cps<br />

Draft, page 32<br />

1.10E-02<br />

1.05E-02<br />

1.00E-02<br />

9.50E-03<br />

9.00E-03<br />

8.50E-03<br />

8.00E-03<br />

Mean: 9.59E-3 cps<br />

Uc: 3.3E-4 cps<br />

%Unc: 3.4%<br />

2007-04-28 2007-08-06 2007-11-14 2008-02-22 2008-06-01 2008-09-09 2008-12-18 2009-03-28 2009-07-06<br />

Figure 19: Background variations<br />

Some laboratories stated that they used gamma detectors with energy range from 200 keV<br />

which might be not suitable for Pb-210 measurement; also some laboratories extrapolated the<br />

efficiency calibration to cover the Pb-210 energy which resulted in not acceptable results. The<br />

most frequent reasons for failure in Pb-210 are discussed in more details in [5, 6].<br />

4.5. Recommendations related to the analysis <strong>of</strong> Ra-226 in moss-soil sample<br />

The majority <strong>of</strong> participating laboratories measured Ra-226 in moss-soil using gamma<br />

spectrometry with high resolution detector <strong>of</strong> different type.<br />

In general, there are two approaches used for Ra-226 determination by gamma spectrometry.<br />

These are the use <strong>of</strong> the 186 keV line from Ra-226, and use <strong>of</strong> gamma lines from the shortlived<br />

progeny Pb-214 and Bi-214. In the former case, the contribution <strong>of</strong> U-235 to the peak<br />

area needs to be allowed for, and if an inappropriate correction is applied, this may lead to<br />

shortcomings in performance.<br />

As shown in Figure 20, most <strong>of</strong> the unacceptable results were due to overestimation <strong>of</strong> the<br />

Ra-226 which could be due to the unresolved U-235 peak or due to an incorrect calibration.<br />

The shortcomings observed in the performance <strong>of</strong> the participating laboratories in gamma<br />

measurements <strong>of</strong> Ra-226 could be attributed to several factors. These factors are related to the<br />

gamma spectrometry procedure utilised by each laboratory. Some <strong>of</strong> these factors are the use<br />

<strong>of</strong> inappropriate calibration standards or procedure, sample geometry, interference from other<br />

gamma lines, and properties <strong>of</strong> the sample container used.<br />