Temperature control7.5 Cooling (R/RH) and heating (RH) notesCentrifuges with cooling (R) or cooling and heating (RH) are equipped with an efficientcooling system. It allows obtaining selected temperatures in the chamber even at maximum spinspeed or fast obtaining desired temperatures (e.g. 4:C and 36:C). Note that time and possibility ofobtaining a set temperature is dependent on multiple factors, including: the power of the coolingsystem, the shape of the rotor, the rotor speed, ambient temperature, etc. The accuracy of thetemperature stability of ± 1 : C is determined by the installation place of the temperature sensor.31