Binder1
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
12 Lead ECG Recorders<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Electrocardiography (ECG)<br />
The heart is responsible for pumping blood around the body by repeated, rhythmic<br />
contractions. The functional unit of the heart is the cardiac muscle cell or cardiomyocyte. Each<br />
cardiomyocyte maintains an electrical charge or potential across its cell membrane and<br />
contracts when this potential is discharged. In order for all of the cardiomyocytes to contract at<br />
the same time and produce an effective muscular contraction, the heart also maintains its own<br />
electrical conducting system which coordinates the electrical activity of the heart.<br />
Fig 1 – The Human Heart<br />
The electrical impulses take a specialised conduction pathway.<br />
This pathway is made up of 5 elements:<br />
1. The Sino-Atrial (SA) node<br />
2. The Atrio-Ventricular (AV) node<br />
3. The bundle of His<br />
4. The left and right bundle branches<br />
5. The Purkinje fibres<br />
Fig 2 – The Human Heart
The SA node is the natural pacemaker of the heart. Pacemakers and Temporary Pacing Wires<br />
(TPWs) are used when the SA node has ceased to function properly.<br />
The SA node releases electrical stimuli at a regular rate; this rate is dictated by the needs of the<br />
body. Each stimulus passes through the myocardial cells of the atria creating a wave of<br />
contraction which spreads rapidly through both atria.<br />
The heart is made up of around half a billion cells; the majority of the cells make up the<br />
ventricular walls. The rapidity of atrial contraction is such that around 100 million myocardial<br />
cells contract in less than one third of a second. The electrical stimulus from the SA node<br />
eventually reaches the AV node and is delayed briefly so that the contracting atria have enough<br />
time to pump all the blood into the ventricles. Once the atria are empty of blood the valves<br />
between the atria and ventricles close. At this point the atria begin to refill and the electrical<br />
stimulus passes through the AV node and Bundle of His into the Bundle branches and Purkinje<br />
fibres.<br />
The Purkinje fibres spread widely across the ventricles. In this way all the cells in the ventricles<br />
receive an electrical stimulus causing them to contract.<br />
As the ventricles contract, the right ventricle pumps blood to the lungs where carbon dioxide is<br />
released and oxygen is absorbed, whilst the left ventricle pumps blood into the aorta from<br />
where it passes into the coronary and arterial circulation.<br />
At this point the ventricles are empty, the atria are full and the valves between them are closed.<br />
The SA node is about to release another electrical stimulus and the process is about to repeat<br />
itself. The SA node recharges whilst the atria are refilling, and the AV node recharges when the<br />
ventricles are refilling. In this way there is no need for a pause in heart function. Again, this<br />
process takes less than one third of a second.<br />
The times given for the 3 different stages are based on a heart rate of 60 bpm, or 1 beat per<br />
second.<br />
The term used for the release (discharge) of an electrical stimulus is "depolarisation", and the<br />
term for recharging is "repolarisation".<br />
The sum total of the simultaneous electrical discharging and re-charging of all the<br />
cardiomyocytes in the heart is sufficient to be detected by sensing probes placed on the exterior<br />
of the body at various positions around the heart. The ECG records the vector sum and produces<br />
a combined trace. This is the principle behind the electrocardiograph or ECG which can be used<br />
to monitor the rhythm of the heart.<br />
Injured cardiomyocytes such as those suffering from lack of oxygen during a heart attack leak<br />
electrical current rather than discharge it in a coordinated manner, the altered electrical signal<br />
of the injured heart results in a characteristic ECG pattern which can lead to the diagnosis of<br />
acute myocardial infarction (Heart Attack).<br />
In contrast, dead cardiomyocytes or scarred cardiac muscle does not carry or maintain an<br />
electrical charge, and this absence of electrical activity is also detectable by ECG. Therefore, a<br />
previously unrecognised or "silent" heart attack can be diagnosed by electrocardiogram, and<br />
even localised to a particular area of the heart by using multiple sensing probes or ECG leads.<br />
(EBME, 2011)<br />
An ECG trace may be obtained with the electrodes attached in a variety of positions, however,<br />
the system of positioning leads for performing a 12-lead ECG is universal. This helps to ensure<br />
that, when a person's ECGs are compared, any changes on the ECG are due to cardiac injury, not<br />
a difference in placement of leads, this is extremely important with the increasing use of foreign<br />
travel. There are universal standards in place throughout the world.
These positions may differ slightly when a patient is on continuous cardiac monitoring. The<br />
leads routinely attached to wrists and ankles will be placed on shoulders and lower abdomen so<br />
that movement of limbs has minimal effect on the rhythm trace. These positions may also differ<br />
if a patient is shaking (maybe due to Parkinson's disease or hypothermia) or has muscle<br />
tremors. In this situation the leads may be moved onto the thighs and forearms.<br />
There are 10 wires on an ECG machine that are connected to specific parts of the body. These<br />
wires break down into 2 groups:<br />
1. 6 chest leads<br />
2. 4 limb or peripheral leads (one of these is "neutral")<br />
Fig 3 – 12 Lead ECG Placement<br />
The 6 leads are labelled as "V" leads and numbered V1 to V6. They are positioned in specific<br />
positions on the rib cage.<br />
Limb leads are made up of 4 leads placed on the extremities: left and right wrist; left and right<br />
ankle.<br />
Fig 4 – ECG Placement on a patient lying down.
Each of the 12 leads represents a particular orientation in space, as indicated below (RA = right<br />
arm; LA = left arm, LL = left foot):<br />
Bipolar limb leads (frontal plane):<br />
Lead I: RA (-) to LA (+) (Right Left, or lateral)<br />
Lead II: RA (-) to LL (+) (Superior Inferior)<br />
Lead III: LA (-) to LL (+) (Superior Inferior)<br />
Augmented unipolar limb leads (frontal plane):<br />
Lead aVR: RA (+) to [LA & LL] (-) (Rightward)<br />
Lead aVL: LA (+) to [RA & LL] (-) (Leftward)<br />
Lead aVF: LL (+) to [RA & LA] (-) (Inferior)<br />
Unipolar (+) chest leads (horizontal plane):<br />
Leads V1, V2, V3: (Posterior Anterior)<br />
Leads V4, V5, V6:(Right Left, or lateral)<br />
The "aV" stands for Augmented Vector, the last letter refers to a position, which are as<br />
follows:<br />
aV R Augmented Vector Right Right wrist/shoulder<br />
aV L Augmented Vector Left Left wrist/shoulder<br />
aV F Augmented Vector Foot Left foot/lower abdomen<br />
These 3 leads create a triangle with the heart in the middle, as below. The lines into the centre<br />
indicate the line of sight of these leads.<br />
Fig 5 – Einthoven’s Triangle.<br />
The 2 leads situated on the right and left wrist (or shoulders), aV R and aV L respectively, and the<br />
lead situated on the left ankle (or left lower abdomen) aV F, make up a triangle, known as<br />
"Einthoven’s Triangle.
ECG Leads have colour coded bands so that they can be easily identified.<br />
The ECG below shows where these leads are when printed.<br />
Fig 6 – ECG Printout<br />
The ECG is usually recorded on a time scale of 0.04 seconds/mm on the horizontal axis and a<br />
voltage sensitivity of 0.1mV/mm on the vertical axis.<br />
On standard ECG recording paper, 1 small square represents 0.04 seconds and one large square<br />
0.2 seconds.<br />
In the normal ECG waveform the P wave represents atrial depolarisation, the QRS complex<br />
ventricular depolarisation and the T wave ventricular repolarisation.<br />
The P - R Interval is taken from the start of the P wave to the start of the QRS complex. The Q - T<br />
interval is taken from the start of the QRS complex to the end of the T wave. This represents the<br />
time taken to depolarise and repolarise the ventricles. The S - T segment is the period between<br />
the end of the QRS complex and the start of the T wave. All cells are normally depolarised<br />
during this phase.
