexing. Centrifuge for 5 seconds in<br />

a microfuge.<br />

3 Incubate samples at 45ºC for 3 hours or overnight at 37°C.<br />

Page 11


SECTION 3<br />

PREPARATION OF SAMPLE AND STANDARDS FOR ELECTROPHORESIS<br />

I. Sample Preparation<br />

1a Add 10 µl water to your sample(s) and E4, then add 20 µl of 2x Loading Buffer. Load 4 µl or 1/10 th<br />

of the total sample volume directly onto the gel. Bring the E5 external control to a total volume of<br />

16 µl, with the addition of 6 µl of 2x Loading Buffer (added to 10 µl labeling reaction). Load 4 µl or<br />

50 pmoles onto the gel.<br />

Alternatively,<br />

1b Dry the samples, E4 and E5 in a centrifugal vacuum evaporator for approximately 15 minutes or<br />

until the sample reaches a viscous gel stage.<br />

2b Resuspend the fluorophore-labeled oligosaccharide from your sample only, in 10-20 µl water.<br />

NOTE: The actual volume of water used to resuspend the sample will depend on the amount of<br />

oligosaccharide present in the sample. Start with 10 µl, this will enable the sample to be diluted further<br />

if necessary.<br />

3b Remove 2 µl of the sample and add 2 µl of 2x Loading Buffer (E1).<br />

4b Resuspend E4 in 20 µl water and 20 µl of 2x Loading Buffer. Resuspend E5 in 8 µl water and 8 µl<br />

of 2x Loading Buffer.<br />

5b Load 4 µl of the sample, E4 and E5 directly onto the gel.<br />

II. Preparation of OLIGO Ladder Standard (E3):<br />

For non-imager users:<br />

1a Resuspend the OLIGO Ladder Standard (E3) in 50 µl of distilled water.<br />

2a Add 50 µl of 2x Loading Buffer (E1). After reconstitution the OLIGO Ladder Standard should be<br />

stored at -70°C.<br />

Page 12


3a Load 4 µl directly onto the gel; the maltotetraose band in the ladder should contain 50 pmoles.<br />

For Imager users:<br />

1b Resuspend the OLIGO Ladder Standard (E3) in 100 µl of distilled water.<br />

2b Add 100 µl of 2x Loading Buffer (E1). After reconstitution, the OLIGO Ladder Standard should be<br />

stored at -70°C.<br />

2c Load 4 µl directly onto the gel; the maltotetraose band in the ladder should contain 25 pmoles.<br />

Page 13


SECTION 4<br />

FACE ELECTROPHORESIS<br />

I. Preparation of 1.5 liter of FACE Oligo Gel Running Buffer<br />

(1.5 liter of buffer is sufficient for one electrophoresis run of one or two gels.)<br />

NOTE:<br />

FACE electrophoresis buffer should be prepared up to one day in advance of the<br />

electrophoresis run and stored at 4°C.<br />

1 Remove one of the Oligo <strong>Profiling</strong> Gel Running Buffer packs from the kit.<br />

2 Cut open one end of the packet and carefully pour the contents into a 2-liter graduated cylinder and<br />

add 1000 ml of distilled water.<br />

3 Rinse buffer packet with 100 ml of water and add to the redissolved buffer.<br />

4 Bring up to a final volume of 1.5 liter with water.<br />

5 Mix well. The buffer must be chilled to 4-6°C before use.<br />

II. Set-up of FACE Electrophoresis Apparatus<br />

FACE Electrophoresis of N-linked gels is best performed at a temperature range of 5-8°C. We<br />

recommend the use of a recirculating chiller, however satisfactory results can be obtained by chilling the<br />

buffer prior to the electrophoresis run.<br />

CAUTION:<br />

To prevent damage to the gel box, connect both quick disconnect fittings to gel box<br />

before turning on the recirculator!<br />

1 Place the FACE electrophoresis tank containing a stir bar on a mechanical stirrer. If using a<br />

recirculating chiller, connect it to the gel box cooling chamber with tubing and quick-connect<br />

fittings. Turn on the circulator and stirrer; set the coolant temperature to 5ºC. Bleed all air out of<br />

system by tilting the gel box in the direction of the outlet.<br />

Page 14


2 Pour pre-cooled OLIGO Gel Running Buffer into the electrophoresis tank up to the level marked on<br />

the outside of the box. The temperature of the buffer can be monitored during the run using a<br />

thermometer inserted through the hole in the lid. The temperature will probably increase a few<br />

degrees during electrophoresis but should not exceed 10°C<br />

III. Loading and Running FACE OLIGO <strong>Profiling</strong> Gels<br />

1 Determine the number of gels required for the samples prepared. Each Oligo <strong>Profiling</strong> Gel contains<br />

