Defibrillators
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Defibrillators</strong><br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Defibrillation<br />
Electrical cardioversion and defibrillation have become routine procedures in the management<br />
of patients with cardiac arrhythmias. Cardioversion is the delivery of energy that is<br />
synchronised to the QRS complex, while defibrillation is the non-synchronised delivery of a<br />
shock randomly during the cardiac cycle.<br />
Most defibrillators are energy-based, meaning that the device charges a capacitor to a selected<br />
voltage and then rapidly delivers a pre-specified amount of energy in joules (J) to the<br />
myocardium to treat cardiac arrhythmias. The capacitance of a capacitor is the amount of<br />
electric charge it can store for every volt applied to it. With regard to defibrillators the amount<br />
of energy stored in a capacitor is very important. It can be calculated using the formula<br />
E = ½CV 2 , where E is the energy in joules, C the capacitance in farads and V the voltage<br />
measured in volts. This energy is dissipated in the patient’s body over a small time interval,<br />
about 10 milliseconds or one hundredth of a second.<br />
If the capacitance is 1000 μF and the voltage is 500 V then the stored energy is 125J.<br />
[E = ½ CV 2 ]<br />
E= ½ (1000 × 10 -6 ) x (500 2 ) = 125 J.<br />
Current European Society of Cardiology and AHA guidelines suggest the following initial energy<br />
selection for specific arrhythmias:<br />
<br />
<br />
<br />
<br />
For atrial fibrillation, 120 to 200 joules for biphasic devices and 200 joules for<br />
monophasic devices.<br />
For atrial flutter, 50 to 100 joules for biphasic devices and 100 joules for monophasic<br />
devices.<br />
For ventricular tachycardia with a pulse, 100 joules for biphasic devices and 200 joules<br />
for monophasic devices.<br />
For ventricular fibrillation or pulseless ventricular tachycardia, at least 150 joules for<br />
biphasic devices and 360 joules for monophasic devices.<br />
They also incorporate an inductor to prolong the duration of the delivered current, and a<br />
rectifier to convert alternating current (AC) to direct current (DC). (Knight, 2014)<br />
A defibrillator can deliver a controlled electrical shock to a heart that has a life-threatening<br />
rhythm, such as ventricular fibrillation (VF). In VF, the heart's chaotic activity prevents blood<br />
from pumping adequately or at all. Voltage stored by the defibrillator conducts electrical<br />
current (a shock) through the chest by way of electrodes or paddles placed on the chest. This<br />
brief pulse of current halts the chaotic activity of heart, by depolarising a large part of the heart<br />
muscle terminating the dysrhythmia allowing normal sinus rhythm to be re-established by the<br />
body’s internal pace maker located in the sinoatrial node of the heart, giving the heart a chance<br />
to re-start with a normal rhythm.<br />
Many factors affect the chance of defibrillation success including; placement of the electrode<br />
pads, time elapsed before the first shock is given, and certain health conditions. Successful<br />
defibrillation requires that enough current be delivered to the heart muscle during the shock. If<br />
the transthoracic impedance level is high the heart may not receive enough current for<br />
defibrillation to be successful. Impedance is the body's resistance to the flow of current; some<br />
people naturally have higher impedance than others. Therefore, it may take more current, a<br />
longer shock duration, and/or increased voltage to ensure success. (EBME, 2003)
The shock is delivered via two electrode pads/paddles placed as shown below.<br />
Fig 1 – Placement of Electrode Pads/Paddles<br />
Modern defibrillators may be manual or automated; they generally produce biphasic waveforms<br />
as opposed to monophasic waveforms, which increase safety and efficacy. Miniature<br />
implantable cardioverter-defibrillators (ICD) may be used in patients with recurrent lifethreatening<br />
arrhythmias. (Chaudhari, 2005)<br />
Monophasic Waveforms<br />
This is a type of defibrillation waveform where current flows in one direction. In this waveform,<br />
there is no ability to adjust for patient impedance, and it is generally recommended that all<br />
monophasic defibrillators deliver 200 - 300 J of energy to a maximum of 360J, applied to adult<br />
patients with the assumed average impedance of 50 ohms, to ensure maximum current is<br />
delivered which in the graph below is ≈ 45 amps.<br />
Biphasic Waveforms<br />
Fig 2 – Graphical representation of a Monophasic Waveform<br />
With biphasic shocks, the direction of current flow is reversed near the halfway point of the<br />
electrical defibrillation cycle. Biphasic waveforms were initially developed for use in<br />
implantable defibrillators and have since become the standard in external defibrillators. With<br />
biphasic waveforms there is a lower defibrillation threshold (DFT) that allows reductions of the<br />
energy levels administrated and may cause less myocardial damage.<br />
While all biphasic waveforms have been shown to allow termination of VF at lower current than<br />
monophasic defibrillators, there are two types of waveforms used in external defibrillators.
