Operating the CS-3 in the FieldFigure 5 - Hemisphere Switch Logic Diagram"R" - Remote 80HZ (JP3 - OFF, JP2 - ON)In the position "R" of the Hemisphere switch, the setting of the operatinghemisphere is actuated remotely by superimposing an 80 Hz sine signal onthe D.C. supply voltage. This can be conveniently achieved by coupling thecontrol signal voltage through a transformer at a convenient remote location.In the absence of the signal, the CS-3 is set to operate in the northernoperating hemisphere.To set the CS-3 for the operation in the southern operating hemisphereminimum AC signal amplitude of 50mV Vpeak is required at the frequencyof (80+/-5)Hz at the power input line to the CS-3. Please note, that thecontrol signal frequency has been selected far away from the Larmor signalfrequency range to make the interference with the magnetic fieldmeasurement negligible. The input amplifier of the magnetic processorsfilters out efficiently the frequencies outside the Larmor frequency range.2-8CS-3 <strong>Manual</strong> - part # 762701Revision 1.0