01.06.2013 Views

Development of a Oxygen Sensor for Marine ... - DTU Nanotech

Development of a Oxygen Sensor for Marine ... - DTU Nanotech

Development of a Oxygen Sensor for Marine ... - DTU Nanotech

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Development</strong> <strong>of</strong> a <strong>Oxygen</strong><br />

<strong>Sensor</strong> <strong>for</strong> <strong>Marine</strong> Environment<br />

Studies<br />

M.Sc. Thesis<br />

Peter Kastberg Hansen s958291<br />

Supervisor: Erik V Thomsen<br />

September 20, 2005


ii<br />

MIC - Department <strong>of</strong> Micro and <strong>Nanotech</strong>nology<br />

Technical University <strong>of</strong> Denmark<br />

<strong>DTU</strong> Building 345east<br />

DK-2800 Kgs. Lyngby


Abstract<br />

This thesis described the first steps towards creating a oxygen sensor <strong>for</strong><br />

measuring dissolved oxygen in the oceans, with the intended use as part <strong>of</strong><br />

a larger project concerning fish behavior studies. The sensor is a Clark type<br />

oxygen sensor, with an thermal platinum sensor added as well.<br />

The first part describes the reasons <strong>for</strong> choosing the Clark sensor and the<br />

theory behind.<br />

The second part describes the design considerations and the fabrication steps<br />

that leads to the layout.<br />

While the final part deals with problems encountered, solutions to them,<br />

results gathered, and ways <strong>of</strong> improving the design.<br />

i


Acknowledgements<br />

I would like to say thanks the following people <strong>for</strong> their help, inspiration and<br />

patience; Erik V. Thomsen (my supervisor), Anders Hyldg˚ard (effectively the<br />

co-supervisor), and the rest <strong>of</strong> the Applied <strong>Sensor</strong>s group at MIC. Also thanks<br />

to Oliver Geschke <strong>for</strong> giving some valuable tips and insight on the workings<br />

<strong>of</strong> the Clark sensor early on. I would also like to thank the Laboratory<br />

Technicians <strong>for</strong> help in the Clean room, and the other master students in<br />

room 119 at MIC <strong>for</strong> a fun and spirited atmosphere, and <strong>for</strong> giving a hand<br />

with a lot <strong>of</strong> small things.<br />

iii


Contents<br />

1 Introduction 1<br />

1.1 Dissolved <strong>Oxygen</strong> Level . . . . . . . . . . . . . . . . . . . . . 2<br />

1.2 Thesis Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . 4<br />

2 Optodes and ISFET 7<br />

2.1 Optodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7<br />

2.2 ISFET (Ion Selective Field Effect Transistor) . . . . . . . . . . 9<br />

2.3 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11<br />

3 Theory <strong>of</strong> the Clark <strong>Sensor</strong> 13<br />

3.1 Chemistry <strong>of</strong> the Clark <strong>Sensor</strong> . . . . . . . . . . . . . . . . . . 14<br />

3.1.1 One layer electrode model . . . . . . . . . . . . . . . . 16<br />

3.1.2 Two layer model . . . . . . . . . . . . . . . . . . . . . 19<br />

3.2 The Potentiostat . . . . . . . . . . . . . . . . . . . . . . . . . 22<br />

3.3 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26<br />

4 System Design 27<br />

4.1 General Design . . . . . . . . . . . . . . . . . . . . . . . . . . 27<br />

4.2 Actual Design . . . . . . . . . . . . . . . . . . . . . . . . . . . 28<br />

4.2.1 Minor Design Variations . . . . . . . . . . . . . . . . . 32<br />

4.3 Temperature <strong>Sensor</strong> . . . . . . . . . . . . . . . . . . . . . . . . 33<br />

v


vi CONTENTS<br />

4.4 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36<br />

4.5 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40<br />

5 Fabrication 41<br />

5.1 Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41<br />

5.2 Back-end Processing . . . . . . . . . . . . . . . . . . . . . . . 45<br />

5.2.1 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . 45<br />

5.2.2 Ag to Ag/AgCl . . . . . . . . . . . . . . . . . . . . . . 47<br />

5.2.3 Electrolyte and Membrane . . . . . . . . . . . . . . . . 47<br />

5.3 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48<br />

6 Problems and Solutions 49<br />

6.1 Wirebonding . . . . . . . . . . . . . . . . . . . . . . . . . . . 49<br />

6.1.1 Conducting Glue . . . . . . . . . . . . . . . . . . . . . 50<br />

6.2 The flex print . . . . . . . . . . . . . . . . . . . . . . . . . . . 53<br />

6.2.1 Non Conducting Glue . . . . . . . . . . . . . . . . . . 54<br />

6.3 The ’O’-ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54<br />

6.3.1 Changing the resting area . . . . . . . . . . . . . . . . 55<br />

6.4 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56<br />

7 Evaluation <strong>of</strong> Results and Measurements 57<br />

7.1 Fabrication results . . . . . . . . . . . . . . . . . . . . . . . . 57<br />

7.2 Temperature <strong>Sensor</strong> . . . . . . . . . . . . . . . . . . . . . . . . 58<br />

7.3 Clark <strong>Sensor</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60<br />

7.4 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61<br />

8 Conclusion 63<br />

A Silicide Recipe 69


CONTENTS vii<br />

B Fabrication Process 71<br />

C Process Sequence 75


viii CONTENTS


Chapter 1<br />

Introduction<br />

Roughly 70 % <strong>of</strong> the world is covered by water[1], and have been an inspiration<br />

