Test Interpretation - QC/Patient Results<br />

A. Test Interpretation<br />

Test Interpretation - QC/Patient Results<br />

<strong>Assay</strong> test results are automatically interpreted by the TIGRIS DTS System <strong>APTIMA</strong> <strong>Trichomonas</strong><br />

<strong>Assay</strong> software. A test result may be negative, positive, or invalid as determined by total RLU in<br />

the detection step (see below). A test result may be invalid due to RLU values outside the normal<br />

expected ranges. Initial invalid test results should be retested. Report the first valid result.<br />

Test Interpretation Total RLU (x1000)<br />

Negative 0* to < 100<br />

Positive 100 to < 2400<br />

Invalid 0* or ≥ 2400<br />

*If the RLU measured on the TIGRIS DTS System is between 0 and<br />

999, a result of “0” is reported in the “Total RLU (000s)” column in the<br />

run report. Measured RLU values less than 690 are reported as invalid.<br />

RLU values between 690 and 999 are reported as valid.<br />

B. Quality Control Results and Acceptability<br />

The <strong>APTIMA</strong> Negative Control for <strong>Trichomonas</strong>, which is labeled “NC CONTROL – TRICH,” and<br />

the <strong>APTIMA</strong> Positive Control for <strong>Trichomonas</strong>, which is labeled “PC CONTROL + TRICH,” act as<br />

controls for the target capture, amplification, and detection steps of the assay. In accordance with<br />

guidelines or requirements of national, regional, and/or local regulations or accrediting<br />

organizations, additional controls for cell lysis and RNA stabilization may be included. The<br />

<strong>APTIMA</strong> Positive Control for <strong>Trichomonas</strong> which is labeled “PC CONTROL + TRICH” contains<br />

non-infectious T. <strong>vaginalis</strong> rRNA.<br />

The <strong>APTIMA</strong> <strong>Trichomonas</strong> <strong>vaginalis</strong> Controls must produce the following test results:<br />

Control Total RLU (x1000) T. <strong>vaginalis</strong> Result<br />

NC Control – TRICH 0* and < 20 Negative<br />

PC Control + TRICH ≥ 500 and < 2400 Positive<br />

*If the RLU measured on the TIGRIS DTS System is between 0 and 999, a result of “0” is<br />

reported in the “Total RLU (000s)” column in the run report. Measured RLU values less than<br />

690 are reported as invalid. RLU values between 690 and 999 are reported as valid.<br />

Each laboratory should implement appropriate control procedures to satisfy the requirements of<br />

CLIA regulations (section 493.1256).<br />

Note: For assistance with out-of-range controls, contact <strong>Gen</strong>-<strong>Probe</strong> Technical Support.<br />

<strong>APTIMA</strong> <strong>Trichomonas</strong> <strong>vaginalis</strong> <strong>Assay</strong> 16 502246 Rev. B