7. Errors<br />
Error Possible cause Remedy<br />
Time display is off even though Power failure Check both the power cable and main<br />
the unit is switched on voltage<br />
Motor fails to run when the unit<br />
is switched on<br />
Cover of the unit is open Close protective cover<br />
Motor is overheated Allow the unit to cool<br />
Capsule is thrown out of the holder Capsule holder is contaminated Blast capsule holder with compressed air<br />
Capsule is too large or heavy<br />
(refer to Technical Data section)<br />
Use suitable capsules only<br />
Error message “Lo” and error signal Cover was opened during Press “Start/Stop” to quit<br />
are issued mixing<br />
Motor or bearing are defective Press “Start/Stop” to quit. For a short<br />
time, the speed (in revs per sec) at which<br />
the device was turned off is shown in the<br />
display.<br />
Have the unit repaired.<br />
Capsule is too heavy Press “Start/Stop” to quit<br />
NOTE: The <strong>RotoMix</strong> device may be repaired by 3M ESPE or 3M ESPE-authorised personnel only.<br />
8. Maintenance and Care<br />
The device requires no maintenance. Clean all components<br />
with a soft cloth and, if necessary, a mild detergent.<br />
Under no circumstances use solvents or scouring agents<br />
as they will permanently damage the plastic.<br />
Do not allow detergent to enter the mixing unit.<br />
The following agents can be used for disinfection:<br />
Pursept-A (Merz) and FD 322 (Dürr).<br />
Pursept-A and FD 322 are not available in all countries.<br />
Customer Information<br />
No person is authorized to provide any information which<br />
deviates from the information provided in this instruction<br />
sheet.<br />
Warranty<br />
3M ESPE warrants this product will be free from<br />
defects in material and manufacture. 3M ESPE MAKES<br />
NO OTHER WARRANTIES INCLUDING ANY IMPLIED<br />
WARRANTY OF MERCHANTABILITY OR FITNESS FOR<br />
A PARTICULAR PURPOSE. User is responsible for determining<br />
the suitability of the product for user’s application.<br />
If this product is defective within the warranty period,<br />
your exclusive remedy and 3M ESPE’s sole obligation<br />
shall be repair or replacement of the 3M ESPE product.<br />
4<br />
Limitation of Liability<br />
Except where prohibited by law, 3M ESPE will not be<br />
liable for any loss or damage arising from this product,<br />
whether direct, indirect, special, incidental or consequential,<br />
regardless of the theory asserted, including warranty,<br />
contract, negligence or strict liability.<br />
Information valid as of 05/03