Calibration ModeNoteSeparate calibration of the measuring ranges for chlorine dioxide,bromine and ozone is not possible. The unit uses the calibrationfor the chlorine measuring range. Factor of 1.9 is used to calculatechlorine, factor 2.25 is used to calculate bromine and factor0.67 is used to calculate ozone, respectively, from the chlorinepolynomial.CALFactory calibration active.cALCalibration has been set by the user.Recommended calibration valuesChlorine: between 0.5 and 1.5 mg/lpH: between 7.6 and 8.0Cyanuric acid: between 30 and 60 mg/l24