The waveforms are shown below and will have the desired effect at current values ranging from<br />
approx. 15 – 35 amps.<br />
Fig 3 – Graphical representation of two Biphasic Waveforms<br />
Types of Defibrillator<br />
Automated External Defibrillator (AED)<br />
AEDs are highly sophisticated, microprocessor-based devices that analyse multiple features of<br />
the surface ECG signal including frequency, amplitude, slope and wave morphology. They<br />
contain various filters for QRS signals, radio transmission and other interferences, as well as for<br />
poor electrode contact. Some devices are programmed to detect patient movement.<br />
The typical controls on an AED include a power button, a display screen on which trained<br />
rescuers can check the heart rhythm and a discharge button. Certain defibrillators have special<br />
controls for internal paddles or disposable electrodes.<br />
In AED Mode, the Defibrillator analyses the patient’s ECG and advises you whether or not to<br />
deliver a shock. Voice prompts guide you through the defibrillation process by providing<br />
instructions and patient information. Voice prompts are reinforced by messages/pictures that<br />
appear on the display. (Lozano, 2013)<br />
Manual Defibrillator<br />
Manual defibrillators are designed to give full control to the clinical users. The defibrillator<br />
records the patients ECG, the user then assess the ECG and selects the appropriate level of<br />
energy for defibrillation.<br />
Capnography<br />
End tidal Carbon Dioxide (EtCO 2 ) is the partial pressure or maximal concentration of carbon<br />
dioxide (CO 2) at the end of an exhaled breath, which is expressed as a percentage of CO 2 or<br />
mmHg. The normal values are 5% to 6% CO 2, which is equivalent to 35-45 mmHg. CO 2 reflects<br />
cardiac output and pulmonary blood flow as the gas is transported by the venous system to the<br />
right side of the heart and then pumped to the lungs by the right ventricles. When CO 2 diffuses<br />
out of the lungs into the exhaled air, a device called capnometer measures the partial pressure<br />
or maximal concentration of CO 2 at the end of exhalation.
Capnography uses an EtCO 2 sensor to continuously monitor the carbon dioxide that is inspired<br />
and exhaled by the patient. It is usually presented as a graph of expiratory CO 2 against time, or<br />
less commonly against expired volume. The sensor employs infrared (IR) spectroscopy to<br />
measure the concentration of CO 2 molecules that absorb infrared light. This consists of a source<br />
of infrared radiation, a chamber containing the gas sample, and a photo-detector. When the<br />
expired CO 2 passes between the beam of infrared light and photo-detector, the absorbance is<br />
proportional to the concentration of CO 2 in the gas sample. The gas samples can be analysed by<br />
the mainstream (in-line) or side-stream (diverting) techniques. (Physio-Control, 2013)<br />
During CPR, the amount of CO 2 excreted by the lungs is proportional to the amount of<br />
pulmonary blood flow; therefore capnography can be used to monitor the effectiveness of CPR<br />
and as an early indication of the Return of Spontaneous Circulation (ROSC).<br />
It has been shown that when a patient experiences ROSC the first indication is often a sudden<br />
rise in EtCO 2 as the rush of circulation washes un-transported CO 2 from tissues, likewise a<br />
sudden drop in EtCO 2 may indicate that the patient has lost pulse and CPR may need to be<br />
restarted. (Paramedicine, 2000)<br />
Maintenance & Service Procedures<br />
<strong>Defibrillators</strong> are serviced annually, during the service functional checks of all controls, displays<br />
and sound outputs are performed. ECG functions are checked including heart rate calibration<br />
and lead off detection, most defibrillators can detect whether the paddles/pads are connected<br />
or disconnected and this should also be checked.<br />
An analyser is used to ensure output energy levels are within specification and a check of all<br />
functions/analysis is performed when in AED mode.<br />
Pacer function and pacer detection are both tested, a functional check of the capnography (if<br />
applicable) and finally an electrical safety test is performed.<br />
There is also scheduled battery and patient lead replacement, the expiry dates on the pads<br />
should be checked to ensure they are still ok to use. If they are past there expiry date the ward<br />
staff should be informed and the pads removed from use and replaced.<br />
During maintenance & service procedures it is vital to ensure a defibrillator is never left alone<br />
charged. When repairing or opening the case for any reason it is important to follow the<br />
manufacturer’s guidelines for discharging the capacitor to ensure no harm comes to yourself or<br />
others.
Bibliography<br />
Chaudhari, M., 2005. Anaesthesia Journal. [Online]<br />
Available at: http://www.anaesthesiajournal.co.uk/article/S1472-0299(06)00175-5/abstract<br />
[Accessed September 2015].<br />
EBME, 2003. EBME - Biphasic Defibrillator. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/12-biphasicdefibrillation?showall=&start=3<br />
[Accessed September 2015].<br />
Knight, B. P., 2014. UpToDate. [Online]<br />
Available at: http://www.uptodate.com/contents/basic-principles-and-technique-ofcardioversion-and-defibrillation<br />
[Accessed September 2015].<br />
Lozano, I. F., 2013. Principles of External defibrillators. [Online]<br />
Available at: http://www.heartrhythmcharity.org.uk/www/media/files/InTech-<br />
Principles_of_external_defibrillators.pdf<br />
[Accessed September 2015].<br />
Paramedicine, 2000. End Tidal CO2. [Online]<br />
Available at: http://www.paramedicine.com/pmc/End_Tidal_CO2.html<br />
[Accessed October 2015].<br />
Phillps Medical Systems, 2005. M4735A (ELD) Heartstream XL Defibrillator Service/User Manual,<br />
Physio-Control, 2013. Lifepak® 20e Defibrillator Service/User manual.