02.04.2013 Views

Calibration Procedure of HACH Pocket Turbidimeter - Up To

Calibration Procedure of HACH Pocket Turbidimeter - Up To

Calibration Procedure of HACH Pocket Turbidimeter - Up To

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Notes:<br />

5.6 Press and hold the “Read” key until the reading is stable. Release the<br />

“Read” key to accept this value.<br />

5.7 Remove this solution from the cell compartment and insert the 1.0 NTU<br />

solution. Close the lid.<br />

5.8 Press “CAL” key until “C1.0” appears on the screen with alternating value<br />

(normally 1.0 NTU).<br />

5.9 Press and hold the “Read” key. When the reading is stable release the<br />

key. The value will be accepted and stored.<br />

5.10 Remove the 1.0 NTU solution from the cell compartment and insert the 20<br />

NTU solution. Close the lid.<br />

5.11 Press “CAL” key until “C20” appears on the screen with alternating value<br />

(normally 20 NTU).<br />

5.12 Press and hold the “Read” key until the reading is stable. Release the key<br />

to save this value.<br />

5.13 Press the “CAL” key to end calibration. “CLd” will be displayed on the<br />

screen. NOTE: If E09 is displayed, then recalibrate the instrument.<br />

If Error messages appear on the screen, check that the standards have been labeled<br />

properly and repeat the calibration. The possible errors are listed with the corrective<br />

actions on pages 32-36 <strong>of</strong> the instrument manual.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!