5.0 Maintenance & TrouboubleshootingModel <strong>306WA</strong>Symptom CauseWhat To DoCalibrator current is erratic.Proper calibration current exists,but there is no cell response.a) Poor electrical connections.a) Short circuit in calibrator.a) Check for properconnections betweenthe control unit andcalibrator on terminalstrip TS5.b) Ensure that brown (13)and yellow (14) wiresconnect to brown andyellow posts respectivelyon calibratorassembly in the cellcompartment.c) Moisten wick. Seesection 4.3.a) Verify cell response to<strong>oxygen</strong> by loosening thecalibrator assembly tointroduce air, and thenbolt back in place.Recorder/meter shouldshow increase.b) If there is no response,inspect the calibrator forobvious shorts. If noteasily repairable,replace the calibrator.5-14TELEDYNE ELECTRONIC TECHNOLOGIES<strong>Analytical</strong> Instruments