Fig 7 – ECG Waveforms<br />
ECG Normal Values<br />
P - R interval - 0.12 - 0.2 seconds (3-5 small squares of standard ECG paper)<br />
QRS complex - duration less than or equal to 0.1 seconds (2.5 small squares)<br />
Q - T interval- corrected for heart rate (QTc) QTc = QT/ RR interval less than or equal to 0.44<br />
seconds.<br />
Interference<br />
(EBME, 2012)<br />
There are a number of reasons for ECG rhythm disturbance, which could either be down to the<br />
patient, the leads or down to the environment the patient is being nursed in.<br />
Muscle tremor is something that can occur for a number of reasons, such as:<br />
<br />
<br />
<br />
Shivering due to cold<br />
Rigors<br />
Parkinson’s disease<br />
This can cause the ECG trace to look like the below image.<br />
Fig 8 – ECG Trace showing Muscle Tremor
Another disturbance, emanating directly from the surrounding environment, is electrical<br />
interference. The ECG machine is designed to pick up electrical activity within the heart but<br />
it could pick up electrical activity from other sources, causing a trace as below.<br />
Fig 9 – ECG Trace showing Electrical Interference<br />
Artefact is the name given to disturbances in rhythm monitoring caused by movement of the<br />
electrodes<br />
The movement can be caused in a number of ways.<br />
Fig 10 -ECG Trace showing Artefact<br />
<br />
<br />
<br />
<br />
<br />
If the electrodes have been in place for a prolonged period of time, the moist inner pad<br />
can dry up and the connection becomes poor.<br />
The weight of the leads can pull the electrode away from the skin and contact is lost<br />
intermittently, such as when the patient leans or roles over.<br />
The electrode has come away from the skin and is stuck to an item of clothing<br />
The patient is fiddling with the electrodes.<br />
When the skin is sweaty/dirty/hairy the electrodes may not stick well, resulting in an<br />
unstable trace<br />
However, the machine can also mistake artefact for fatal arrhythmias. In this case the machine<br />
will release a "fatal arrhythmia" alarm.<br />
The cables from the electrodes terminate in a single cable, which is plugged into the port on the<br />
ECG Monitor and are electrically isolated for patient safety. Circuitry also protects the monitor<br />
from any high voltages produced by a defibrillator or electro-cautery signals. The signal is<br />
filtered, amplified, conditioned and processed for display. Pulse rate is also calculated and<br />
displayed, with high and low alarms set with the standard being 150bpm high, 50bpm low.<br />
(The University of Nottingham, 2015)
Maintenance and Service Procedures<br />
Firstly check the ECG leads for noise, this is done by connecting the leads to a patient/ECG<br />
simulator and gently flexing the leads while watching the screen carefully for any noise. If any<br />
noise occurs the leads should be replaced.<br />
The lead select function must then be checked to ensure the monitor switches between all<br />
available leads.<br />
Lead fault detection must then be checked this is done by removing a lead from the simulator,<br />
the ECG machine must recognise this as a lead becoming detached and alarming this to ensure<br />
users don’t mistake a lead becoming detached for a clinical problem.<br />
Check heart rate calibration, set simulator to various rates and ensure monitor reads within<br />
specification, Clinical engineering general specification is for rates under 160bpm ±2bpm and<br />
for rates over 160bpm ±3bpm.<br />
Visually inspect screen gain and scroll rates, also perform a print out and check this.<br />
The 50Hz Filter is checked.<br />
The audible alarms are then checked by lowering and raising the rate on the simulator below<br />
and above the heart rate limits to ensure all alarms are working as expected.<br />
If applicable pacer detection is checked, to do this heart activity and pacer signals are simulated<br />
the monitor should display the heart rate that has been applied and also the pacer signal should<br />
be visible in the waveform. Then set the simulator to only give the pacer signal, the monitor<br />
should recognise this as the pacer signal and heart rate should read as zero.<br />
An Electrical Safety Test is then carried out.
Bibliography<br />
EBME, 2011. EBME - Cardiology. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/10-cardiology<br />
[Accessed Febuary 2014].<br />
EBME, 2012. EBME - Cardiac Monitoring. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/20-cardiac-monitoring<br />
[Accessed March 2014].<br />
Jones, A., 2009. PHYSICAL PRINCIPLES OF INTRA-ARTERIAL BLOOD PRESURE MEASUREMENTS,<br />
s.l.: s.n.<br />
The University of Nottingham, 2015. Cardiology Teaching Resource. [Online]<br />
Available at:<br />
http://www.nottingham.ac.uk/nursing/practice/resources/cardiology/function/index.php<br />
[Accessed June 2015].
Defibrillators<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Defibrillation<br />
Electrical cardioversion and defibrillation have become routine procedures in the management<br />
of patients with cardiac arrhythmias. Cardioversion is the delivery of energy that is<br />
synchronised to the QRS complex, while defibrillation is the non-synchronised delivery of a<br />
shock randomly during the cardiac cycle.<br />
Most defibrillators are energy-based, meaning that the device charges a capacitor to a selected<br />
voltage and then rapidly delivers a pre-specified amount of energy in joules (J) to the<br />
myocardium to treat cardiac arrhythmias. The capacitance of a capacitor is the amount of<br />
electric charge it can store for every volt applied to it. With regard to defibrillators the amount<br />
of energy stored in a capacitor is very important. It can be calculated using the formula<br />
E = ½CV 2 , where E is the energy in joules, C the capacitance in farads and V the voltage<br />
measured in volts. This energy is dissipated in the patient’s body over a small time interval,<br />
about 10 milliseconds or one hundredth of a second.<br />
If the capacitance is 1000 μF and the voltage is 500 V then the stored energy is 125J.<br />
[E = ½ CV 2 ]<br />
E= ½ (1000 × 10 -6 ) x (500 2 ) = 125 J.<br />
Current European Society of Cardiology and AHA guidelines suggest the following initial energy<br />
selection for specific arrhythmias:<br />
<br />
<br />
<br />
<br />
For atrial fibrillation, 120 to 200 joules for biphasic devices and 200 joules for<br />
monophasic devices.<br />
For atrial flutter, 50 to 100 joules for biphasic devices and 100 joules for monophasic<br />
devices.<br />
For ventricular tachycardia with a pulse, 100 joules for biphasic devices and 200 joules<br />
for monophasic devices.<br />
For ventricular fibrillation or pulseless ventricular tachycardia, at least 150 joules for<br />
biphasic devices and 360 joules for monophasic devices.<br />
They also incorporate an inductor to prolong the duration of the delivered current, and a<br />
rectifier to convert alternating current (AC) to direct current (DC). (Knight, 2014)<br />
A defibrillator can deliver a controlled electrical shock to a heart that has a life-threatening<br />
rhythm, such as ventricular fibrillation (VF). In VF, the heart's chaotic activity prevents blood<br />
from pumping adequately or at all. Voltage stored by the defibrillator conducts electrical<br />
current (a shock) through the chest by way of electrodes or paddles placed on the chest. This<br />
brief pulse of current halts the chaotic activity of heart, by depolarising a large part of the heart<br />
muscle terminating the dysrhythmia allowing normal sinus rhythm to be re-established by the<br />
body’s internal pace maker located in the sinoatrial node of the heart, giving the heart a chance<br />
to re-start with a normal rhythm.<br />
Many factors affect the chance of defibrillation success including; placement of the electrode<br />
pads, time elapsed before the first shock is given, and certain health conditions. Successful<br />
defibrillation requires that enough current be delivered to the heart muscle during the shock. If<br />
the transthoracic impedance level is high the heart may not receive enough current for<br />
defibrillation to be successful. Impedance is the body's resistance to the flow of current; some<br />
people naturally have higher impedance than others. Therefore, it may take more current, a<br />
longer shock duration, and/or increased voltage to ensure success. (EBME, 2003)
The shock is delivered via two electrode pads/paddles placed as shown below.<br />
Fig 1 – Placement of Electrode Pads/Paddles<br />
Modern defibrillators may be manual or automated; they generally produce biphasic waveforms<br />
as opposed to monophasic waveforms, which increase safety and efficacy. Miniature<br />
implantable cardioverter-defibrillators (ICD) may be used in patients with recurrent lifethreatening<br />
arrhythmias. (Chaudhari, 2005)<br />
Monophasic Waveforms<br />
This is a type of defibrillation waveform where current flows in one direction. In this waveform,<br />
there is no ability to adjust for patient impedance, and it is generally recommended that all<br />
monophasic defibrillators deliver 200 - 300 J of energy to a maximum of 360J, applied to adult<br />
patients with the assumed average impedance of 50 ohms, to ensure maximum current is<br />
delivered which in the graph below is ≈ 45 amps.<br />
Biphasic Waveforms<br />
Fig 2 – Graphical representation of a Monophasic Waveform<br />
With biphasic shocks, the direction of current flow is reversed near the halfway point of the<br />
electrical defibrillation cycle. Biphasic waveforms were initially developed for use in<br />
implantable defibrillators and have since become the standard in external defibrillators. With<br />
biphasic waveforms there is a lower defibrillation threshold (DFT) that allows reductions of the<br />
energy levels administrated and may cause less myocardial damage.<br />
While all biphasic waveforms have been shown to allow termination of VF at lower current than<br />
monophasic defibrillators, there are two types of waveforms used in external defibrillators.
The waveforms are shown below and will have the desired effect at current values ranging from<br />
approx. 15 – 35 amps.<br />
Fig 3 – Graphical representation of two Biphasic Waveforms<br />
Types of Defibrillator<br />
Automated External Defibrillator (AED)<br />
AEDs are highly sophisticated, microprocessor-based devices that analyse multiple features of<br />
the surface ECG signal including frequency, amplitude, slope and wave morphology. They<br />
contain various filters for QRS signals, radio transmission and other interferences, as well as for<br />
poor electrode contact. Some devices are programmed to detect patient movement.<br />
The typical controls on an AED include a power button, a display screen on which trained<br />
rescuers can check the heart rhythm and a discharge button. Certain defibrillators have special<br />
controls for internal paddles or disposable electrodes.<br />
In AED Mode, the Defibrillator analyses the patient’s ECG and advises you whether or not to<br />
deliver a shock. Voice prompts guide you through the defibrillation process by providing<br />
instructions and patient information. Voice prompts are reinforced by messages/pictures that<br />
appear on the display. (Lozano, 2013)<br />
Manual Defibrillator<br />
Manual defibrillators are designed to give full control to the clinical users. The defibrillator<br />
records the patients ECG, the user then assess the ECG and selects the appropriate level of<br />
energy for defibrillation.<br />
Capnography<br />
End tidal Carbon Dioxide (EtCO 2 ) is the partial pressure or maximal concentration of carbon<br />
dioxide (CO 2) at the end of an exhaled breath, which is expressed as a percentage of CO 2 or<br />
mmHg. The normal values are 5% to 6% CO 2, which is equivalent to 35-45 mmHg. CO 2 reflects<br />
cardiac output and pulmonary blood flow as the gas is transported by the venous system to the<br />
right side of the heart and then pumped to the lungs by the right ventricles. When CO 2 diffuses<br />
out of the lungs into the exhaled air, a device called capnometer measures the partial pressure<br />
or maximal concentration of CO 2 at the end of exhalation.
