LKS Faculty of Medicine Version 1.1<br />

University of Hong Kong<br />

18.Adjust the Micrometer Dial (next to the sort block), if necessary, to<br />

obtain evenly brightest bead spots for the four side streams. If possible,<br />

reduce the center core stream bead spot brightness to minimize the<br />

discrepancy amongst the center core stream and the other four.<br />

Core Stream Bead Spots TOO BRIGHT Bead Spots UNEVENLY BRIGHT<br />

19.Check the fridge if an Accudrop bead solution has already been prepared.<br />

Only prepare fresh stock if the label on the tube is NOT WITHIN 2 DAYS.<br />

In brief, dilute 1 drop of Accudrop beads into ~350µL FACSFlow solution,<br />

followed by thorough vortexing. Load the tube into the chamber.<br />

20.In the Laser tab, set the Window Extension to ZERO.<br />

21.Go to Browser Window, and click to ‘Create a new Sort Layout’.<br />