etween blade and handle<br />

6 Paediatric Fiberoptic Laryngoscope<br />

1. Miller blade<br />

2. Sizes of Neonate<br />

3. Handle made of SS with bulb in built in handle<br />

4. Xeon lamp to prevent tissue burns<br />

5. Xeon lamp mounted on handle.<br />

D-2 Ortho<br />

1 C-ARM MOBILE IMAGE INT<strong>ESI</strong>FIER<br />

1. Mechanical C-arm<br />

Fully counterbalanced C-arm having:<br />

Orbital movement : + 90 degrees / -35 degrees<br />

Angulations : + - 190 degress<br />

Vertical : 400 mm motorized<br />

Horizontal : 200 mm<br />

Vertical movement : Motorized<br />

Horizontal movement : Manual<br />

Lateral Movement : Steering wheel<br />

Swivel Range : +/ - 12.5 degrees<br />

Source to 1.1 distance : 880 mm<br />

Depth of immersion : 630 mm<br />

Clear space : 680 mm<br />

2. T.V. System<br />

- Immage Intensifier should be of size 9” with zoom facility.<br />

- System should have two 15” Monitors with work station.<br />

- Grid : 8:1, 40 lines / cm (Stationary)<br />

- Disk storage up to 100 images.<br />

- Digital zoom.<br />

- image annotation.<br />

- Image reversal.<br />

- Live & stored image rotation.<br />

- Reference measurement<br />

- Patient directory<br />

- CD Writing facility.<br />

3. X-Ray Parameters:<br />

High frequency X-ray generator 2.5 KW<br />

The X-ray generator should have facility of Radiography & fluoroscopy.<br />

The X-ray tube should have a dual focus stationary anode X-ray tube 0.6 mm &<br />

1.5 mm for fluoroscopy and Radiography with Anode heat storage capacity<br />

min 35 KJoules.<br />

The equipment should have:<br />

Fluroscopy KV range : 40-110 KV<br />

MA rabge : 5 mA & 7 mA<br />

The equipment should have:<br />

Radiography KV range : 40-100 KV<br />

MA range : upt to 63 mA<br />

MAs : Upto 200 mAs<br />

Collimation : Fixed slot Collimator.<br />

4. Power requirement: 160 V to 260 Volt single phase without any external<br />

transformer:<br />

5. Accessories<br />

i Sterile cover<br />

40<br />

2 1000<br />

1 30000