MIL-STD-3009<br />

APPENDIX A<br />

measurement in order to assure that the photometer meets the requirements of this<br />

document. Records of calibrations and checks shall be maintained by the contractor<br />

and shall be available for Government inspection.<br />

A.4.2 Photometer calibration.<br />

The photometer shall be calibrated using methods that are traceable to NIST<br />

standards.<br />

A.4.2.1 Photometer sensitivity.<br />

The full-scale sensitivity shall be 3.43 cd/m 2 (1.0 fL) or less, with a spot size of no<br />

greater than 0.18 mm (0.007 inch).<br />

A.4.2.2 Photometer accuracy.<br />

The measured luminance of a NIST traceable luminance standard shall be within ±2%<br />

of the NIST certified luminance.<br />

A.4.2.3 Photometer sensitivity and accuracy verification.<br />

The full-scale sensitivity and accuracy of the photometer shall be verified using a NIST<br />

traceable standard of luminance set to a luminance value less than or equal to 3.43<br />

cd/m 2 (1.0 fL) and also equal to the known full-scale sensitivity value of one of the<br />

photometer ranges. Using a spot size no greater than 0.18 mm (0.007 inch) the<br />

photometer full-scale sensitivity shall be within ±2% of the NTIS traceable standard of<br />

luminance value.<br />

A.4.3 Readout resolution.<br />

The unit shall have a digital readout with a resolution better than or equal to 0.1% of full<br />

scale.<br />

A.4.4 Photometer optics.<br />

The optics shall be capable of allowing measurements of spot sizes down to 0.18 mm<br />

(0.007 inch) while meeting the sensitivity requirements of A.4.2.1. The optics shall be<br />

capable of focusing to within 10 cm (4.0 inches).<br />

A.4.5 Photometer viewing <strong>system</strong>.<br />

The viewing <strong>system</strong> must be capable of locating the spot to be measured with a<br />

maximum error of 0.05 mm (0.002 inch).<br />

42