7 PERIPHERAL NERVE LOCATOR<br />
Battery operated;<br />
Facility for identification of peripheral nerves in conduction anaesthesia and for<br />
percutaneous stimulation for neuromuscular stimulation under GA;<br />
LCD display of current, impulse pattern, pulse width and pulse amplitude;<br />
TOF, Twitch, DBS & Tetanus modes of stimulation;<br />
Autoclavable biopolar handle for visualizing intant muscle response without applying<br />
electrodes;<br />
To be supplied with rechargeable battery with charge<br />
8 PHARMACEUTAL REFRIGERATOR<br />
Temperature maintenance 2-8 degree C<br />
- Capacity 360-500 L, double wall steel cabinet<br />
- Alarm on lower and upper temperature deviation<br />
- Audio alarm for low and high temperature<br />
- Digital temperature display<br />
- Preferably with castor wheels for smooth movement<br />
- Quality certified CE or ISO or equivalent preferred<br />
- Operation at 220V/ 50 Hz / single phase power<br />
9 TECHNICAL SPECIFICATIOS FOR AUTOMATIC POINT OF CARE HAEMOSTASIS ANALYSER<br />
ALONG WITH ACCESSORIES;<br />
1) The Equipment should work on electromagnetic principle and measure secondary<br />
oscillation of the wire induced by blood clot.<br />
2) The Equipment should be a modular system which is upgradable from 1 to 4<br />
analyzing channels with in the machine and should be able to run all the analyzing<br />
channels simultaneously .Analyzing channels should be integrated with in the same<br />
machine<br />
3) Complete procedure from loading and unloading of sample filled cups, heating of<br />
blood samples to mechanical synchronization with electronic components should be<br />
fully automatic<br />
4) It should come with prefixed torsion wire sensor<br />
5) It should have an integrated notebook with in the main equipment<br />
6)The equipment should have automatic electromagnetic brake system for protection<br />
of sensor when not in use<br />
7) The Machine should have software which provide automatic facility for<br />
repositioning the cup carrier/holder and for engaging pin with torsion wire.<br />
8) The following parameter should be analyzed by the machine and provide the<br />
different reference values of all parameters with and without the reagents:<br />
i) R-time<br />
ii) K-time<br />
iii) Alpha angle<br />
iv) Maximum amplitude<br />
v) LY 30/60<br />
vi) A10 and A 20<br />
9) Blood sample required for single test should be less than o.5ml<br />
10) Demonstration of the machine would be required<br />
11) Essential accessories:<br />
a) consumables for at least 500 samples<br />
b) Branded printer<br />
c) Branded U P S.<br />
12) It should have interface facility/ integration for data transfer from equipment to<br />
third party hospital management application.<br />