Capnography uses an EtCO 2 sensor to continuously monitor the carbon dioxide that is inspired<br />
and exhaled by the patient. It is usually presented as a graph of expiratory CO 2 against time, or<br />
less commonly against expired volume. The sensor employs infrared (IR) spectroscopy to<br />
measure the concentration of CO 2 molecules that absorb infrared light. This consists of a source<br />
of infrared radiation, a chamber containing the gas sample, and a photo-detector. When the<br />
expired CO 2 passes between the beam of infrared light and photo-detector, the absorbance is<br />
proportional to the concentration of CO 2 in the gas sample. The gas samples can be analysed by<br />
the mainstream (in-line) or side-stream (diverting) techniques. (Physio-Control, 2013)<br />
During CPR, the amount of CO 2 excreted by the lungs is proportional to the amount of<br />
pulmonary blood flow; therefore capnography can be used to monitor the effectiveness of CPR<br />
and as an early indication of the Return of Spontaneous Circulation (ROSC).<br />
It has been shown that when a patient experiences ROSC the first indication is often a sudden<br />
rise in EtCO 2 as the rush of circulation washes un-transported CO 2 from tissues, likewise a<br />
sudden drop in EtCO 2 may indicate that the patient has lost pulse and CPR may need to be<br />
restarted. (Paramedicine, 2000)<br />
Maintenance & Service Procedures<br />
Defibrillators are serviced annually, during the service functional checks of all controls, displays<br />
and sound outputs are performed. ECG functions are checked including heart rate calibration<br />
and lead off detection, most defibrillators can detect whether the paddles/pads are connected<br />
or disconnected and this should also be checked.<br />
An analyser is used to ensure output energy levels are within specification and a check of all<br />
functions/analysis is performed when in AED mode.<br />
Pacer function and pacer detection are both tested, a functional check of the capnography (if<br />
applicable) and finally an electrical safety test is performed.<br />
There is also scheduled battery and patient lead replacement, the expiry dates on the pads<br />
should be checked to ensure they are still ok to use. If they are past there expiry date the ward<br />
staff should be informed and the pads removed from use and replaced.<br />
During maintenance & service procedures it is vital to ensure a defibrillator is never left alone<br />
charged. When repairing or opening the case for any reason it is important to follow the<br />
manufacturer’s guidelines for discharging the capacitor to ensure no harm comes to yourself or<br />
others.
Bibliography<br />
Chaudhari, M., 2005. Anaesthesia Journal. [Online]<br />
Available at: http://www.anaesthesiajournal.co.uk/article/S1472-0299(06)00175-5/abstract<br />
[Accessed September 2015].<br />
EBME, 2003. EBME - Biphasic Defibrillator. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/12-biphasicdefibrillation?showall=&start=3<br />
[Accessed September 2015].<br />
Knight, B. P., 2014. UpToDate. [Online]<br />
Available at: http://www.uptodate.com/contents/basic-principles-and-technique-ofcardioversion-and-defibrillation<br />
[Accessed September 2015].<br />
Lozano, I. F., 2013. Principles of External defibrillators. [Online]<br />
Available at: http://www.heartrhythmcharity.org.uk/www/media/files/InTech-<br />
Principles_of_external_defibrillators.pdf<br />
[Accessed September 2015].<br />
Paramedicine, 2000. End Tidal CO2. [Online]<br />
Available at: http://www.paramedicine.com/pmc/End_Tidal_CO2.html<br />
[Accessed October 2015].<br />
Phillps Medical Systems, 2005. M4735A (ELD) Heartstream XL Defibrillator Service/User Manual,<br />
Physio-Control, 2013. Lifepak® 20e Defibrillator Service/User manual.
Diagnostic Ultrasound<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Diagnostic Ultrasound<br />
Ultrasound is the term used to describe sound of frequencies above 20 kilohertz (kHz), at this<br />
frequency it is beyond the range of human hearing.<br />
Diagnostic ultrasound is an imaging technique used for visualising all body regions that are not<br />
situated behind expanses of bone or air-containing tissue, such as the lungs. Examinations<br />
through thin, flat bones are possible at lower frequencies. It is also possible to bypass obstacles<br />
with endoscopes (endoscopic sonography). Frequencies of 2–20 Megahertz (MHz) are typical<br />
for diagnostic ultrasound.<br />
Transcutaneous ultrasound is used mainly for evaluating:<br />
<br />
<br />
<br />
<br />
<br />
Neck: thyroid gland, lymph nodes, abscesses, vessels (angiology).<br />
Chest: wall, pleura, peripherally situated disorders of the lung, mediastinal tumours<br />
(The mediastinum is the cavity that separates the lungs from the rest of the chest. It<br />
contains the heart, esophagus, trachea, thymus, and aorta), and the heart as a whole<br />
(echocardiography).<br />
Abdomen: The abdominal cavity is the space bounded by the vertebrae, abdominal<br />
muscles, diaphragm, and pelvic floor. The intraperitoneal space located within the<br />
abdominal cavity, but wrapped in peritoneum (membrane that forms the lining of the<br />
abdominal cavity). The structures within the intraperitoneal space e.g. the stomach.<br />
The structures in the abdominal cavity that are located behind the intraperitoneal space<br />
"retroperitoneal" e.g. the kidneys, and those structures below the intraperitoneal space<br />
called "subperitoneal" or "infraperitoneal" e.g. the bladder and small pelvis: organs, fluid<br />
containing structures, gastrointestinal tract, great vessels and lymph nodes, tumours<br />
and abnormal fluid collections.<br />
Extremities (joints, muscles and connective tissue, vessels).<br />
Obstetric sonography is commonly used during pregnancy.<br />
Diagnostic ultrasound imaging depends on the computerised analysis of reflected ultrasound<br />
waves, which non-invasively build up fine images of internal body structures.<br />
Higher frequencies have a shorter wavelength and are therefore capable of reflecting and<br />
scattering off smaller structures giving higher resolution, however, the use of high frequencies<br />
is limited by their greater attenuation (loss of signal strength) in tissue and thus shorter depth<br />
of penetration limiting the depth when producing an image. Lower frequencies produce less<br />
resolution but image deeper into the body. For this reason, different ranges of frequency are<br />
used for examining different parts of the body values typically used are:<br />
<br />
<br />
<br />
<br />
<br />
<br />
2.5 MHz - deep abdomen, obstetric and gynaecological imaging<br />
3.5 MHz - general abdomen, obstetric and gynaecological imaging<br />
5.0 MHz - vascular, breast, pelvic imaging<br />
7.5 MHz - breast, thyroid<br />
10.0 MHz - breast, thyroid, superficial veins, superficial masses, musculoskeletal<br />
imaging.<br />
15.0 MHz - superficial structures, musculoskeletal imaging.<br />
(Morgan, 2015)
Transducer<br />
Ultrasound waves are produced by a transducer (a hand-held probe), which can both emit<br />
ultrasound waves, as well as detect the ultrasound echoes reflected back. The probe produces<br />
the ultrasound waves and receives the echoes using a principle called the piezoelectric effect.<br />
Inside the probe there are one or more piezoelectric crystals, piezoelectric crystals are able to<br />
produce sound waves when an electric current passes through them, but can also work<br />
in reverse, producing electricity when a sound wave hits them. This happens as the mechanical<br />
and electrical energy causes them to change shape, by contracting or expanding.<br />
When used in an ultrasound scanner, the transducer sends out a directed beam of sound waves<br />
into the body, and the sound waves are reflected back to the transducer from the tissues and<br />
organs in the path of the beam. When these echoes hit the transducer, they generate electrical<br />
signals that the ultrasound scanner converts into images of the tissues and organs.<br />
The probe has a sound absorbing substance to eliminate back reflections from the probe itself,<br />
and an acoustic lens to help focus the emitted ultrasound waves.<br />
Ultrasound Techniques<br />
The echo principle forms the basis of all common ultrasound techniques. The distance between<br />
the transducer and the reflector or scatterer in the tissue is measured by the time between the<br />
emission of a pulse and reception of its echo. Additionally, the intensity of the echo can be<br />
measured. With Doppler techniques, comparison of the Doppler shift of the echo with the<br />
emitted frequency gives information about any movement of the reflector. The various<br />
ultrasound techniques used are described below.<br />
A-mode<br />
A-mode (A-scan, amplitude modulation) is a one-dimensional examination technique in which a<br />
transducer with a single crystal is used. The echoes are displayed on the screen along a time<br />
(distance) axis as peaks proportional to the intensity (amplitude) of each signal. The method is<br />
rarely used today, as it conveys limited information, e.g. measurement of distances. For an<br />
example, ophthalmologists can use it to measure the diameter of the eye ball<br />
B-mode<br />
B-mode (brightness modulation) is a similar technique, but the echoes are displayed as points of<br />
different grey-scale brightness corresponding to the intensity (amplitude) of each signal.<br />
B-scan, 2D<br />
The arrangement of many (e.g. 256) one-dimensional lines in one plane makes it possible to<br />
build up a two-dimensional (2D) ultrasound image (2D B-scan). The single lines are generated<br />