<strong>for</strong> mankind over the ages, from explorers such as Columbus to works <strong>of</strong><br />

fiction, such as Hemingway’s ’The Old Man and the Sea’. Several time the<br />

oceans have shown the tremendous <strong>for</strong>ces that rest within in various <strong>for</strong>ms<br />

<strong>of</strong> natural disasters such as tidal waves, floods, tsunamis and many other.<br />

However the oceans is also a source <strong>of</strong> life, in fact the original source <strong>of</strong> life,<br />

not just <strong>for</strong> us but <strong>for</strong> the fish that lives below.<br />

This thesis have been subpart <strong>of</strong> a larger project (Fish and Chips), where<br />

the main goal is the development <strong>of</strong> a multi-sensor <strong>for</strong> marine environmental<br />

studies. In that, the prototype <strong>for</strong> a Micro Electronic Mechanical System<br />

(MEMS) have been developed, with then intended use <strong>for</strong> fish tagging. While<br />

this thesis does not have quite as wide a scope, it does focus on one <strong>of</strong> the<br />

more important aspects <strong>of</strong> it, namely oxygen, or rather dissolved oxygen.<br />

There are commercially available dissolved oxygen sensors already today,<br />

however these are more <strong>of</strong>ten than not macro scale sensors attached to a<br />

buoy, and there<strong>for</strong>e only measuring in one place. They can also be attached<br />

to ships, however they will only measure the oxygen level at the surface, and<br />

there<strong>for</strong>e not give a full accurate picture <strong>of</strong> the oxygen level.<br />

Why dissolved oxygen? Well there are many reasons, mainly like us the<br />

fish needs to breathe, and the gills <strong>of</strong> a fish are not all that different <strong>for</strong>m<br />

our lungs[2]. Basically there needs to be enough dissolved oxygen in the<br />

water to favor diffusion <strong>of</strong> oxygen from the water, through the gills <strong>of</strong> the<br />

fish and into the blood. When the oxygen falls below a certain level this<br />

passive diffusion fails to be able to drive, the oxygen into the blood and the<br />

fish suffocates much like a human might at high altitudes. So by monitoring<br />

1


2 CHAPTER 1. INTRODUCTION<br />

the oxygen level it is possible to learn more about the behavior and health<br />

<strong>of</strong> the fish. Also given the relationship between dissolved oxygen levels and<br />

temperature(more about this later), it provides an idea <strong>of</strong> where the fish have<br />

been.[3, 4]<br />

Figure 1.1: Sketch <strong>of</strong> a fish’s gills, with oxygen leaving the water, and entering<br />

the bloodstream, and then subsequently moved around to<br />

the rest <strong>of</strong> the fish’s body. Reproduced from [5]<br />

Furthermore given the amount <strong>of</strong> pollution, algae growth and other problems<br />

in the world oceans today, it supplies additional info on the general<br />

health state <strong>of</strong> the oceans. The use <strong>of</strong> the sensor doesn’t need be limited<br />

to the oceans however, as it in theory should also be useful <strong>for</strong> on site measurement<br />

<strong>of</strong> the dissolved oxygen levels in rivers, creeks and lakes. Not to<br />

mention the medical uses involving the oxygen levels in blood. Also in these<br />

cases having a small easy to use sensor is preferable over a macro scale one.[6]<br />

1.1 Dissolved <strong>Oxygen</strong> Level<br />

How much dissolved oxygen is there in seawater? Under the assumption that<br />

the only source <strong>for</strong> the dissolved oxygen in water is the atmosphere (two<br />

other sources being by rapid movement(aeration), and as a waste product<br />

<strong>of</strong> photosynthesis. Both are natural processes, which however is affected by<br />

pollution.), the relationship between air and water can be represented with<br />

the equation:<br />

This has the equilibrium constant<br />

O2(g) ⇔ O2(aq) (1.1)


1.1. DISSOLVED OXYGEN LEVEL 3<br />

K = aO2<br />

fO2<br />

(1.2)<br />

Where aO2 is the activity <strong>of</strong> oxygen and fO2 the fugacity 1 <strong>of</strong> oxygen in<br />

the gas phase.<br />

However when the fugacity <strong>of</strong> oxygen is approximated by its partial pressure<br />

PO2 and the activity <strong>of</strong> oxygen in water by its concentration CO2(aq),<br />

Henry’s law 2 will be expressed as:<br />

CH = CO2(aq)<br />

PO2<br />

(1.3)<br />

Where CH is the Henry’s law constant, CH(O2) = 1.3 × 10 −3 Mol/atm.<br />

PO2 can be calculated by multiplying the fraction volume <strong>of</strong> oxygen in the<br />

air (21 %) with the pressure at sea level ( 760 mmHg, atmospheric pressure).<br />

