Test Interpretation - QC/Patient Results<br />
A. Test Interpretation<br />
Test Interpretation - QC/Patient Results<br />
<strong>Assay</strong> test results are automatically interpreted by the TIGRIS DTS System <strong>APTIMA</strong> <strong>Trichomonas</strong><br />
<strong>Assay</strong> software. A test result may be negative, positive, or invalid as determined by total RLU in<br />
the detection step (see below). A test result may be invalid due to RLU values outside the normal<br />
expected ranges. Initial invalid test results should be retested. Report the first valid result.<br />
Test Interpretation Total RLU (x1000)<br />
Negative 0* to < 100<br />
Positive 100 to < 2400<br />
Invalid 0* or ≥ 2400<br />
*If the RLU measured on the TIGRIS DTS System is between 0 and<br />
999, a result of “0” is reported in the “Total RLU (000s)” column in the<br />
run report. Measured RLU values less than 690 are reported as invalid.<br />
RLU values between 690 and 999 are reported as valid.<br />
B. Quality Control Results and Acceptability<br />
The <strong>APTIMA</strong> Negative Control for <strong>Trichomonas</strong>, which is labeled “NC CONTROL – TRICH,” and<br />
the <strong>APTIMA</strong> Positive Control for <strong>Trichomonas</strong>, which is labeled “PC CONTROL + TRICH,” act as<br />
controls for the target capture, amplification, and detection steps of the assay. In accordance with<br />
guidelines or requirements of national, regional, and/or local regulations or accrediting<br />
organizations, additional controls for cell lysis and RNA stabilization may be included. The<br />
<strong>APTIMA</strong> Positive Control for <strong>Trichomonas</strong> which is labeled “PC CONTROL + TRICH” contains<br />
non-infectious T. <strong>vaginalis</strong> rRNA.<br />
The <strong>APTIMA</strong> <strong>Trichomonas</strong> <strong>vaginalis</strong> Controls must produce the following test results:<br />
Control Total RLU (x1000) T. <strong>vaginalis</strong> Result<br />
NC Control – TRICH 0* and < 20 Negative<br />
PC Control + TRICH ≥ 500 and < 2400 Positive<br />
*If the RLU measured on the TIGRIS DTS System is between 0 and 999, a result of “0” is<br />
reported in the “Total RLU (000s)” column in the run report. Measured RLU values less than<br />
690 are reported as invalid. RLU values between 690 and 999 are reported as valid.<br />
Each laboratory should implement appropriate control procedures to satisfy the requirements of<br />
CLIA regulations (section 493.1256).<br />
Note: For assistance with out-of-range controls, contact <strong>Gen</strong>-<strong>Probe</strong> Technical Support.<br />
<strong>APTIMA</strong> <strong>Trichomonas</strong> <strong>vaginalis</strong> <strong>Assay</strong> 16 502246 Rev. B