one after the other by moving (rotating or swinging) transducers or by electronic multi-element<br />
transducers.<br />
Electronic transducers are made from a large number of separate elements arranged on a plane<br />
(linear array) or a curved surface (curved array). A group of elements is triggered<br />
simultaneously to form a single composite ultrasound beam that will generate one line of the<br />
image. The whole two-dimensional image is constructed step-by-step, by stimulating one group<br />
after the other over the whole array.<br />
The lines can run parallel to form a rectangular (linear array) or a divergent image (curved<br />
array)
Three- and four-dimensional techniques<br />
The main prerequisite for construction of three-dimensional (3D) ultrasound images is very fast<br />
data acquisition. The transducer is moved by hand or mechanically perpendicular to the<br />
scanning plane over the region of interest.<br />
The collected data are processed at high speed, so that real-time presentation on the screen is<br />
possible. This is called the four-dimensional (4D) technique (4D = 3D + real time). The 3D image<br />
can be displayed in various ways, such as transparent views of the entire volume of interest or<br />
images of surfaces, as used in obstetrics and not only for medical purposes. It is also possible to<br />
select two-dimensional images in any plane, especially those that cannot be obtained by a 2D B-<br />
scan.<br />
M-mode or TM-mode<br />
Fig 1 - 3D Ultrasound scan Image<br />
M-mode or TM-mode (time motion) is used to analyse moving structures, such as heart valves.<br />
The echoes generated by a stationary transducer (one-dimensional B-mode) are recorded<br />
continuously over time.<br />
The ultrasound images can show:<br />
Presence, position, size and shape of organs.<br />
Stasis, concretions and dysfunction of hollow organs and structures.<br />
Tumour diagnosis and differentiation of focal lesions.<br />
Inflammatory diseases.<br />
Metabolic diseases causing macroscopic alterations of organs.<br />
Abnormal fluid collection in body cavities or organs, including ultrasound-guided<br />
diagnostic and therapeutic interventions;<br />
Evaluating transplants;<br />
Diagnosis of congenital defects and malformations.<br />
Additionally, ultrasound is particularly suitable for checks in the management of chronic<br />
diseases and for screening, because it is low risk, comfortable for patients and cheaper than<br />
other imaging modalities. (World Health Organization, 2011)<br />
Signal Processing and image capture<br />
The main ultrasound system takes care of processing signals from the ultrasound probe and<br />
capturing and recording images. The computer based system runs dedicated software that<br />
controls the ultrasound probes, processes the signal returned from the probe, displays the<br />
image and allows the image to be captured on internal storage or sent through the hospital<br />
network
Maintenance & Service Procedures<br />
A visual inspection paying attention to the case, cables, any foot pedals and the ultrasound<br />
probe should be carried out, also checking that any ventilation grilles are free from dust/debris.<br />
The ultrasound system should be running quietly when switched on and a check that all the<br />
buttons are functioning and the screen displays a clear image.<br />
Check for ‘dropout’ (i.e. crystal failure) on all probes using a paperclip ensuring that it is<br />
displayed as expected. If an ultrasound phantom is available confirm the following;<br />
<br />
<br />
Image quality (uniformity, no axial banding, excessive noise etc.)<br />
Ultrasound phantom targets imaged as expected.<br />
An EST should then be performed on the system with the probe placed in a saline solution with<br />
the applied part lead.
Bibliography<br />
Morgan, D. M. A., 2015. Ultrasound Frequencies, s.l.: NIBIB.<br />
NIBI, 2012. National Institute of Biomedical Imaging. [Online]<br />
Available at: http://www.nibib.nih.gov/science-education/science-topics/ultrasound<br />
[Accessed 2016].<br />
Stern, B., 2016. Basic Concepts of Utrasound Scanning. [Online]<br />
Available at: http://www.yale.edu/ynhti/curriculum/units/1983/7/83.07.05.x.html<br />
[Accessed 2016].<br />
World Health Organization, 2011. Manual of Diagnostic Ultrasound. 2nd ed. Geneva: WHO Press.
Electro-Surgery Equipment<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Electro-Surgery<br />
Electro-surgery is the application of a high-frequency electric current to biological tissue as a<br />
means to cut, coagulate (form blood clots), desiccate (remove water), or fulgurate (destroy and<br />
remove) tissue. Its benefits include the ability to make precise cuts with limited blood loss.<br />
Electrosurgical devices are frequently used during surgical operations in hospital operating<br />
rooms or in outpatient procedures.<br />
Electro-surgery is performed using an electrosurgical generator (also referred to as a power<br />
supply or waveform generator), in a circuit made up of an active electrode, the patient and an<br />
active return electrode. The tissue is heated by an AC current; the AC current is used to directly<br />
heat the tissue itself, whereas in electro-cautery the tip of the instrument is heated.<br />
Fig 1 – Mono-polar Electro-surgery<br />
Fig 2 – Bi-polar Electro-surgery<br />
Standard electrical current alternates at a frequency of 50Hz, at this frequency the current<br />
transmitted to the body tissue would cause excessive neuromuscular stimulation and possibly<br />
electrocution. Nerve and muscle stimulation cease at 100 kHz so at frequencies above this,<br />
typically 200 kHz – 3.3 MHz, electro-surgery can be performed with the energy passing through<br />
the patient with minimal neuromuscular stimulation and no risk of electrocution meaning<br />
electro-surgery can be performed safely.
Mono-polar<br />
Mono-polar is the most commonly used electro-surgical mode. In mono-polar electro-surgery,<br />
the active electrode is in the surgical site, with the patient return electrode somewhere else on<br />
the patient’s body. The current passes through the patient as it completes the circuit from the<br />
active electrode to the return electrode. The return electrode placement should be on<br />
conductive tissue over a large surface area to prevent a build-up of heat which could potentially<br />
cause a burn. Other factors that may cause the return electrode to be compromised are<br />
excessive hair, adipose tissue (fat), bony prominences, fluid invasion, metal implants and scar<br />
tissue.<br />
Fig 3 – Patient return electrode placement<br />
Return Electrode Contact Quality Monitoring (RECQM)<br />
RECQM was developed to protect patients from burns due to inadequate contact of the return<br />
electrode pad site by using a split plate system. RECQM equipped generators actively monitor<br />
the amount of impedance between the two split plates. If the impedance is high, due to poor<br />
contact on either side of the split plate, the electrosurgical unit will alarm to warn the clinical<br />
user of the problem. If the impedance between the two plates is very high or not symmetrical<br />
the unit will alarm and the system will deactivate the generator before an injury can occur<br />
Fig4 – Return Electrode Contact Quality Monitoring<br />
(RECQM)
Bi-polar<br />
In bipolar electro-surgery, both the active electrode and return electrode functions are<br />
performed at the site of the surgery. The tines of the forceps perform the active & return<br />
electrode with only the tissue being grasped included in the circuit. As one of the tines of the<br />
forceps performs the return function no patient return electrode is needed. Bipolar diathermy<br />
is perceived to be safer as the current pathway is much shorter than that utilised in mono-polar<br />
diathermy. Bipolar diathermy is generally utilised in the following situations:<br />
<br />
<br />
<br />
<br />
When coagulation only is required.<br />
When coagulation is required in peripheral areas of the body such as hands or feet or<br />
other areas where channelling (tissue damage caused by heat) may occur.<br />
In procedures where pinpoint or micro coagulation is required.<br />
When a patient has a pacemaker in situ.<br />
Electrosurgical Modes<br />
Electrosurgical generators are able to produce a variety of electrical waveforms. As waveforms<br />
change, so will the corresponding tissue effects. Using a constant waveform, like cut, the<br />
surgeon is able to vaporize or cut tissue. This waveform produces heat very rapidly.<br />
Using an intermittent waveform, like coagulation, causes the generator to modify the waveform<br />
so that the duty cycle (on time) is reduced. This interrupted waveform will produce less heat.<br />
Instead of tissue vaporization, a coagulum is produced.<br />
A blended current is not a mixture of both cutting and coagulation current but rather a<br />
modification of the duty cycle. In the example below, as you go from Blend 1 to Blend 3 the<br />
duty cycle is progressively reduced. A lower duty cycle produces less heat. Consequently, Blend<br />
1 is able to vaporize tissue with minimal haemostasis whereas Blend 3 is less effective at cutting<br />
but has maximum haemostasis.<br />
Fig 5 – Electrosurgical Modes<br />
The only variable that determines whether one waveform vaporizes tissue and another<br />
produces a coagulum is the rate at which heat is produced. High heat produced rapidly causes<br />
vaporization. Low heat produced more slowly creates a coagulum.<br />
Any one of the five waveforms can accomplish both tasks by modifying the variables that impact<br />
tissue effect.
