-Features.<br />

-pressure range from 0 – 20 mm Hg, CPAP, range 2 – 15 cm H2O.<br />

-Facility to adjust FiO2 of delivered Air.<br />

-Alarms:- High / Low pressure, apnea.<br />

-Digital display of - CPAP pressure in CPAP mode<br />

- Delivered pressure in IPPR mode.<br />

power 230 V AC, 50 Hz.<br />

-Servo controlled humidifier system with airway temperature probe.<br />

-Rechargeable and autoclave water chamber<br />

-Continuous use with Back-up batteries with 6 hr life.<br />

-Accessories with each machine:<br />

Soft reusable/disposable nasal cannulas / Reusable air<br />

Tubings with water traps / Reusable humidifying chambers/head<br />

cap & straps to fix nasal cannula snugly<br />

-Head Fixator 1 piece.<br />

-Test Lung 1 piece.<br />

- Nasal prongs! piece.<br />

Make - Indigenous<br />

4 Neonatal Ventilator<br />

-Should be microprocessor based, time-cycled suitable, pressure control for Neonatal and<br />

Pediatric patients.<br />

-Built in Air compressor and heated humidifier. Air compressor should be from the same<br />

principle.<br />

-Reusable circuit with online bacterial filter.<br />

-Should be able to provide ventilation in CMV, ACMV, IMV, CPAP and SIMV modes with<br />

apnea backup.<br />

-Should be able to run on compressor as well as on compressed air supply, the inlets<br />

should be compatible with oxygen points of the hospital.<br />

-Should display digitally both set as well as monitored parameters.<br />

-Frequency of ventilation, FiO2, PIP, PEEP, MAP<br />

-Derived parameters:- I;E ratio, TV, Minute ventilation.<br />

-The unit should be able to provide the following parameters:-<br />

- Tidal volume:- Wide range upto 1500 ml<br />

- Respiratory rate:- 0 – 150 breaths per minute<br />

- Pressure support:- 0 to 50 CM, H2O or equiv. mbar<br />

- PEEP:- 0 to 25 cm water or equivalent mbar<br />

- Fio2:- 21% to 100%<br />

- Inspiratory time:- 0.1 – 3 secs<br />

- Gas Flow:- 5 – 15 Lpm<br />

-Should have audio visual alarm for the following:-<br />

- Low or high pressure (PIP)<br />

- Low PEEP / CPAP<br />

- Patient disconnection / tube obstruction<br />

- Power / compressor failure<br />

- Insufficient Exp. Time<br />

-Humidifier:-<br />

- Flow resistance upto 1 cm H2 O/1/sec<br />

- Temperature range 31 – 40 0 C<br />

- Temperature control + 2 0 C<br />

- Digital display of temperature range<br />

- Water level indicator<br />

- Warm up time less than 15 mins.<br />

- Alarms<br />

Heater wire on:<br />

- Airway temp tracking + 2 0 C from set temp.<br />

- Chamber temp:- If chamber temp varies + 4 0 C from set temp.<br />

Heater wire off:<br />

Chamber temp limited to 60 max.<br />

Airway temp fixed at 41 C high & 29.5 0 C Low<br />

-Power 220 – 240 V, 50-6-Hz<br />

-Should have a battery backup for at least 60 minutes.<br />

-Spares with each ventilator:- Reusable circuit with Y piece, Heater wire, temperature<br />

probe and adapter<br />

34<br />