12.05.2016 Views

Defibrillators

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Defibrillators</strong><br />

Steven Lewis<br />

Clinical Engineering<br />

United Lincolnshire Hospital Trust


Defibrillation<br />

Electrical cardioversion and defibrillation have become routine procedures in the management<br />

of patients with cardiac arrhythmias. Cardioversion is the delivery of energy that is<br />

synchronised to the QRS complex, while defibrillation is the non-synchronised delivery of a<br />

shock randomly during the cardiac cycle.<br />

Most defibrillators are energy-based, meaning that the device charges a capacitor to a selected<br />

voltage and then rapidly delivers a pre-specified amount of energy in joules (J) to the<br />

myocardium to treat cardiac arrhythmias. The capacitance of a capacitor is the amount of<br />

electric charge it can store for every volt applied to it. With regard to defibrillators the amount<br />

of energy stored in a capacitor is very important. It can be calculated using the formula<br />

E = ½CV 2 , where E is the energy in joules, C the capacitance in farads and V the voltage<br />

measured in volts. This energy is dissipated in the patient’s body over a small time interval,<br />

about 10 milliseconds or one hundredth of a second.<br />

If the capacitance is 1000 μF and the voltage is 500 V then the stored energy is 125J.<br />

[E = ½ CV 2 ]<br />

E= ½ (1000 × 10 -6 ) x (500 2 ) = 125 J.<br />

Current European Society of Cardiology and AHA guidelines suggest the following initial energy<br />

selection for specific arrhythmias:<br />

<br />

<br />

<br />

<br />

For atrial fibrillation, 120 to 200 joules for biphasic devices and 200 joules for<br />

monophasic devices.<br />

For atrial flutter, 50 to 100 joules for biphasic devices and 100 joules for monophasic<br />

devices.<br />

For ventricular tachycardia with a pulse, 100 joules for biphasic devices and 200 joules<br />

for monophasic devices.<br />

For ventricular fibrillation or pulseless ventricular tachycardia, at least 150 joules for<br />

biphasic devices and 360 joules for monophasic devices.<br />

They also incorporate an inductor to prolong the duration of the delivered current, and a<br />

rectifier to convert alternating current (AC) to direct current (DC). (Knight, 2014)<br />

A defibrillator can deliver a controlled electrical shock to a heart that has a life-threatening<br />

rhythm, such as ventricular fibrillation (VF). In VF, the heart's chaotic activity prevents blood<br />

from pumping adequately or at all. Voltage stored by the defibrillator conducts electrical<br />

current (a shock) through the chest by way of electrodes or paddles placed on the chest. This<br />

brief pulse of current halts the chaotic activity of heart, by depolarising a large part of the heart<br />

muscle terminating the dysrhythmia allowing normal sinus rhythm to be re-established by the<br />

body’s internal pace maker located in the sinoatrial node of the heart, giving the heart a chance<br />

to re-start with a normal rhythm.<br />

Many factors affect the chance of defibrillation success including; placement of the electrode<br />

pads, time elapsed before the first shock is given, and certain health conditions. Successful<br />

defibrillation requires that enough current be delivered to the heart muscle during the shock. If<br />

the transthoracic impedance level is high the heart may not receive enough current for<br />

defibrillation to be successful. Impedance is the body's resistance to the flow of current; some<br />

people naturally have higher impedance than others. Therefore, it may take more current, a<br />

longer shock duration, and/or increased voltage to ensure success. (EBME, 2003)


The shock is delivered via two electrode pads/paddles placed as shown below.<br />

Fig 1 – Placement of Electrode Pads/Paddles<br />

Modern defibrillators may be manual or automated; they generally produce biphasic waveforms<br />

as opposed to monophasic waveforms, which increase safety and efficacy. Miniature<br />

implantable cardioverter-defibrillators (ICD) may be used in patients with recurrent lifethreatening<br />

arrhythmias. (Chaudhari, 2005)<br />

Monophasic Waveforms<br />

This is a type of defibrillation waveform where current flows in one direction. In this waveform,<br />

there is no ability to adjust for patient impedance, and it is generally recommended that all<br />

monophasic defibrillators deliver 200 - 300 J of energy to a maximum of 360J, applied to adult<br />

patients with the assumed average impedance of 50 ohms, to ensure maximum current is<br />

delivered which in the graph below is ≈ 45 amps.<br />

Biphasic Waveforms<br />

Fig 2 – Graphical representation of a Monophasic Waveform<br />

With biphasic shocks, the direction of current flow is reversed near the halfway point of the<br />

electrical defibrillation cycle. Biphasic waveforms were initially developed for use in<br />

implantable defibrillators and have since become the standard in external defibrillators. With<br />

biphasic waveforms there is a lower defibrillation threshold (DFT) that allows reductions of the<br />

energy levels administrated and may cause less myocardial damage.<br />

While all biphasic waveforms have been shown to allow termination of VF at lower current than<br />

monophasic defibrillators, there are two types of waveforms used in external defibrillators.


