Automatic overload protection for CO sensorThe CO sensor is safeguarded from overload by an internal programme.Above a pre-programmed limit, a flushing pump is switchedon (P2) to supply fresh air to the CO sensor (red arrows).The measurement of the other gas components (black arrows)NO x /SO 2filtercan be resumed without problems also during the few secondsuntil automatic re-connection. To extend its life span it is possibleto switch off the CO sensor manually with simultaneous sensorpurging.Purging pumpP2Measuring gas pumpSensorsCOP1 O 2 NO NO 2 SO 2Service-friendly designThe <strong>MK</strong> <strong>9000</strong> is designed insuch a way that all wearing partscan be quickly exchanged. Sensorsand battery can be substitutedwithout instrument disassembly.Important componentsare installed within the robustmetal housing which can bedismounted quickly for repair ormaintenance works.Easy opening of thebattery compartmentEasily exchangeablesensorsMaintenance-free thermal quick printerThe integral thermal quick printer provides with an immediate documentationof the analysed values issued with a printing speedof maximal 75 mm/s. The printout can be personalized with theoperator address (8 x 24 characters). Optionally the <strong>MK</strong> <strong>9000</strong>can be equipped with a matrix printer in order to secure a longlastingprintouts availability.Maintenance-freeMatrix quick printer5