-Features.<br />
-pressure range from 0 – 20 mm Hg, CPAP, range 2 – 15 cm H2O.<br />
-Facility to adjust FiO2 of delivered Air.<br />
-Alarms:- High / Low pressure, apnea.<br />
-Digital display of - CPAP pressure in CPAP mode<br />
- Delivered pressure in IPPR mode.<br />
power 230 V AC, 50 Hz.<br />
-Servo controlled humidifier system with airway temperature probe.<br />
-Rechargeable and autoclave water chamber<br />
-Continuous use with Back-up batteries with 6 hr life.<br />
-Accessories with each machine:<br />
Soft reusable/disposable nasal cannulas / Reusable air<br />
Tubings with water traps / Reusable humidifying chambers/head<br />
cap & straps to fix nasal cannula snugly<br />
-Head Fixator 1 piece.<br />
-Test Lung 1 piece.<br />
- Nasal prongs! piece.<br />
Make - Indigenous<br />
4 Neonatal Ventilator<br />
-Should be microprocessor based, time-cycled suitable, pressure control for Neonatal and<br />
Pediatric patients.<br />
-Built in Air compressor and heated humidifier. Air compressor should be from the same<br />
principle.<br />
-Reusable circuit with online bacterial filter.<br />
-Should be able to provide ventilation in CMV, ACMV, IMV, CPAP and SIMV modes with<br />
apnea backup.<br />
-Should be able to run on compressor as well as on compressed air supply, the inlets<br />
should be compatible with oxygen points of the hospital.<br />
-Should display digitally both set as well as monitored parameters.<br />
-Frequency of ventilation, FiO2, PIP, PEEP, MAP<br />
-Derived parameters:- I;E ratio, TV, Minute ventilation.<br />
-The unit should be able to provide the following parameters:-<br />
- Tidal volume:- Wide range upto 1500 ml<br />
- Respiratory rate:- 0 – 150 breaths per minute<br />
- Pressure support:- 0 to 50 CM, H2O or equiv. mbar<br />
- PEEP:- 0 to 25 cm water or equivalent mbar<br />
- Fio2:- 21% to 100%<br />
- Inspiratory time:- 0.1 – 3 secs<br />
- Gas Flow:- 5 – 15 Lpm<br />
-Should have audio visual alarm for the following:-<br />
- Low or high pressure (PIP)<br />
- Low PEEP / CPAP<br />
- Patient disconnection / tube obstruction<br />
- Power / compressor failure<br />
- Insufficient Exp. Time<br />
-Humidifier:-<br />
- Flow resistance upto 1 cm H2 O/1/sec<br />
- Temperature range 31 – 40 0 C<br />
- Temperature control + 2 0 C<br />
- Digital display of temperature range<br />
- Water level indicator<br />
- Warm up time less than 15 mins.<br />
- Alarms<br />
Heater wire on:<br />
- Airway temp tracking + 2 0 C from set temp.<br />
- Chamber temp:- If chamber temp varies + 4 0 C from set temp.<br />
Heater wire off:<br />
Chamber temp limited to 60 max.<br />
Airway temp fixed at 41 C high & 29.5 0 C Low<br />
-Power 220 – 240 V, 50-6-Hz<br />
-Should have a battery backup for at least 60 minutes.<br />
-Spares with each ventilator:- Reusable circuit with Y piece, Heater wire, temperature<br />
probe and adapter<br />