5.2 <strong>Calibration</strong> in Water Saturated Air<br />
5.2.1 The DO probe to the meter Attach and turn on the instrument.<br />
5.2.2 Prepare the storage and calibration chamber for calibration. This is the black<br />
plastic tube with a small attached lid.<br />
5.2.3 Open the small lid <strong>of</strong> the chamber and insert a wet-soaked sponge. A dry<br />
compact sponge is provided with the supplies. Wet this with distilled water and<br />
insert.<br />
5.2.4 Insert the DO probe into the calibration chamber. The tip <strong>of</strong> the probe should not<br />
be flooded with water. Ensure not to over wet the sponge.<br />
5.2.5 Let the DO probe inserted in the chamber stand for 10 minutes.<br />
5.2.6 Press the DO key to put the meter in the DO reading mode.<br />
5.2.7 Press the “CAL” key.<br />
5.2.8 The display will show 100%. Press the enter key.<br />
5.2.9 The stabilizing icon will appear while the meter completes the calibration.<br />
5.2.10 When calibration is complete, the meter will return to the reading mode.<br />
Notes:<br />
For more information/steps on the calibration procedure, please refer to pages 50-55 <strong>of</strong><br />
the instrument manual.