Viewports<br />

CCS/CRIUS viewports<br />

experiment:<br />

• using a black body<br />

calibrated EpiTT<br />

• measuring identical<br />

temperature ramps<br />

through 5 nominal<br />

identical viewports of<br />

"new type"<br />

• comparing to one "old<br />

type" viewport<br />

about 40K variation!<br />

T Py /°C<br />

950<br />

900<br />

850<br />

800<br />

T 950<br />

T bb<br />

viewport comparison<br />

750<br />

750 800 850 900 950<br />

T bb /°C<br />

23 June 2009 <strong>Temperature</strong> <strong>calibration</strong><br />