3-13. Installation3.1 Transmitter MountingTransmitter may be mounted in any position.After transmitter is mounted, tighten all bolts.Pipe may be clamped to another pipe, or flangedand bolted to floor or wall.U-bolt secures assembly to 2" pipe. U-bolt may berevolved 90° for use with horizontal pipe.For fixed regulator and associated parts, refer toCustomer Maintenance Parts List involved.Optional air-set can be mounted as illustratedbelow.3.2 Air Supply and TransmissionPipingTransmitter1/4" (8A) TubingOutput gaugeSupply gauge(optional)Shut off valveAir supplyAir filter-regulator (optional)ReceiverF0303.aiFigure 3.3Air Supply and Transmission PipingU-bolt2" PipeEither set ofmounting holesmay be used.BoltNOTE• Air supply must be regulated at 1.4 kgf/cm 2or bar, 140 kPa, or 20 psi.• Transmitter uses 0.5 Nm 3 /h of air in normaloperation.• Air must be clean and dry. Blow out filterregularly.• Transmission line must be free of leaks.F0301.aiFigure 3.1Transmitter MountingLowpressureHighpressureBypassF0302.aiFigure 3.2Typical Transmitter PipingIM 02C01B06-01E