26 <strong>Setrox</strong> S Steroid-Eluting <strong>Leads</strong> Technical ManualSetscrew AdjustmentCautionThe pulse generator’s setscrew(s) must be retracted priorto inserting the lead connector. Failure to back off the pulsegenerator’s setscrew(s) may result in damage to the lead(s),and/or difficulty connecting the lead(s).4. Insert the lead connector into the pulse generator’s headerreceptacle following the manufacturer’s directions for leadinsertion. The method of insertion depends upon the headerconfiguration of the pulse generator.5. Securely tighten the setscrew(s) of the pulse generator connector.CautionTightening SetscrewDo not over-tighten the setscrew(s). Use only a torquewrench, which automatically prevents over-tightening.6. Assure correct insertion of the connector pin by verifying that thetip of the pin is visible past the header connector block.7. Visually and mechanically check the integrity of the connectionafter tightening the setscrew(s).