The electrosurgical generator or unit can produce three distinct surgical effects, known as<br />
fulguration, desiccation and cutting. The electrosurgical generator creates different wave forms<br />
which are determined by the setting on the machine, universally known as COAG, CUT and<br />
BLEND. The settings and desired effects are linked as follows:<br />
Cutting<br />
The CUT waveform is a continuous waveform at a lower voltage but higher current than COAG.<br />
This creates a high density of current in a specific tissue area within a short period of time<br />
causing cellular fluid to burst into steam and disrupt the structure which results in vaporisation<br />
of tissue. This results in cell explosion due to the localised but intense heat<br />
Coagulation<br />
Fig 6 – Electrosurgical Cut<br />
Coagulation is performed using waveforms with a lower average power, generating insufficient<br />
heat for explosive vaporisation, but performing a thermal coagulum instead.<br />
Fulguration<br />
When the COAG waveform is used at a high power setting it will create the effect known as<br />
fulguration. The high power generates sparks which create intermittent heating of tissue<br />
causing cells to dry out quickly rather than explode into steam. In fulguration mode the<br />
electrode is held away from the tissue so that when the air gap between the electrode and the<br />
tissue is ionised an electric arc discharge develops. Fulguration coagulates and chars the tissue<br />
over a wide area. In this application the burning of the tissue is more superficial; therefor this<br />
technique is used for very superficial or protrusive lesions such as skin tags.<br />
Desiccation<br />
Fig 7 – Electrosurgical Coagulate<br />
Electrosurgical desiccation occurs when the electrode is in direct contact with the tissue.<br />
Desiccation is achieved most efficiently with the cutting current. By touching the tissue with the<br />
electrode, the current concentration is reduced. Less heat is generated and no cutting action<br />
occurs. The cells dry out and form a coagulum rather than vaporize and explode.
Many surgeons routinely cut with the coagulation current. Likewise, you can coagulate with the<br />
cutting current by holding the electrode in direct contact with tissue. It may be necessary to<br />
adjust power settings and electrode size to achieve the desired surgical effect. The benefit of<br />
coagulating with the cutting current is that you will be using far less voltage. Likewise, cutting<br />
with the cut current will also accomplish the task with less voltage. This is an important<br />
consideration during minimally invasive procedures.<br />
Variables Impacting Tissue Effect<br />
In addition to waveform and power setting, other variables impact tissue effect. They include:<br />
Size of the electrode: The smaller the electrode, the higher the current concentration.<br />
Consequently, the same tissue effect can be achieved with a smaller electrode, even though the<br />
power setting is reduced.<br />
Time: At any given setting, the longer the generator is activated, the more heat is produced. And<br />
the greater the heat, the farther it will travel to adjacent tissue (thermal spread).<br />
Manipulation of the electrode: This can determine whether vaporization or coagulation<br />
occurs. This is a function of current density and the resultant heat produced while sparking to<br />
tissue versus holding the electrode in direct contact with tissue.<br />
Type of Tissue: Tissues vary widely in resistance.<br />
Eschar: Eschar (dead tissue produced by a thermal burn) is relatively high in resistance to<br />
current. Keeping electrodes clean and free of eschar will enhance performance by maintaining<br />
lower resistance within the surgical circuit.<br />
Fig 8 – Electrosurgical Mode Effects
Argon-Enhanced Electro-surgery<br />
Argon-enhanced electro-surgery incorporates a stream of argon gas to improve the surgical<br />
effectiveness of the electrosurgical current.<br />
Argon gas is inert and non-combustible, making it a safe medium through which to pass<br />
electrosurgical current. Electrosurgical current easily ionizes argon gas, making it more<br />
conductive than air. This highly conductive stream of ionized gas provides the electrical current<br />
an efficient pathway.<br />
There are many advantages to argon-enhanced electrosurgical cutting and coagulation.<br />
<br />
<br />
<br />
<br />
<br />
Non-combustible<br />
Decreased smoke & odour<br />
Non-contact in coagulation mode<br />
Decreased blood loss.<br />
Decreased tissue damage<br />
Fig 9 – Argon enhanced electro-surgery<br />
Surgical Smoke<br />
Surgical smoke is created when tissue is heated and cellular fluid is vaporized by the thermal<br />
action of an energy source. Viral DNA, bacteria, carcinogens and irritants are known to be<br />
present in electrosurgical smoke. Universal precautions indicate a smoke evacuation system<br />
should be used. The National Institute of Occupational Safety and Health (NIOSH) and the<br />
Centre for Disease Control (CDC) have also studied electrosurgical smoke at length. The<br />
organisations concluded: “Research studies have confirmed that this smoke plume can contain<br />
toxic gases and vapours such as benzene, hydrogen cyanide, formaldehyde, bio-aerosols, dead<br />
and live cellular material (including blood fragments) and viruses.”<br />
When electro-surgery is used in the context of minimally invasive surgery, it raises a new set of<br />
safety concerns. Two of these are insulation failure and direct coupling of current.<br />
Direct Coupling<br />
Direct coupling occurs when the user accidentally activates the generator while the active<br />
electrode is near another metal Instrument (scopes, graspers, etc.). Electrical current will flow<br />
from the first conductor into the secondary one and energize it. This energy will seek a pathway<br />
to complete the circuit to the patient return electrode. There is potential for significant patient<br />
injury.
Insulation Failure<br />
Many surgeons routinely use the coagulation waveform. This waveform is comparatively high<br />
in voltage. This voltage can spark through compromised insulation. Also, high voltage can cause<br />
breaks in weak insulation. Breaks in insulation can create an alternate route for the current to<br />
flow. If this current is concentrated, it can cause significant injury. Also radio frequency is not<br />
always confined by insulation and current leakage does occur. It is recommended that Cords<br />
not be wrapped around metal instruments and Cords not be bundled together.<br />
Fig 10 – Possible injury through compromised<br />
insulation<br />
Further precautions to take for safe use are; the electrosurgical device should not be used in the<br />
presence of flammable agents such as alcohol and /or tincture based (alcoholic extract of plant<br />
or animal) agents.<br />
Use of electrosurgical devices should be avoided in oxygen enriched environments<br />
Active electrodes should be placed in clean, dry, well insulated safety holsters when not in use<br />
to avoid any accidental injury if the generator is accidently activated.<br />
Maintenance and service procedures<br />
Maintenance procedures should include a thorough visual inspection of all elements of the<br />
electrosurgical device.<br />
Error logs should be checked and any regular or serious errors should be rectified.<br />
The system that checks pad resistance should be checked to ensure the correct errors occur at<br />
the correct resistances.<br />
The power output should be checked across all setting at various levels and effects.<br />
Argon plasma units should be checked alongside the correct electrosurgical device.<br />
Flow rates of the argon gas output should be measured. All functions should be checked.<br />
A full electrical safety test should be carried out, it is important to consult the service manual/<br />
protocol to ensure the unit is in the correct mode to ensure the output relays are all closed for<br />
the duration of the safety test.
Bibliography<br />
Covidien, 2014. Principles in Electro-surgery. [Online]<br />
Available at: http://www.asit.org/assets/documents/Prinicpals_in_electrosurgery.pdf<br />
[Accessed July 2015].<br />
EBME, 2009. EBME - Apnoea Alarms. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/6-apnoea-alarms<br />
[Accessed March 2014].<br />
Gov.uk, 2014. Electro-surgery Equipment Safety. [Online]<br />
Available at:<br />
https://www.gov.uk/government/uploads/system/uploads/attachment_data/file/403384/Ele<br />
ctro-surgery_equipment_safety_poster.pdf<br />
[Accessed July 2015].<br />
MHRA, n.d. Electrosurgical Module. [Online]<br />
Available at: info.mhra.gov.uk/learning/ESUGenericModule/player.html<br />
[Accessed July 2015].<br />
NATN, 2013. Electro-surgery managing the risk – National Association of Theatre Nurses. [Online]<br />
Available at: www.afpp.org.uk/filegrab/electrosurgeryguidance.pdf?ref=5<br />
[Accessed July 2015].
