Date: 2003-07 <strong>ISO</strong>/<strong>TC</strong> <strong>108</strong>/<strong>SC</strong> 3 N <strong>388</strong>-<strong>CEN</strong>/<strong>TC</strong> <strong>231</strong> N <strong>439</strong><br />
Report of voting on doc. <strong>108</strong>/3 N 367-<strong>CEN</strong>/<strong>TC</strong> <strong>231</strong> N 402<br />
Second <strong>ISO</strong>/CD 8041 "Human response to vibration - Measuring instrumentation"<br />
1 2 (3) 4 5 (6) (7)<br />
Comment (justification for change) by the MB Proposed change by the MB Secretariat observations<br />
on each comment submitted<br />
Type<br />
of<br />
comment2<br />
Paragraph/<br />
Figure/Table/<br />
Note<br />
(e.g. Table 1)<br />
Clause No./<br />
Subclause No./<br />
Annex<br />
(e.g. 3.1)<br />
MB 1<br />
UK 05.02.2 Para 1 ed Clarify definition of tolerance Change to “1% of displayed value” Accepted<br />
Accepted<br />
Replace paragraph with:<br />
“The tolerance limits in this standard include the<br />
associated expanded uncertainties of<br />
measurement calculated for a coverage factor of<br />
2, corresponding to a level of confidence of<br />
approximately 95%, in accordance with the<br />
guidance given in the Guide to the expression of<br />
te Need to specify coverage factor for expanded<br />
uncertainties of measurement, e.g <strong>ISO</strong> 61672-1 para 2 of<br />
Annex A<br />
UK 05.05 Penultimate<br />
para<br />
uncertainty in measurement.”<br />
UK 05.06.5 Equation 11 ed Equation variables need to be in italics Change to variables to italics Accepted<br />
Accepted<br />
f < f1*10^(-2/10)<br />
f1*10^(-2/10)