8 lanes. One outside lane should be used for the Oligo Ladder Standard (E3) leaving the other 7<br />

lanes for samples. Optional: Load 2 µl of the E2 tracking dye to an outside lane to monitor<br />

electrophoresis. This leaves 6 lanes for samples. The FACE electrophoresis box will accept one or<br />

two gels.<br />

2 Cut open 1 or 2 packages containing FACE Oligo <strong>Profiling</strong> Gels. Gently remove the comb(s) from<br />

the gel(s). To avoid distorting the wells, gently wiggle each comb to free the teeth from the gel, then<br />

lift up slowly until the comb is released.<br />

3 It is essential that the wells of the gel are thoroughly rinsed with Oligo Gel Running Buffer from the<br />

upper buffer reservoir prior to sample loading. This is best accomplished by using a syringe with a<br />

blunt needle (a Pasteur pipette is not recommended because of the possibility of breakage in the<br />

wells).<br />

4 Place the gel cassette(s), one on each side of the center core unit of the gel box with the short glass<br />

plate against the gasket. Be sure the cassette is centered and that the cassette is resting on the "feet"<br />

at the bottom of the apparatus. If running only one gel, place the buffer dam on the other side.<br />

5 Place one wedge down each side of the cassette. Then push wedges down to obtain a seal between<br />

the inner short plate and the gasket. Repeat this procedure on the other side. Both sides should now<br />

be sealed against the gaskets.<br />

6 Fill the upper buffer reservoir formed between the cassettes with approximately 100 ml of chilled<br />

Oligo Gel Running Buffer and check for leaks. Final buffer level should be just below the electrode<br />

supports. If a leak occurs, first check that the apparatus was assembled properly, then try pushing<br />

down the wedge with slightly greater pressure. If the leak persists check the "Troubleshooting"<br />

section of this manual.<br />

Page 15


7 With the core unit containing the gels placed securely on the bench, load samples into the wells by<br />

underlaying the upper buffer. Use flat sequencing pipette tips (e.g. Sigma T-1656) to load by<br />

delivering the sample to the bottom of each well. Optimal resolution will be achieved by using 4 µl<br />

of sample per lane.<br />

NOTE: For the most reliable quantitation of oligosaccharide bands the use of a positive displacement<br />

pipette is recommended.<br />

8 Place the core unit containing the loaded gels into the electrophoresis tank and place the lid on<br />

electrophoresis tank.<br />

9 First connect the power cords to the electrophoresis tank then connect the power supply. Connect<br />

the positive (red) lead to the electrode marked with a red dot on the electrophoresis box. Connect the<br />

negative (black) lead to the electrode marked with a black dot.<br />

10 A thermometer may be placed into the lower buffer chamber through the hole in the lid to monitor<br />

the temperature.<br />

NOTE: The initial temperature of the lower buffer should be between 5°C and 8°C.<br />

11 Turn on the power supply and select the proper current. FACE OLIGO <strong>Profiling</strong> Gels should be run<br />

at a constant current of 20 mA per gel (40 mA for 2 gels). Limits on the power supply should be set<br />

for 1000 volts and 60 watts. These run conditions will result in voltages of 100-400 V at the<br />

beginning of the run and may approach 800 V at the end of the run. If the initial voltage is<br />

significantly different, check to be sure that the leads are connected properly and that the buffers are<br />

at the recommended levels.<br />

12 Monitor electrophoresis by following the migration of the fast moving Orange Dye (E1). Generally,<br />

electrophoresis is complete when the orange dye just exits the bottom of the gel in approximately<br />

1 hour and 5 minutes. If the tracking dye (E2) is used, 4 colors should separate during the course of<br />

the electrophoresis run.<br />

13 Most N-linked oligosaccharides fall in the Glucose 4 – Glucose 12 range in the OLIGO Ladder<br />

Standard (E3) (see Figure 4, page 25) so the time of electrophoresis should be adjusted to optimize<br />

the separation of this region of the gel.<br />

Page 16


14 When the electrophoresis is complete, turn off the power supply. Disconnect the power cords from<br />

the power supply and the electrophoresis tank. Turn off the recirculating chiller if used.<br />

NOTE: If using the FACE Imager, begin warm-up now.<br />

Page 17


SECTION 5<br />

PROCESSING OF FACE OLIGO PROFILING GELS<br />

Following Step 14, above, carefully remove the gels from the electrophoresis tank.<br />

I. Gel Imaging using the UV Transilluminator (lightbox)<br />

CAUTION:<br />

UV-protective eyewear or face shield should be worn. Avoid prolonged exposure to UV<br />

light.<br />

1 Allow UV lightbox to "warm-up" for at least 2 minutes in order to get maximum intensity output.<br />

The lightbox must be long-wave UV (UVP Model TL-33 or equivalent) and have a peak output at<br />

approximately 360 nm; this is not the type of box typically used for ethidium-stained DNA gels.<br />

