<strong>pulse</strong> <strong>oximeters</strong>NEW MouseOx Pulse Oximeter for Rodents!Respiration<strong>pulse</strong> <strong>oximeters</strong>MRIThroat ClipCollar ClipSpecificationsOxygen Saturation (SpO2):RangeResolutionResponse Time0 to 100% Arterial Blood Oxygen Saturation≤ 1.5% across entire measurement rangeReported after every heartbeat – screen refreshedevery 0.72 secondsPulse Rate (BPM):Range90 to 900 BPMResolution Pulse Rate (BPM) Resolution (BPM)210 2.4300 4.9400 8.7500 13.5600 19.4700 26.2800 34.0900 42.9300 8.7Response Time Reported after every heartbeat – screen refreshedevery 0.72 secondsBreath Rate (BrPM):Range25 to 450 BrPMResolution Breath Rate (BrPM) Resolution (BrPM)25 0.1100 0.6200 2.2300 4.9400 8.7Response Time Reported every 1.7 seconds and is a moving averageof the rate of breathing for the previous 10 breath ratemeasurements. The screen updates every 0.72 seconds,but new breath rate data is presented at the lower rate.Pulse Distention:Range0 to 800 mmResolution ≤ 2.4% of measurementResponse Time Reported after every heartbeat – screen refreshedevery 0.72 secondsBreath Distention:Range0 to 800 mmResolution ≤ 2.4% of measurementResponse Time Reported every 1.7 secondsF60Harvard Apparatus • phone 508.893.8999 • toll free U.S. 800.272.2775 • fax 508.429.5732 • www.harvardapparatus.com