4) Preferably should have Automatic tube compensation inspiratory as well<br />
as in expiratory cycle.<br />
5) Should be upgradable for independent lung ventilation.<br />
6) Should have apnoea back-up.<br />
7) Preferably should have automatic adjustment of pressure and flow (Auto<br />
Flow) within a set PIP<br />
8) Should have Automatic compliance and leakage compensation for circuit<br />
and ET tube.<br />
Should have the following settings;<br />
Tidal Volume 20ml to 2000 ml<br />
Respiratory Rate 5 – 10 breaths per minute<br />
Peak Inspiratory Pressure 0 – 80 cmH2O<br />
CPAP/PEEP 0 – 35 cmH2O<br />
Pressure support 0 – 80 cmH2O<br />
FiO2 21 – 100%<br />
Inspiratory flow 0 – 120 litres per minutes<br />
SIMV Respiratory Rate 1 – 40 breaths per minute<br />
ß Should have monitoring and display of the following parameters:<br />
1. Airway pressure – Peak, Plateau, Mean, CPAP/PEEP<br />
2. Tidal Volume – Inspired and expired<br />
3. Minute Volume – Inspired and expired and spontaneous<br />
4. Respiratory Rate – Inspirtory, I:E Ratio<br />
5. FiO2<br />
6. Lung Mechanics – Resistance, Compliance<br />
7. Lung mechanics indicators – NIF, RSB,<br />
8. Occlusion pressure and Intrinsic (Auto) PEEP<br />
ß Should have built in alarms for the following:<br />
1. High/low Airway pressure<br />
2. High/low Minute Volume<br />
3. High Respiratory Rate<br />
4. High Tidal Volume<br />
5. Apnoea / apnoea alarm time<br />
6. High/low O2 %<br />
7. Power Failure<br />
8. Compressed air failure<br />
9. End tidal CO2 (Optional)<br />
Should have the real time waveform/loops for the following:<br />
A Waveforms : Pressure, Flow & Volume<br />
B Loops : Pressure Vs Flow, pressure Vs Volume, Volume Vs Flow<br />
C Trends User defined 24 hour graphical trends<br />
It should be supplied with following accessories to be supplied by the same<br />
manufacturer-<br />
A- Humidifier:- Should be servo controlled thermo compensated Airway.<br />
temperature to be controlled, measured and displayed<br />
ßTemperature probe should be provided for connect ion at Y piece of patient hose<br />
system<br />
ßHumidifier clamp / bracket also should be provided.<br />
ßOne additional humidifier chamber should also be provided.<br />
B- Non-corrosive trolley and hinged arm<br />
C- Reusable Adult and Pediatric breathing circuits :- Steam Sterilisable, made of<br />
silicon, Adult Six & Pediatric Two<br />
D- Disposable adult and paediatric breathing circuits compatible with humidifier,<br />
along with heater wire adapter – 50 Adult circuits and 25 Paediatrics circuits.<br />
E- HME filters – 50 Nos.<br />
F- Air & Oxygen hoses 1 each<br />
G- Orofacial masks for noninvasive ventilation – Autoclavable, in large, medium<br />
and small size<br />