The waveforms are shown below and will have the desired effect at current values ranging from<br />

approx. 15 – 35 amps.<br />

Fig 3 – Graphical representation of two Biphasic Waveforms<br />

Types of Defibrillator<br />

Automated External Defibrillator (AED)<br />

AEDs are highly sophisticated, microprocessor-based devices that analyse multiple features of<br />

the surface ECG signal including frequency, amplitude, slope and wave morphology. They<br />

contain various filters for QRS signals, radio transmission and other interferences, as well as for<br />

poor electrode contact. Some devices are programmed to detect patient movement.<br />

The typical controls on an AED include a power button, a display screen on which trained<br />

rescuers can check the heart rhythm and a discharge button. Certain defibrillators have special<br />

controls for internal paddles or disposable electrodes.<br />

In AED Mode, the Defibrillator analyses the patient’s ECG and advises you whether or not to<br />

deliver a shock. Voice prompts guide you through the defibrillation process by providing<br />

instructions and patient information. Voice prompts are reinforced by messages/pictures that<br />

appear on the display. (Lozano, 2013)<br />

Manual Defibrillator<br />

Manual defibrillators are designed to give full control to the clinical users. The defibrillator<br />

records the patients ECG, the user then assess the ECG and selects the appropriate level of<br />

energy for defibrillation.<br />

Capnography<br />

End tidal Carbon Dioxide (EtCO 2 ) is the partial pressure or maximal concentration of carbon<br />

dioxide (CO 2) at the end of an exhaled breath, which is expressed as a percentage of CO 2 or<br />

mmHg. The normal values are 5% to 6% CO 2, which is equivalent to 35-45 mmHg. CO 2 reflects<br />

cardiac output and pulmonary blood flow as the gas is transported by the venous system to the<br />

right side of the heart and then pumped to the lungs by the right ventricles. When CO 2 diffuses<br />

out of the lungs into the exhaled air, a device called capnometer measures the partial pressure<br />

or maximal concentration of CO 2 at the end of exhalation.


Capnography uses an EtCO 2 sensor to continuously monitor the carbon dioxide that is inspired<br />

and exhaled by the patient. It is usually presented as a graph of expiratory CO 2 against time, or<br />

less commonly against expired volume. The sensor employs infrared (IR) spectroscopy to<br />

measure the concentration of CO 2 molecules that absorb infrared light. This consists of a source<br />

of infrared radiation, a chamber containing the gas sample, and a photo-detector. When the<br />

expired CO 2 passes between the beam of infrared light and photo-detector, the absorbance is<br />

proportional to the concentration of CO 2 in the gas sample. The gas samples can be analysed by<br />

the mainstream (in-line) or side-stream (diverting) techniques. (Physio-Control, 2013)<br />

During CPR, the amount of CO 2 excreted by the lungs is proportional to the amount of<br />

pulmonary blood flow; therefore capnography can be used to monitor the effectiveness of CPR<br />

and as an early indication of the Return of Spontaneous Circulation (ROSC).<br />

It has been shown that when a patient experiences ROSC the first indication is often a sudden<br />

rise in EtCO 2 as the rush of circulation washes un-transported CO 2 from tissues, likewise a<br />

sudden drop in EtCO 2 may indicate that the patient has lost pulse and CPR may need to be<br />

restarted. (Paramedicine, 2000)<br />

Maintenance & Service Procedures<br />

<strong>Defibrillators</strong> are serviced annually, during the service functional checks of all controls, displays<br />

and sound outputs are performed. ECG functions are checked including heart rate calibration<br />

and lead off detection, most defibrillators can detect whether the paddles/pads are connected<br />

or disconnected and this should also be checked.<br />

An analyser is used to ensure output energy levels are within specification and a check of all<br />

functions/analysis is performed when in AED mode.<br />

Pacer function and pacer detection are both tested, a functional check of the capnography (if<br />

applicable) and finally an electrical safety test is performed.<br />

There is also scheduled battery and patient lead replacement, the expiry dates on the pads<br />

should be checked to ensure they are still ok to use. If they are past there expiry date the ward<br />

staff should be informed and the pads removed from use and replaced.<br />

During maintenance & service procedures it is vital to ensure a defibrillator is never left alone<br />

charged. When repairing or opening the case for any reason it is important to follow the<br />

manufacturer’s guidelines for discharging the capacitor to ensure no harm comes to yourself or<br />

others.


Bibliography<br />

Chaudhari, M., 2005. Anaesthesia Journal. [Online]<br />

Available at: http://www.anaesthesiajournal.co.uk/article/S1472-0299(06)00175-5/abstract<br />

[Accessed September 2015].<br />

EBME, 2003. EBME - Biphasic Defibrillator. [Online]<br />

Available at: http://www.ebme.co.uk/articles/clinical-engineering/12-biphasicdefibrillation?showall=&start=3<br />

[Accessed September 2015].<br />

Knight, B. P., 2014. UpToDate. [Online]<br />

Available at: http://www.uptodate.com/contents/basic-principles-and-technique-ofcardioversion-and-defibrillation<br />

[Accessed September 2015].<br />

Lozano, I. F., 2013. Principles of External defibrillators. [Online]<br />

Available at: http://www.heartrhythmcharity.org.uk/www/media/files/InTech-<br />

Principles_of_external_defibrillators.pdf<br />

[Accessed September 2015].<br />

Paramedicine, 2000. End Tidal CO2. [Online]<br />

Available at: http://www.paramedicine.com/pmc/End_Tidal_CO2.html<br />

[Accessed October 2015].<br />

Phillps Medical Systems, 2005. M4735A (ELD) Heartstream XL Defibrillator Service/User Manual,<br />

Physio-Control, 2013. Lifepak® 20e Defibrillator Service/User manual.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!