Sep 7/04<br />
<strong>FutureSuite</strong> <strong>UPDATE</strong> <strong>NOTICE</strong> <strong>REV</strong> <strong>3.01</strong><br />
(12) The Shmoo plot was terminated when the following conditions were satisfied:<br />
Three-dimensional Normal Shmoo was used<br />
[Slice] was set as the display type.<br />
The following procedures were used to operate Shmoo.<br />
Set the value that was displayed in the axis to be sliced.<br />
4. IMPROVED ITEMS<br />
Read the Shmoo data file, in which scan conditions of the number of steps which was<br />
smaller than the set value.<br />
Relevant software revisions: <strong>REV</strong> 3.00 through <strong>REV</strong>3.00-6<br />
Improvement The Shmoo data file has been read correctly.<br />
(13) The unit displayed on Index of the Scan Condition dialog box was "V" when the following<br />
condition was satisfied:<br />
ILH or ILL was specified as a unit condition in the Scan Condition dialog box.<br />
Relevant software revisions: <strong>REV</strong> 3.00 through <strong>REV</strong>3.00-6<br />
Improvement The unit displayed on Index has been "A".<br />
(14) The display position of the original point marker of the Z-axis was wrong when the following<br />
conditions were satisfied:<br />
The file saved with the Shmoo plot tool of Rev1.xx was read.<br />
The data of this file was 3D Normal Shmoo.<br />
The display type was [Slice].<br />
The target axis to slice was not the Z-axis.<br />
Relevant software revisions: <strong>REV</strong> 3.00 through <strong>REV</strong>3.00-6<br />
Improvement The original point marker of the Z-axis has been displayed on the correct<br />
position.<br />
(15) The timing set number of RATE in the original point data of the outputted text was not enclosed<br />
in parenthesis"" when the following conditions were satisfied:<br />
RATE was set as the scanning conditions of an axis.<br />
The Shmoo data was outputted to a text.<br />
Relevant software revisions: <strong>REV</strong> 3.00 through <strong>REV</strong>3.00-6<br />
Improvement The timing set number of RATE in the original point data of the outputted<br />
text bas been enclosed in parenthesis"".<br />