LKS Faculty of Medicine Version 1.1<br />

University of Hong Kong<br />

29.Optimize the drop delay again, by adjusting the value in 0.03-drop<br />

increment until the left side stream intensity is consistently > 95%; increase<br />

the flow rate to ≤ 4.0 if necessary. Wait ~ 2 seconds after each change for a<br />

complete response to the delay change.<br />

Note: Two percentage values must add up to 100 %.<br />

30.Click Optical Filter to move the emission filter away from camera.<br />

31.Click Sort to stop sorting Accudrop beads; select Cancel when asked if<br />

preferred to save the report.<br />

32.Reset the Window Extension to 2.00 (for 70µm, 85µm and 100µm<br />

nozzles; 4.00 for 130µm nozzle) on the Laser tab.<br />

33.Click Voltage on the Side Stream Window to turn off the deflection plates,<br />

before unloading the tube.<br />

34.Return to the experiment and select the Tube Pointer.<br />

BD <strong>FACSAria</strong> <strong>SORP</strong> <strong>Operation</strong> Manual<br />