2 Optional: To increase sensitivity you may wish to peel the tape from the gel cassette, and<br />

disassemble the cassette by carefully prying open the two glass plates, removing the gel completely<br />

from the cassette and placing it directly on the UV lightbox.<br />

Images of gels can be recorded using a Polaroid camera. The proper choice of light source, filters and<br />

film must be made. A filter must be fitted to the camera lens, which completely covers the glass of the<br />

lens, stray UV contacting the lens will cause it to fluoresce and subsequently lower the sensitivity of the<br />

film. A Kodak Wratten Number 8 filter is suggested and can be purchased from Kodak and Kodak<br />

distributors. A suitable filter will have no inherent fluorescence, peak transmission at approximately<br />

500 nm, and bandwidth of 80 nm FWHM.<br />

A medium speed, medium resolution, Polaroid film is recommended. Use Polaroid 53 film for cameras<br />

that use single 4"x 5" sheet film; use Polaroid 553 film for cameras that use 8 sheet film cartridges.<br />

Photographing Gels:<br />

1 Photograph the gel using the lowest practical F setting on the lens, with the gel filling as much of<br />

the frame as possible. In our experience, exposures at F5.6 using Polaroid 53 film have ranged from<br />

5 to 40 seconds using the equipment specified above. KEEP UV EXPOSURE OF THE GEL TO A<br />

MINIMUM TO PREVENT BLEACHING.<br />

Page 18


2 Develop film according to manufacturer's instructions.<br />

II.<br />

Gel Imaging Using the FACE Imaging System<br />

1 Turn on the FACE Imager and allow a warm-up period of at least 10 minutes before acquiring an<br />

image. Allow the computer to initialize the Imager.<br />

2 Clean the glass plates completely with lint-free tissue. A little distilled water should be used to<br />

clean any obvious residue on the plates before imaging. The gel cassettes should be relatively dry<br />

and free of dust before imaging.<br />

3 Open the door to the Imager and place the cassette containing the gel into the cassette holder (see<br />

the "FACE Software Manual" for imaging instructions).<br />

III. Gel Handling<br />

After imaging, gels may be processed in a number of ways depending on the needs of the investigator.<br />

• If the gel is no longer needed it should be properly discarded.<br />

• As long as the gel cassette is intact, it can be placed back in the electrophoresis apparatus and the<br />

run continued in order to improve the resolution of the oligosaccharide bands.<br />

• Following imaging of the oligosaccharide gels, the glass plates can be separated and the gels<br />

dried on a flat bed gel drier between sheets of Teflon membrane at 80°C for 1 hour. After the gel<br />

is dry, carefully peel the Teflon sheets away from the gel. Gels dried in this manner may be<br />

stored indefinitely and re-imaged at any time.<br />

NOTE: Drying fixes the bands so that they do not continue to diffuse through the gel.<br />

IV. N-<strong>Linked</strong> <strong>Oligosaccharide</strong> Preparation from FACE <strong>Profiling</strong> Gels<br />

FACE <strong>Profiling</strong> Gels may be used for the preparation of oligosaccharides (e.g. for subsequent enzymatic<br />

sequencing) by loading aliquots of the same sample in all lanes. The following procedure is<br />

recommended for maximum recovery of oligosaccharides. Due to loss of signal that may occur during<br />

this procedure, accurate quantitation may not be possible on purified oligosaccharides.<br />

Page 19


1 Following electrophoresis, bands on the gel can be visualized by placing the gel on a long wave UV<br />

lightbox.<br />

CAUTION:<br />

UV protective eyewear or face shield should be worn. Avoid prolonged exposure to UV<br />

light.<br />

2 Remove the tape and carefully separate the cassette with a broad spatula. Using a scalpel or razor<br />

blade, carefully excise the band(s) of interest. This should be done as quickly as possible to<br />

minimize photobleaching.<br />

3 Place the gel slices in microcentrifuge tubes.<br />

4 Soak the gel slices in a minimum volume of 100% ethanol for ½ hour.<br />

5 Pipet off the ethanol and suspend the gel slices in a minimum volume of water, enough to cover the<br />

gel slices.<br />

6 Vortex the gel slices gently, place at 4°C overnight or all day (12-16 hours).<br />

7 Remove the supernatant and place in a microcentrifuge tube.<br />

8 If maximum recovery is required wash the gel slices, again, in a minimum volume of water. Place<br />

the gel slices at 4°C for 1 to 4 additional hours.<br />

9 Remove the supernatant and pool with the first supernatant.<br />

10 If any particulate material is visible, filter the supernatant pool through a 0.65µ filter.<br />

11 Dry the samples (supernatants).<br />

Page 20


TROUBLESHOOTING GUIDE<br />

Sample not moving or moving slowly on gel.<br />

Leads may be reversed. Check leads to power supply and gel box. Check upper buffer level is above the<br />

top of the short glass plate. Check the lower buffer level. If the level of the buffer is above the orange<br />

gasket, remove 20-50 mls of buffer until the level is below the gasket.<br />

Erratic voltage, voltage and/or current leak. (At the beginning of the run voltage is greater than<br />