Multi-Parameter Monitoring<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Multi-parameter Monitors<br />
Multi-parameter Monitors are intended to be used for monitoring, displaying, reviewing, storing<br />
and the transferring of multiple physiological parameters including, ECG, heart rate (HR),<br />
respiration (Resp), temperature (Temp), SPO 2 (pulse oxygen saturation), pulse rate (PR), noninvasive<br />
blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), airways<br />
gases such as; carbon dioxide (CO 2), oxygen (O 2), anaesthetic gas (AG).<br />
Monitoring vital signs for example; a patient’s blood pressure, pulse rate, and respiration rate is<br />
a crucial aspect of patient care in hospital. Vital signs indicate a patient’s clinical condition, are<br />
necessary to calculate national early warning scores (NEWS) and used to determine the<br />
monitoring, escalation and interventions that are required subsequently.<br />
In a hospital setting, patient monitoring is used in operating theatres, intensive care and critical<br />
care units, and many other critical and non-critical areas.<br />
Continuous multi-parameter monitoring has shown to be an effective mechanism for triggering<br />
early detection of changes in a patient’s condition by notifying the nursing staff that the patient<br />
needs attention. This is beneficial as by assessing the situation sooner and making the right<br />
clinical decision to intervene as appropriate.<br />
In addition to monitoring patients on an individual basis, the remote observation of multiple<br />
patients is possible through central patient monitoring systems. These monitoring systems are<br />
typically made up of networked machines consisting of one or more sensors, display devices,<br />
processing components, and communication links for displaying or recording the results<br />
elsewhere through the network. For portable monitors or in areas where a hard wired network<br />
is not practical wireless monitors are used, the signal is sent from the telemetry unit and picked<br />
up by aerial’s located around the hospital. This is particularly useful for areas such as Accident &<br />
Emergency and Coronary Care Unit or where multiple areas need to monitor the same patient at<br />
the same time or while a patient is being transported from one area to another.<br />
In non-critical areas the signs being monitored will predominantly be Heart Rate, SPO 2, Blood<br />
Pressure and ECG. In a critical care area such as ICU or an operating theatre, further signs will<br />
be monitored these may include; Invasive Blood Pressure (IBP), Respiration Rate and airways<br />
gases such as CO 2, O 2, and other gases that may be respired such as anaesthetic gases during a<br />
surgical procedure.<br />
Respiration Rate<br />
The function of the respiratory system is to supply adequate oxygen to the tissues and to<br />
remove the waste product carbon dioxide. This is achieved with the inspiration and expiration<br />
of air. With each breath there is a pause after expiration. The rate of respiration will vary with<br />
age and gender. A respiratory rate of 12-18 breaths per minute in a healthy adult is considered<br />
normal. Rates outside of this normal range can be classed as;<br />
<br />
<br />
<br />
<br />
<br />
Tachypnoea: the rate is regular but over 20 breaths per minute.<br />
Bradypnoea: the rate is regular but less than 12 breaths per minute.<br />
Apnoea: there is an absence of respiration for several seconds - this can lead to<br />
respiratory arrest.<br />
Dyspnoea: difficulty in breathing, the patient gasps for air.<br />
Cheyne-Stokes respiration: the breathing gets increasingly deeper then shallower,<br />
very slow and laboured with periods of apnoea. This type of breathing is often seen in<br />
the dying patient.
Hyperventilation: patients may breathe rapidly due to a physical or psychological<br />
cause, for example if they are in pain or panicking. Hyperventilation reduces the carbon<br />
dioxide levels in the blood, causing tingling and numbness in the hands; this may cause<br />
further distress. In adults, more than 20 breaths a minute is considered moderate, more<br />
than 30 breaths is severe. (Mallett & Dougherty, 2004)<br />
Multi-parameter monitors have a function to measure the respiration rate of the patient. This is<br />
the number of breaths the patient takes per minute. Most multi-parameter monitors record this<br />
by monitoring the resistance between two ECG electrodes connected to the patient. As the<br />
patient breaths in and the chest expands the resistance between the two electrodes increases<br />
and then decreases as the patient exhales and the chest contracts. Others ways to monitor<br />
respiration could be using capnography which involves CO 2 measurements, referred to as EtCO 2<br />
or end-tidal carbon dioxide concentration. The respiratory rate monitored as such is called<br />
AWRR or airway respiratory rate. Other monitors may record spirometry flow volume loops,<br />
which will show the flow, volume and time taken to inhale and exhale, or it may be monitored<br />
by simply counting the number of breaths in one minute by recording how many times the chest<br />
rises.<br />
Capnography<br />
Capnography is the measurement of carbon dioxide (CO 2) in exhaled breath; capnography gives<br />
medical professionals another tool for determining whether blood is flowing to vital organs like<br />
the heart and brain. CO 2 levels reflect cardiac output and pulmonary blood flow; as the gas is<br />
transported by the venous system to the right side of the heart and then pumped to the lungs by<br />
the right ventricles.<br />
Capnography is particularly important during surgery, where patients are continuously<br />
monitored while under anaesthesia to ensure safety. In addition to its use in surgical settings,<br />
capnography can help physicians and emergency medical personnel determine whether a<br />
patient is having a heart attack or hyperventilating. It can also help them determine whether<br />
CPR is working.<br />
The two primary methods used for measuring CO 2 in expired air are mass spectroscopy and<br />
infrared spectroscopy. In mass spectroscopy gases and vapours of different molecular weights<br />
are separated and a breakdown of what gases and percentages can be displayed.<br />
End tidal Carbon Dioxide (EtCO 2) is the partial pressure or maximal concentration of carbon<br />
dioxide at the end of an exhaled breath, which is expressed as a percentage of CO 2 or in mmHg.<br />
Infrared (IR) spectroscopy uses an EtCO 2 sensor to continuously monitor the carbon dioxide<br />
that is inspired and exhaled by the patient. It is usually presented as a graph of expiratory CO 2<br />
against time, or less commonly against expired volume.<br />
The EtCO 2 sensor consists of an infrared source, a chamber through which the gas sample<br />
passes, and a photo-detector. When the expired CO 2 passes between the beam of infrared light<br />
and photo-detector it leads to a reduction in the amount of light falling on the sensor, this is due<br />
to the principle that CO 2 absorbs infrared radiation. The absorbance is proportional to the<br />
concentration of CO 2 in the gas sample. (Physio-Control, 2013)<br />
Capnometers can be categorised based on the sensing device location. The gas samples can be<br />
analysed by mainstream or side-stream techniques.
Mainstream capnometers are in-line with the patient tubing; the housing is heated to prevent<br />
condensation. The advantage of mainstream analysis is that it gives a real-time measurement<br />
(i.e., an immediate response rate of
eat, and a waveform (a graph of pressure against time) can be displayed.<br />
The components of an intra-arterial monitoring system can be viewed as three main parts:<br />
<br />
<br />
<br />
The measuring apparatus<br />
The transducer<br />
The monitor<br />
The measuring apparatus consists of an arterial cannula connected to tubing containing a<br />
continuous column of saline. The pressure waveform of the arterial pulse is transmitted via the<br />
column of fluid, to a pressure transducer, in the case of intra-arterial monitoring the transducer<br />
consists of a flexible diaphragm which in turn moves strain gauges converting the pressure<br />
waveform into an electrical signal. Monitors amplify the output signal from the transducer, filter<br />
the noise and also display the arterial waveform in real time on a screen. They also usually give<br />
a digital numeric display of systolic, diastolic and mean arterial blood pressure (MAP).<br />
The arterial line is also connected to a flushing system consisting of a bag of saline pressurised<br />
to 300 mm/Hg via a flushing device.<br />
Fig 3 – Invasive Blood Pressure being monitored<br />
For a pressure transducer to read accurately, atmospheric pressure must be discounted from<br />
the pressure measurement. This is done by exposing the transducer to atmospheric pressure<br />
and calibrating the pressure reading to zero. A transducer should be zeroed several times per<br />
day to eliminate any baseline drift.<br />
The pressure transducer must be set at the appropriate level in relation to the patient in order<br />
to measure blood pressure correctly. This is usually taken to be level with the patient’s heart.<br />
Failure to do this results in an error due to hydrostatic pressure (the pressure exerted by a<br />
column of fluid – in this case, blood) being measured in addition to blood pressure. This can be<br />
significant – every 10cm error in levelling will result in a 7.4mmHg error in the pressure<br />
measured; a transducer too low over reads, a transducer too high under reads.<br />
IBP monitoring has numerous advantages; IBP allows continuous ‘beat-to-beat’ blood pressure<br />
monitoring. This is useful in patients who are likely to display sudden changes in blood pressure<br />
(e.g. vascular surgery), patients who require close control of blood pressure (e.g. head injured<br />
patients), or in patients receiving drugs to maintain blood pressure.<br />
The IBP technique also allows accurate blood pressure readings at very low pressures, for<br />
example in shocked patients.<br />
An invasive blood pressure reading could also allow for improvement of patient comfort if blood<br />
pressure monitoring is required for a long period of time. IBP monitoring avoids the trauma of<br />
repeated cuff inflations.<br />
Other advantages include that the arterial cannula is convenient for repeated arterial blood<br />
sampling, for instance for arterial blood gases. (Jones, 2009)
Maintenance & Service Procedures<br />
Invasive Blood Pressure<br />
This is checked using a patient simulator at various pressures, it is essential that a zeroing<br />
procedure is performed before any measurements are taken.<br />
Respiration Rate<br />
This is also checked using a patient simulator at various rates to ensure accuracy.<br />
Capnography & Airways Gases<br />
Capnograpy can be tested by attaching an ETCO 2 circuit and breathing into the airway adapter at<br />
a set rate, the rise and fall of the CO 2 graph will be displayed on the monitor. The accuracy of<br />
any measured airways gases is also to be checked by attaching a cylinder filled with various<br />
gases such as; O 2, CO 2, nitrogen and typically an anaesthetic gas at set percentages. The monitor<br />
should show the levels of the various gases, if the values displayed are not correct a calibration<br />
procedure should be carried out.
