etween blade and handle<br />
6 Paediatric Fiberoptic Laryngoscope<br />
1. Miller blade<br />
2. Sizes of Neonate<br />
3. Handle made of SS with bulb in built in handle<br />
4. Xeon lamp to prevent tissue burns<br />
5. Xeon lamp mounted on handle.<br />
D-2 Ortho<br />
1 C-ARM MOBILE IMAGE INT<strong>ESI</strong>FIER<br />
1. Mechanical C-arm<br />
Fully counterbalanced C-arm having:<br />
Orbital movement : + 90 degrees / -35 degrees<br />
Angulations : + - 190 degress<br />
Vertical : 400 mm motorized<br />
Horizontal : 200 mm<br />
Vertical movement : Motorized<br />
Horizontal movement : Manual<br />
Lateral Movement : Steering wheel<br />
Swivel Range : +/ - 12.5 degrees<br />
Source to 1.1 distance : 880 mm<br />
Depth of immersion : 630 mm<br />
Clear space : 680 mm<br />
2. T.V. System<br />
- Immage Intensifier should be of size 9” with zoom facility.<br />
- System should have two 15” Monitors with work station.<br />
- Grid : 8:1, 40 lines / cm (Stationary)<br />
- Disk storage up to 100 images.<br />
- Digital zoom.<br />
- image annotation.<br />
- Image reversal.<br />
- Live & stored image rotation.<br />
- Reference measurement<br />
- Patient directory<br />
- CD Writing facility.<br />
3. X-Ray Parameters:<br />
High frequency X-ray generator 2.5 KW<br />
The X-ray generator should have facility of Radiography & fluoroscopy.<br />
The X-ray tube should have a dual focus stationary anode X-ray tube 0.6 mm &<br />
1.5 mm for fluoroscopy and Radiography with Anode heat storage capacity<br />
min 35 KJoules.<br />
The equipment should have:<br />
Fluroscopy KV range : 40-110 KV<br />
MA rabge : 5 mA & 7 mA<br />
The equipment should have:<br />
Radiography KV range : 40-100 KV<br />
MA range : upt to 63 mA<br />
MAs : Upto 200 mAs<br />
Collimation : Fixed slot Collimator.<br />
4. Power requirement: 160 V to 260 Volt single phase without any external<br />
transformer:<br />
5. Accessories<br />
i Sterile cover<br />
40<br />
2 1000<br />
1 30000