06.03.2013 Views

FACSAria SORP Standard Operation Protocol Basic Operation

FACSAria SORP Standard Operation Protocol Basic Operation

FACSAria SORP Standard Operation Protocol Basic Operation

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

LKS Faculty of Medicine Version 1.1<br />

University of Hong Kong<br />

# If required to change the collecting tubes, click Waste Drawer prior to the<br />

retrieval of the device. Users are reminded to return the waste drawer to<br />

‘SORT’ position following the replacement.<br />

BD <strong>FACSAria</strong> <strong>SORP</strong> <strong>Operation</strong> Manual<br />

During the retrieval of the device<br />

‘SORT’ position following the collecting tube replacement<br />

70.Upon completion, click Sort on the Sort Layout Window and click OK<br />

on the Confirm Prompt; then click ‘Unload’ on Acquisition Dashboard to<br />

unload the current sample tube.<br />

Cleaning the System<br />

71.To 3 x 5mL tubes add 2 ml of FACSClean® Solution, FACSRinse®<br />

Solution, and Milli-Q H2O.<br />

72. Load each tube into the loading chamber.<br />

73.Increase the flow rate to 11.0<br />

74.Clean the tubing by running the three solutions for no less than 5 minutes<br />

each. Repeat the same with FACSRinse® Solution for 10 min, if propidium<br />

iodide (PI) is involved.<br />