FACSAria SORP Standard Operation Protocol Basic Operation
FACSAria SORP Standard Operation Protocol Basic Operation
FACSAria SORP Standard Operation Protocol Basic Operation
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
LKS Faculty of Medicine Version 1.1<br />
University of Hong Kong<br />
18.Adjust the Micrometer Dial (next to the sort block), if necessary, to<br />
obtain evenly brightest bead spots for the four side streams. If possible,<br />
reduce the center core stream bead spot brightness to minimize the<br />
discrepancy amongst the center core stream and the other four.<br />
Core Stream Bead Spots TOO BRIGHT Bead Spots UNEVENLY BRIGHT<br />
19.Check the fridge if an Accudrop bead solution has already been prepared.<br />
Only prepare fresh stock if the label on the tube is NOT WITHIN 2 DAYS.<br />
In brief, dilute 1 drop of Accudrop beads into ~350µL FACSFlow solution,<br />
followed by thorough vortexing. Load the tube into the chamber.<br />
20.In the Laser tab, set the Window Extension to ZERO.<br />
21.Go to Browser Window, and click to ‘Create a new Sort Layout’.<br />
BD <strong>FACSAria</strong> <strong>SORP</strong> <strong>Operation</strong> Manual<br />
8