Bibliography<br />
EBME, 2014. Capnography. [Online]<br />
Available at: http://www.ebme.co.uk/articles/clinical-engineering/16-capnometrycapnography<br />
[Accessed December 2015].<br />
Fukuda Denshi UK, 2014. DS-8500 System Maintenance Manual V.09 (2014). V 9 ed. Tokyo:<br />
Fukuda Denshi.<br />
Jones, A., 2009. Physical Principles of Intra-Arterial Blood Pressure Measurements, Salford: s.n.<br />
Mallett , J. & Dougherty, L., 2004. The Royal Marsden Hospital Manual of Clinical Nursing<br />
Procedures. 6th ed. Oxford: Blackwell Science.<br />
Paramedicine, 2000. End Tidal CO2. [Online]<br />
Available at: http://www.paramedicine.com/pmc/End_Tidal_CO2.html<br />
[Accessed October 2015].<br />
Physio-Control, 2013. Lifepak® 20e Defibrillator Service/User manual.<br />
Tilakaratna, P., 2014. How Equipment Works. [Online]<br />
Available at: http://www.howequipmentworks.com/capnography/<br />
[Accessed December 2015].
Neonatal Phototherapy<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Neonatal Phototherapy<br />
Jaundice is one of the most common conditions needing medical attention in newborn babies.<br />
Jaundice refers to yellow colouration of the skin and the sclerae (white outer coat of the<br />
eyeball). The yellow coloration of the skin and sclera in newborns with jaundice is the result of<br />
an accumulation of unconjugated bilirubin. In most infants, unconjugated bilirubin reflects a<br />
normal transitional phenomenon. However, in some infants, serum bilirubin levels may rise<br />
excessively, a condition known as hyper-bilirubinaemia, which can be cause for concern<br />
because unconjugated bilirubin is neurotoxic and can cause death in newborns and lifelong<br />
neurologic complications (bilirubin encephalopathy) in infants who survive. The term<br />
kernicterus is used to denote the clinical features of acute or chronic bilirubin encephalopathy.<br />
Newborns are especially vulnerable to hyper-bilirubinemia induced neurological damage and<br />
therefore must be carefully monitored for alterations in their serum bilirubin levels. (GOSH,<br />
2014)<br />
Neonatal phototherapy is a widely used and accepted form of treatment for neonatal hyperbilirubinaemia.<br />
Effective phototherapy needs to satisfy three important criteria; effective<br />
spectrum, sufficiently high irradiance and large effective treatment area. The aim of<br />
phototherapy is to lower the concentration of circulating bilirubin or keep it from increasing.<br />
Phototherapy achieves this by using light energy to change the shape and structure of bilirubin,<br />
converting it to molecules that can be easily excreted.<br />
Fig 1 – Effect of Blue Light on Bilirubin<br />
Bilirubin absorbs light most strongly in the blue region of the spectrum near 460nm. Only<br />
wavelengths that penetrate tissue and are absorbed by bilirubin have a phototherapeutic effect.<br />
Phototherapy units with outputs predominantly in the 460-490nm blue region of the spectrum<br />
are the most effective for treating hyper-bilirubinaemia.<br />
Blue light (400-480nm) has the potential to cause photochemical induced retinal injury, usually<br />
during therapy the neonate’s eyes will be covered using commercial or in-house eye protection.<br />
The baby’s temperature must also be carefully monitored during treatment because there may<br />
be increased heat from the phototherapy lamps. (Wentworth, 2005)
Maintenance and Service Procedures<br />
Phototherapy units require a thorough visual inspection, a light output check every 3-6 months<br />
and a filter clean and electrical safety test. During service or maintenance procedures it is<br />
important to wear the correct PPE; in this case the most important piece is the correct eye<br />
protection, and safety glasses that block the blue light from entering the eye should be worn.<br />
Fig 2 – Example of Blue Light Blocking Glasses (Amber Lens)<br />
Bibliography<br />
GOSH, 2014. GOSH - Neonatal Jaundice and Phototherapy. [Online]<br />
Available at: http://www.gosh.nhs.uk/health-professionals/clinical-guidelines/neonataljaundice-and-phototherapy<br />
[Accessed August 2015].<br />
NHS, 2013. Jaundice Newborns & Phottherapy. [Online]<br />
Available at: http://www.nhs.uk/conditions/jaundice-newborn/pages/treatment.aspx<br />
[Accessed August 2015].<br />
NICE, 2014. NICE Guidance. [Online]<br />
Available at: https://www.nice.org.uk/guidance/cg98/chapter/Introduction<br />
[Accessed August 2015].<br />
Wentworth, S. D., 2005. Neonatal Phototherapy, Cardiff: Rookwood.
Video Stacks & Endoscopy<br />
Systems<br />
Steven Lewis<br />
Clinical Engineering<br />
United Lincolnshire Hospital Trust
Endoscopy Systems & Video Stacks<br />
Endoscopy is a minimally invasive medical procedure directly visualising any part of the inside<br />
of the body using an endoscope, this will be used with a stack containing various equipment<br />
such as a monitor, light source, insufflator, camera, and printer all connected through an<br />
isolation transformer. These stacks are used in theatres and various clinics, such as Ear, Nose,<br />
and Throat (ENT).<br />
Endoscopes<br />
An endoscope is a long, thin, rigid or flexible tube that consists of two or three main optical<br />
cables, each of which comprises up to 50,000 separate optical fibres (made from optical-quality<br />
glass or plastic). One or two of the cables carry light down into the patient's body; another one<br />
carries reflected light (the image of the patient's body) back up to the physician's eyepiece (or<br />
into a camera, which can display it on a monitor).<br />
The optics of an endoscope are similar to those in a telescope. At the remote (distal) end, there<br />
is an objective lens, which links to one or more bendy sections of fibre-optic cable that carry the<br />
light back out of the patient's body to a second lens in an eyepiece or to a camera, which can be<br />
manipulated to adjust the focus.<br />
Endoscopes can be inserted into the body through a natural opening, such as the mouth and<br />
down the throat, or through the anus. Alternatively, an endoscope can be inserted through a<br />
small surgical cut made in the skin (known as keyhole surgery). Images of the inside of the body<br />
are relayed to a Monitor. The instrument may not only provide an image for visual inspection<br />
and photography, but may also be capable of taking biopsies or the retrieval of foreign objects.<br />
Some of the most commonly used types of endoscope include:<br />
<br />
<br />
<br />
<br />
Colonoscopes: used to examine the large intestine (colon).<br />
Gastroscopes: used to examine the oesophagus and stomach.<br />
Endoscopic Retrograde Cholangiopancreatography (ERCP): used to check for<br />
gallstones.<br />
Broncoscopes: used to examine the lungs and airways.<br />
Other types of endoscope include:<br />
<br />
<br />
<br />
Arthroscopes: used to examine joints.<br />
Hysteroscopes: used to examine the womb (uterus).<br />
Cystoscopes: used to examine the bladder.<br />
Endoscopy can involve;<br />
<br />
<br />
<br />
<br />
The Gastrointestinal Tract (GI tract): oesophagus, stomach and duodenum<br />
(esophagogastroduodenoscopy), small intestine, colon, (colonoscopy, proctosigmoidoscopy).<br />
The Respiratory Tract: the nose (rhinoscopy), the lower respiratory tract<br />
(bronchoscopy), the urinary tract (cystoscopy).<br />
The Female Reproductive System: The cervix (colposcopy), the uterus<br />
(hysteroscopy), the Fallopian tubes (Falloscopy).<br />
Normally closed body cavities (through a small incision): the abdominal or pelvic<br />
cavity (laparoscopy), the interior of a joint (arthroscopy), Organs of the chest<br />
(thoracoscopy and mediastinoscopy).
