Precision, accuracy and linearity <strong>of</strong><br />

ADCC Reporter Bioassay<br />

Antibody Measured Mean<br />

RSD<br />

Sample Potency (%) Potency (%) SD % Recovery (%) (%)<br />

day 1 48.5<br />

day 2 50% 45.2 48.9 3.9 97.7 7.9<br />

day 3 52.9<br />

day 1 63.1<br />

day 2 75% 62.9 66.4 5.9 88.5 8.9<br />

day 3 73.2<br />

day 1 112.1<br />

day 2 125% 136.3 123.0 12.3 98.4 10.0<br />

day 3 120.5<br />

day 1 148.8<br />

day 2 150% 150.4 147.6 3.6 98.4 2.4<br />

day 3 143.6<br />

Confidential and Proprietary. Not for Further Disclosure.<br />

Linearity<br />

y= 1.016x – 0.052<br />

R 2 = 0.995<br />

<strong>Assay</strong> with frozen, thaw-and-use<br />

effec<strong>to</strong>r and WIL2-S target cells<br />

Observed potency<br />

175%<br />

150%<br />

125%<br />

100%<br />

75%<br />

50%<br />

25%<br />

0%<br />

Precision<br />

(Av <strong>of</strong> RSD(%))<br />

7.3%<br />

y = 1.0161x - 0.0516<br />

R² = 0.9949<br />

0% 25% 50% 75% 100% 125% 150% 175%<br />

Expected potency<br />

Accuracy<br />

(Av <strong>of</strong> % Recovery)<br />

95.8%<br />