4.3 Weighing System MeasurementMeasure -5 - 35.00mVdcMeasure (0-20.000Vdc)Measure (0-24.000mA)mA--++MEASUREmVSOURCE++++MEASUREmASOURCECALOG+Ex+-mV-mV+Ex-Example: Weighing System MeasurementUsed for measuring loadcell output millivolts from existing operatingsystems. Connect mV measure from CALOG - LC II to loadcell output anddisplay mV. For mass display also measure excitation voltage (on senseterminals) or set a value for excitation with mV/V and units in the mA&masssubmenu. You can then display the loadcell mV as well as the gross masson the loadcell/s (not excluding tare weight). The mA measurement isoptional should you wish to see the mA output from the loadcell transmitter(LCT) whilst measuring mV or mass.Note : mA measure is isolated<strong>Zemic</strong> Europe info@zemic.nl www.zemic.nl8