An endoscopy is normally carried out while a patient is conscious. It is not usually painful, but<br />
can be uncomfortable, so a local anaesthetic or sedative may be given.<br />
The exception is keyhole surgery, such as a laparoscopy or an arthroscopy, which are performed<br />
under general anaesthetic. (Martin, 2014)<br />
Endoscopy procedures are usually safe, and the risk of serious complications is low. Possible<br />
complications of an endoscopy include an infection in the part of the body that the endoscope is<br />
used to examine, damage to the body part which may in turn cause excessive bleeding.<br />
Monitors<br />
Monitors used on modern video stacks are high quality high definition (HD) flat panel screens,<br />
with excellent colour reproduction and a wide viewing angle. Most will have various video<br />
connections on the back such as S-Video, BNC, DVI, HDMI, etc. Some older video stacks may<br />
have a non HD flat panel monitor or a CRT monitor.<br />
Light Source & Lamps<br />
The light source may form part of a video or endoscopic stack, be a stand-alone unit often used<br />
in ENT procedures, or be attached to a microscope providing illumination to the area under<br />
investigation, the light is usually transmitted via a fibre optic cable to the lens end of a scope.<br />
High-Intensity Discharge Lamps<br />
High Pressure Sodium (HPS), Metal Halide, Mercury Vapour and Self-Ballasted Mercury Lamps<br />
are all high intensity discharge lamps (HID). Compared to fluorescent and incandescent lamps,<br />
HID lamps produce a large quantity of light from a relatively small bulb.<br />
HID lamps produce light by striking an electrical arc across tungsten electrodes housed inside a<br />
specially designed inner glass tube. This tube is filled with both gas and metals. The gas aids in<br />
the starting of the lamps, then, the metals produce the light once they are heated to a point of<br />
evaporation. High intensity discharge lamps, such as xenon, are ubiquitous within the field of<br />
endoscopy for minimally invasive surgery and diagnosis. Coupling light into a small diameter<br />
fibre bundle is difficult, so an extremely bright light source is required. Historically, the light<br />
source of choice has been the 180 to 300 W xenon lamp, which could deliver 1,000+ lumens of<br />
light at the distal end of the fibre bundle. While xenon bulbs achieve the technical requirements<br />
for endoscopy, they have a very short life, 500 - 1,000 hours, and can be expensive.<br />
LED Lighting<br />
The medical device industry is constantly changing. New technologies and products enter the<br />
market, replacing outdated or inefficient equipment. LED lighting is one of these, and there are<br />
numerous benefits in using LEDs for medical illumination applications including; longer life, less<br />
heat, dynamic control, lower energy consumption, and in many cases, lower cost. LED<br />
technology has improved significantly over the years, and is currently being integrated into<br />
medical devices, including surgical lighting, exam lights, phototherapy, and endoscopy. Ongoing<br />
advancements in LEDs that are driving the technology’s use in medical devices include<br />
improvement in light intensity, product size and weight; long-term reliability, and<br />
heat/temperature management, a line of high performance LEDs optimised to displace xenon<br />
technology for endoscopy have been developed.
Fibre Optic<br />
The light in a fiber-optic cable travels through the core by constantly bouncing from the<br />
cladding (mirror-lined walls), a principle called total internal reflection. The cladding does not<br />
absorb any light from the core, enabling the light wave to travel great distances.<br />
Fig 1 - Fibre Optic Cable<br />
Light source intensity is either adjusted manually via a dial or slider on the unit. If a camera and<br />
processor are attached they can monitor the amount of light at the scope end and adjusts the<br />
intensity automatically.<br />
Different wavelengths of light are needed for different procedures;<br />
Auto Fluorescence Imaging (AFI) is based on the detection of natural tissue fluorescence<br />
emitted by endogenous molecules (fluorophores) such as collagen, flavins, and porphyrins.<br />
After excitation by a short-wavelength light source, mucosal tissue emits a green fluorescence.<br />
A difference in the intensity of this fluorescence is seen between healthy and unhealthy tissue.<br />
These colour differences in fluorescence emission can be captured in real time during<br />
endoscopy and used for lesion detection or characterisation.<br />
Narrow Band Imaging (NBI) is a powerful optical image enhancement technology that<br />
improves the visibility of blood vessels and other structures on the bladder mucosa. This makes<br />
it an excellent tool for diagnosing bladder cancer during cystoscopy.<br />
White light is composed of an equal mixture of wavelengths. The shorter wavelengths only<br />
penetrate the top layer of the mucosa, while the longer wavelengths penetrate deep into the<br />
mucosa. NBI light is composed of just two specific wavelengths that are strongly absorbed by<br />
haemoglobin.<br />
The shorter wavelength in NBI is 415 nm light, which only penetrates the superficial layers of<br />
the mucosa. This is absorbed by capillary vessels in the surface of the mucosa and shows up<br />
brownish on the video image. This wavelength is particularly useful for detecting tumours,<br />
which are often highly vascularised. The second NBI wavelength is 540 nm light, which<br />
penetrates deeper than 415 nm light. It is absorbed by blood vessels located deeper within the<br />
mucosal layer, and appears cyan on the NBI image. This wavelength allows a better<br />
understanding of the vasculature of suspect lesions. (Olympus, 2014)<br />
Fig 2 – Absorption of Narrow Band Illumination
Insufflator<br />
An Insufflator provides distension of the required cavity for diagnostic or operative procedures.<br />
The insufflator pumps CO 2 into the cavity to a set pressure, this stretches the area being worked<br />
on making it easier to manipulate tools, assess the surgical site or perform the operative<br />
procedure. The maximum pressures of cavity’s should be observed when using an insufflator to<br />
prevent any problems, for example, Prolonged intra-abdominal pressures greater than 20mmHg<br />
should be avoided as they can cause any of the following problems;<br />
<br />
<br />
<br />
Decreased respiration due to pressure on the diaphragm.<br />
Decreased venous return.<br />
Decreased Cardiac output.<br />
CO 2 is now preferred over air as CO 2 is absorbed 150 times faster than the nitrogen in air, and is<br />
promptly eliminated via the lungs. This also means that patients are not subjected to extended<br />
discomfort from bloating and cramping.<br />
Camera System<br />
The camera on a video stack consists of two parts, the camera head, where scopes and light<br />
guides are attached, and a processer unit which feeds the image taken by the camera to a<br />
monitor, image capture device or printer.<br />
The aperture opens and closes to control how much light travels from the subject through a<br />
series of curved lens and through coloured filters focused on a digital sensor, which converts it<br />
into a digital (numerical) format. This is then processed and fed to the image capture system,<br />
monitor and printer.<br />
Image Capture System<br />
Image capture systems are image management systems that have the ability to record images<br />
and videos to their internal hard drive. They can also be connected to the hospital network<br />
where they can be integrated with PACS and other similar systems.<br />
Many image capture systems are generally windows based systems with a dedicated user<br />
interface. Very little maintenance of this system is required, at ULHT the only thing in the way<br />
service and repair we carry out is to take an image of the hard-drive at acceptance in case of any<br />
faults during its life, functional and visual checks and electrical safety testing.<br />
Isolation Transformer<br />
An isolation transformer is a transformer used to transfer electrical power from a source of<br />
alternating current (AC) power to some equipment or device while isolating the powered device<br />
from the power source, usually for safety reasons. Isolation transformers are used to protect<br />
against electric shock, to suppress electrical noise in sensitive devices, or to transfer power<br />
between two circuits which must not be connected. All true transformers are isolating as the<br />
primary and secondary are not connected physically but only by induction. However, in an<br />
isolation transformer the windings are completely insulated from each other to ensure good<br />
isolation.
In a normal supply the earth conductor is referenced to the neutral<br />
connector at source, which gives a potential voltage between live<br />
and earth of 240V. This means that touching either the live or<br />
neutral makes you part of the return path and can result in<br />
electrocution.<br />
With an isolation transformer the output voltage is not referenced<br />
to ground, so you could safely touch the live conductor and ground<br />
and not received a shock. This system is safer but if there is a fault,<br />
the operator touched both live and neutral connections or there<br />
was capacitive coupling between the secondary windings and<br />
earth. This would allow current to flow from either of the<br />
secondary connections through the operator/ patient to ground<br />
and electrocution could still occur.<br />
Maintenance & Service Procedures<br />
Endoscopes<br />
These are generally managed by the user, this includes cleaning.<br />
Cameras and Image Capture devices<br />
Cameras should have visual inspection and function test, checking image quality and focus also<br />
ensuring the various buttons work as they should.<br />
Image capture devices should also be visually inspected paying attention to the screen and case<br />
all functions should then be checked including the touch screen responds if applicable. This<br />
should then be safety tested.<br />
Light Sources<br />
A thorough visual inspection to include the case, mains lead, and the bulb for any<br />
damage/pitting should be performed. A filter clean/change and an electrical safety test are<br />
often all that is required.<br />
Light sources that used arc lamps or HID lamps require a scheduled lamp change, usually at<br />
approx. 500 hours of use.<br />
Before HID lamps are disposed of in an appropriate manner the xenon gas should be discharged.<br />
Insufflators<br />
A Visual inspection should be performed taking special care to inspect any gas connections and<br />
hoses including the pin index.<br />
A Check of output pressure and flow rates for conformity and accuracy against manufacturer’s<br />
specification. Check all controls work and that any alarms sound, including excess pressure and<br />
empty cylinder alarm.<br />
An Electrical Safety Test (EST) should then be completed to the appropriate standard.
Bibliography<br />
Blackwell Publishing, 2014. Basic Endoscopic Equipment. [Online]<br />
Available at: https://www.blackwellpublishing.com/xml/dtds/4-<br />
0/help/10003420_chapter_1.pdf<br />
[Accessed December 2015].<br />
Martin, T., 2014. Principles of Gastrointestinal Endoscopy. Surgery (Oxford), 32(3), pp. 139-144.<br />
Olympus, 2014. Olympus - Narrow Band Imaging. [Online]<br />
Available at:<br />
http://www.olympus.co.uk/medical/en/medical_systems/applications/urology/bladder/narro<br />
w_band_imaging__nbi_/narrow_band_imaging__nbi_.html<br />
[Accessed Febuary 2014].<br />
ustudy, 2011. Endoscopy Working Principle. [Online]<br />
Available at: http://www.ustudy.in/node/5066<br />
[Accessed December 2015].