Parameters of centrifugation8.2 Radius• RADIUS [mm] - control of the radius ofthe rotor within the range from R min toR max . Avalaible values depends on chosenrotor, see ––––– / ––––– (LIST OFROTORS fold).• When radius is changed is activated,symbol is visible on the screen.• Reducing of the rotor radius (and theresulting change of displayed RCF value)applies until switching off the powersupply of the centrifuge or setting theRmax maximum radius once again(loading the program does not changethis setting!).• To change the rotor radius selectRADIUS [mm] with ▲▼◄► keys.[MM]. Press SET. Set demanded value bypressing ▲▼. Press SET to acceptchanges.8.3 Sample density• DENSI (g/cm 3 )) – default density is set to1,2 g/cm 3 (possible values 1,2 ÷ 9,9g/cm 3 ).• When density is changed, symbol isvisible on the screen.• Increasing density of the sample above1,2 g/cm 3 (and limiting of the maximumspeed of centrifuging resulting from it)applies until switching off power supplyof the centrifuge or setting the deviceback to 1,2 g/cm 3 .• To change the density select DENSI(g/cm 3 )) with ▲▼◄► keys. [MM].Press SET. Set demanded value bypressing ▲▼. Press SET to acceptchanges.33