APTIMA Trichomonas vaginalis Assay - Gen-Probe, Inc.
APTIMA Trichomonas vaginalis Assay - Gen-Probe, Inc.
APTIMA Trichomonas vaginalis Assay - Gen-Probe, Inc.
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Limitations<br />
P. The performance of the TIGRIS DTS System has not been determined at altitudes above<br />
2240 m (7355 feet). Additional volumetric verifications and assay specific studies will be<br />
performed prior to, or as part of, the installation and acceptance process in laboratories<br />
above 2240 m (7355 feet) altitude.<br />
Q. If a specimen has a small number of T. <strong>vaginalis</strong> organisms, uneven distribution of these<br />
trichomonads may occur, which may affect the ability to detect T. <strong>vaginalis</strong> rRNA in the<br />
collected material. If negative results from the specimen do not fit with the clinical impression,<br />
a new specimen may be necessary.<br />
R. Customers must independently validate an LIS transfer process.<br />
<strong>APTIMA</strong> <strong>Trichomonas</strong> <strong>vaginalis</strong> <strong>Assay</strong> 18 502246 Rev. B