347.2 What if ...?Error/Error messageíååååìCauseOverloadRectification➞ Remove sample from weighing pan, zero again (tare).î____ïUnderload➞ Check whether weighing pan is positioned properly.Error 1Error 2No stability• in taring or adjusting (calibration)• when reference weight for piece counting is placed on panWrong adjustment weight on pan or none at all➞ Wait for stability before pressing key.➞ Ensure more stable ambient conditions.➞ Remove weighing pan and clean if necessary➞ Place required adjustment weight in center <strong>of</strong> pan.Error 3Reference number for piece counting too small➞ Increase number <strong>of</strong> reference pieces.Error 4Error 60.000 gInternal faultNo standard calibrationWrong weighing pan or pan missing➞ Contact <strong>METTLER</strong> <strong>TOLEDO</strong> customer service.➞ Contact <strong>METTLER</strong> <strong>TOLEDO</strong> customer service.➞ Mount correct weighing pan.AbortAdjustment aborted with the «C» key