Corona ® Charged Aerosol DetectorApplication noteTime %A %B Flow Rate0.00 100 0 1.55.00 100 0 1.511.00 96 4 1.524.00 37 63 1.535.00 32 68 1.545.00 18 82 1.958.00 15 85 1.960.00 5 95 1.563.00 100 0 1.565.00 100 0 1.5Table 1. Gradient Conditions.Results <strong>and</strong> DiscussionReproducibility <strong>of</strong> the method was tested by compar<strong>in</strong>gfour different <strong>in</strong>jections <strong>of</strong> the palm oil sample at 74µgcolumn load. The peak areas for an <strong>in</strong>dividual peak aswell as groups <strong>of</strong> peaks were observed <strong>and</strong> the resultsare shown <strong>in</strong> Table 2. The peak retention times exhibitedless than 1% variation over the gradient run for allsamples observed. Four overlays <strong>of</strong> the palm oil areshown <strong>in</strong> Figure 1. The chromatograms for the fourst<strong>and</strong>ards run are illustrated <strong>in</strong> Figures 2A <strong>and</strong> 2B. Thecorrelation is done us<strong>in</strong>g a polynomial fit <strong>and</strong> was greaterthen >0.999 for all po<strong>in</strong>ts (Figure 3). The LOD <strong>of</strong> the was