4) Preferably should have Automatic tube compensation inspiratory as well<br />

as in expiratory cycle.<br />

5) Should be upgradable for independent lung ventilation.<br />

6) Should have apnoea back-up.<br />

7) Preferably should have automatic adjustment of pressure and flow (Auto<br />

Flow) within a set PIP<br />

8) Should have Automatic compliance and leakage compensation for circuit<br />

and ET tube.<br />

Should have the following settings;<br />

Tidal Volume 20ml to 2000 ml<br />

Respiratory Rate 5 – 10 breaths per minute<br />

Peak Inspiratory Pressure 0 – 80 cmH2O<br />

CPAP/PEEP 0 – 35 cmH2O<br />

Pressure support 0 – 80 cmH2O<br />

FiO2 21 – 100%<br />

Inspiratory flow 0 – 120 litres per minutes<br />

SIMV Respiratory Rate 1 – 40 breaths per minute<br />

ß Should have monitoring and display of the following parameters:<br />

1. Airway pressure – Peak, Plateau, Mean, CPAP/PEEP<br />

2. Tidal Volume – Inspired and expired<br />

3. Minute Volume – Inspired and expired and spontaneous<br />

4. Respiratory Rate – Inspirtory, I:E Ratio<br />

5. FiO2<br />

6. Lung Mechanics – Resistance, Compliance<br />

7. Lung mechanics indicators – NIF, RSB,<br />

8. Occlusion pressure and Intrinsic (Auto) PEEP<br />

ß Should have built in alarms for the following:<br />

1. High/low Airway pressure<br />

2. High/low Minute Volume<br />

3. High Respiratory Rate<br />

4. High Tidal Volume<br />

5. Apnoea / apnoea alarm time<br />

6. High/low O2 %<br />

7. Power Failure<br />

8. Compressed air failure<br />

9. End tidal CO2 (Optional)<br />

Should have the real time waveform/loops for the following:<br />

A Waveforms : Pressure, Flow & Volume<br />

B Loops : Pressure Vs Flow, pressure Vs Volume, Volume Vs Flow<br />

C Trends User defined 24 hour graphical trends<br />

It should be supplied with following accessories to be supplied by the same<br />

manufacturer-<br />

A- Humidifier:- Should be servo controlled thermo compensated Airway.<br />

temperature to be controlled, measured and displayed<br />

ßTemperature probe should be provided for connect ion at Y piece of patient hose<br />

system<br />

ßHumidifier clamp / bracket also should be provided.<br />

ßOne additional humidifier chamber should also be provided.<br />

B- Non-corrosive trolley and hinged arm<br />

C- Reusable Adult and Pediatric breathing circuits :- Steam Sterilisable, made of<br />

silicon, Adult Six & Pediatric Two<br />

D- Disposable adult and paediatric breathing circuits compatible with humidifier,<br />

along with heater wire adapter – 50 Adult circuits and 25 Paediatrics circuits.<br />

E- HME filters – 50 Nos.<br />

F- Air & Oxygen hoses 1 each<br />

G- Orofacial masks for noninvasive ventilation – Autoclavable, in large, medium<br />

and small size<br />