400V or readings are unstable.)<br />

Make sure that the electrical leads are not arcing due to condensation on the lid of the tank. Dry the area<br />

around the electrical posts on the inner box. Check that the leads to the power supply are connected<br />

securely. Verify that the power supply is operating properly.<br />

Band distortion in gel<br />

From time to time band distortions may occur when running O-linked gels, for various reasons,<br />

including:<br />

• Sample may be overloaded, use a maximum of 1/5 th of the volume of the labeling reaction for<br />

each lane.<br />

• Wells may have been torn when comb was removed. Remove comb slowly using a gentle back<br />

and forth rocking motion and lift vertically.<br />

• “Smile effect” on the gel may be the result of a gel not being cooled uniformly. Check that the<br />

cooling system is on and working properly, and is free of air bubbles in the electrophoresis box.<br />

Make sure that a stirring bar has been placed in the electrophoresis box and that the lower buffer<br />

is being mixed. Check that power supply is set for proper current level - see Section 4-III-11 for<br />

proper settings.<br />

• “Fuzzy bands” may be due to too high current used. Check Section 4-III-11 for proper settings.<br />

Page 21


Upper buffer chamber leaks when cassettes are in place.<br />

Check that the plates are clean, not cracked or chipped, and that they are centered on the inner core<br />

assembly. Once the wedge has been brought into position against the glass, gently push down to engage<br />

the sealing gasket. Check that the gaskets are not cracked and that they are seated properly in place (do<br />

not grease gaskets). Pushing down with slightly greater pressure can stop most minor leaks.<br />

If you are using a FACE Imaging System and would like us to take a look at your gel for troubleshooting<br />

purposes, please e-mail a copy of the .raw or .IM2 file to us at glyko@prozyme.com<br />

Trademarks<br />

FACE is a registered trademark of <strong>ProZyme</strong>, Inc., San Leandro, CA, USA<br />

The following designations are trademarks owned by other companies:<br />

Eppendorf is a trademark of Eppendorf Geratebau + Hinz, GmbH<br />

Kodak is a trademark of the Eastman Kodak Company<br />

Milli-Q is a trademark of the Millipore Corporation<br />

N-Glycanase is a registered trademark of Genzyme Corporation, Boston, MA, USA. Recombinant<br />

PNGase F is covered by US Patent No 5,238,821 and its foreign equivalents. The product is<br />

manufactured under exclusive license by <strong>ProZyme</strong>, Inc<br />

Nonidet is a trademark of Shell International Petroleum Company<br />

Polaroid is a trademark of the Polaroid Corporation<br />

Teflon is a trademark of E.I. DuPont de Nemours and Company<br />

Tris is a trademark of Rohm and Haas<br />

Page 22


Separation of N-<strong>Linked</strong> <strong>Oligosaccharide</strong>s Released from Various Glycoproteins<br />

Figure 1.<br />

Lane 1 FACE Oligo Ladder Standard<br />

Lane 2 Fetuin N-linked oligosaccharides<br />

Lane 3 HIV gp120 N-linked oligosaccharides<br />

Lane 4 Human α 1 -acid glycoprotein<br />

Lane 5 Bovine ribonuclease B<br />

Lane 6 Human chorionic gonadotrophin<br />

Lane 7 Human α 1 -trypsin<br />

Lane 8 Bovine deoxyribonuclease<br />

Page 23


Fluorophore Labeling<br />

Figure 2.<br />

Fluorophore labeling by reductive amination<br />

Figure 3. Fluorophore labeling efficiency of a di-sialylated biantennary oligosaccharide.<br />

Page 24


1 2<br />

<br />

Figure 4.<br />

Lane 1 E3 – FACE Oligo Ladder Standard<br />

Lane 2 E4 – Oligo <strong>Profiling</strong> Control<br />

NOTE: The band marked in Lane 1 is Glucose 4 , maltotetraose, which is provided individually as an<br />

internal standard (E5).<br />

Page 25


<strong>ProZyme</strong>, Inc.<br />

3832 Bay Center Place, Hayward, CA 94545-3619 USA<br />

Tel: 1(510) 638-6900 Toll-free: 1(800) 457-9444<br />

Fax: 1(510) 638-6919<br />

E.mail: glyko@prozyme.com<br />

www.prozyme.com/glyko<br />

AD

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!