<strong>Antegrade</strong> approach –Reaming (optional)Reaming medullary canal (optional)Optional instruments189.060 SynReam Intramedullary Reaming System352.032 Reaming Rod 2.5 mm with ball tip,length 950 mmIf necessary enlarge the femoral canal with the medullaryreamer up to the desired diameter.Check fracture reduction under the image intensifier.Inserting the reaming rodInsert the reaming rod into the medullary canal.ReamingStarting with the diameter 8.5 mm, ream the medullarycanal in 0.5 mm increments. The holding forceps isused to control the rotation of the reaming rod. Advancethe reamer head with slight forward and backwardmovements.Do not use force. Continue reaming until the diameter ofthe canal is 0.5 to 1.5 mm larger than the nail diameter.Note: All <strong>Expert</strong> <strong>Retrograde</strong> /<strong>Antegrade</strong> <strong>Femoral</strong> <strong>Nail</strong>s can beinserted over the Reaming Rod. The tip of the reaming rodmust be correctly positioned in the medullary canal since itdetermines the final distal position of the nail.Synthes 61