12.07.2015 Views

PEG 400 by UHPLC with corona ultra

PEG 400 by UHPLC with corona ultra

PEG 400 by UHPLC with corona ultra

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

The Corona ® <strong>ultra</strong>Charged Aerosol DetectorResponse (peak area)3000002500002000001500001000005000001.11 min 1.34 min 1.56 min 1.76 min 1.94 min 2.09 min 2.24 min 2.37 min 2.49 min 2.60 minR 2 = 0.9945R 2 = 0.9925R 2 = 0.9944R 2 = 0.9949R 2 = 0.9965R 2 = 0.9965R 2 = 0.9992R 2 = 0.9991R 2 = 0.9998R 2 = 0.9990 500 1000 1500 2000 2500Total Mass on Column (ng)Figure 4. Response Curves for 18 ng to 2 μg Total Mass (oncolumn).ConclusionSensitive methods for the analysis of <strong>PEG</strong><strong>400</strong> can be obtained <strong>by</strong> either HPLC or<strong>UHPLC</strong> <strong>with</strong> Charged Aerosol Detection.The transfer of the method from HPLC to<strong>UHPLC</strong> offered a four-fold decrease inanalysis time and solvent consumption.The ability to detect low nanogramquantities of <strong>PEG</strong> <strong>400</strong> <strong>with</strong> limits ofdetection

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!