LKS Faculty of Medicine Version 1.1<br />
University of Hong Kong<br />
# If required to change the collecting tubes, click Waste Drawer prior to the<br />
retrieval of the device. Users are reminded to return the waste drawer to<br />
‘SORT’ position following the replacement.<br />
BD <strong>FACSAria</strong> <strong>SORP</strong> <strong>Operation</strong> Manual<br />
During the retrieval of the device<br />
‘SORT’ position following the collecting tube replacement<br />
70.Upon completion, click Sort on the Sort Layout Window and click OK<br />
on the Confirm Prompt; then click ‘Unload’ on Acquisition Dashboard to<br />
unload the current sample tube.<br />
Cleaning the System<br />
71.To 3 x 5mL tubes add 2 ml of FACSClean® Solution, FACSRinse®<br />
Solution, and Milli-Q H2O.<br />
72. Load each tube into the loading chamber.<br />
73.Increase the flow rate to 11.0<br />
74.Clean the tubing by running the three solutions for no less than 5 minutes<br />
each. Repeat the same with FACSRinse® Solution for 10 min, if propidium<br />
iodide (PI) is involved.<br />