Also in dilute solutions and <strong>for</strong> perfect gases K = CH, hence the<br />

concentration <strong>of</strong> oxygen in the ocean is<br />

C = CHPO2n = 1.3 × 10 −3 · 0.21 · 760 · 32 = 6.6mg/L (1.4)<br />

6.6 mg/L might not tell much on its own, however is the level <strong>of</strong> dissolved<br />

oxygen water drops below 5.0 mg/L, fish and other aquatic life will start to<br />

suffocate. The lower it drops the worse it gets, once the oxygen levels drop<br />

below 1-2 mg/L it will only take a few hours be<strong>for</strong>e most or all the fish are<br />

dead.[8]<br />

Now as mentioned earlier there is also a connection between the concentration<br />

<strong>of</strong> dissolved oxygen and temperature, or rather between water<br />

temperature and the saturation <strong>of</strong> oxygen. This is the point where water<br />

have the maximum amount <strong>of</strong> oxygen it can have, relative to pressure and<br />

temperature. It is calculated from the following equation[9]:<br />

Cp = C ∗ <br />

PW V (1 − )(1 − ΘP )<br />

P P<br />

(1 − PW V )(1 − Θ)<br />

(1.5)<br />

1 The measure <strong>of</strong> the tendency <strong>of</strong> a gas to escape or expand[7]<br />

2 The mass <strong>of</strong> a gas that dissolves in a definite volume <strong>of</strong> liquid is directly proportional<br />

to the pressure <strong>of</strong> the gas provided the gas does not react with the solvent.


4 CHAPTER 1. INTRODUCTION<br />

Where<br />

CP = equilibrium oxygen concentration at nonstandard pressure. mg/L.<br />

C ∗ = equilibrium oxygen concentration at 1 atm, mg/L.<br />

P = nonstandard pressure, atm.<br />

PW V = partial pressure <strong>of</strong> water vapor, atm, calculated from<br />

lnPW V = 11.8571 − 3840.70<br />

T<br />

T = temperature, Kelvin<br />

t = temperature, o C.<br />

− 216.961<br />

T 2<br />

Θ = 0.000975 − 1.426 × 10 −5 t + 6.436 × 10 −8 t 2<br />

(1.6)<br />

(1.7)<br />

As can be seen from the equation, the warmer it gets, the less oxygen<br />

there is the water, as it becomes saturated faster.<br />

Summed up all this gives an idea <strong>of</strong> how high, or low depending on how<br />

you look at it, the measured values can be expected to be.<br />

1.2 Thesis Outline<br />

In Chapter 2 other ways <strong>of</strong> using micro technology to measure oxygen is<br />

looked at, and the reason <strong>for</strong> picking the Clark type sensor.<br />

In Chapter 3 the theory behind the Clark <strong>Oxygen</strong> sensor, amperometry,<br />

and the potentionstat is explained.<br />

In Chapter 4 the reasons <strong>for</strong> why the sensor ended up looking as it did<br />

is explained.<br />

In Chapter 5 contains a description <strong>of</strong> the process to create the sensor.<br />

In Chapter 6 the problems encountered while fabricating the sensor is discussed,<br />

as well as how they were solved.<br />

In Chapter 7 various measurements, tests, and results are presented.


1.2. THESIS OUTLINE 5<br />

In Chapter 8 the conclusion <strong>for</strong> the whole project is made.


6 CHAPTER 1. INTRODUCTION


Chapter 2<br />

Optodes and ISFET<br />

There are several ways to measure oxygen, and the concept <strong>of</strong> doing it is<br />

not entirely new, and have been done <strong>for</strong> several decades, in oceanography,<br />

medicine, mountaineering as well as in many other areas. However while<br />

several ways exists on the macro scale, there are not quite as many ways on<br />

the micro scale. In this chapter the pros and cons <strong>of</strong> two <strong>of</strong> the ways are<br />

described, the theory behind them is briefly described, though without going<br />

deeply into the details. A comparison to the Clark sensor is also made (the<br />

theory <strong>of</strong> Clark sensor is described in Chapter 3).<br />

2.1 Optodes<br />

Figure 2.1: Commercial oxygen optode from Aanderaa, actual size <strong>of</strong> the<br />

sensor is 36 × 86mm. Reproduced from [10]<br />

Traditional oxygen sensors, such as the Clark sensor described in Chapter<br />

3 are based on the reduction at a working electrode, which is immersed a<br />

electrolyte and then separated from other interfering gasses by a membrane<br />

7


8 CHAPTER 2. OPTODES AND ISFET<br />

only permeable to oxygen. The oxygen flow through the membrane being<br />

induced by the consumption <strong>of</strong> O2 at this electrode. With the electrical flow<br />

between the electrodes being proportional to this flow <strong>of</strong> oxygen. Problem<br />

is that this flow is highly dependant on the diffusion though the membrane,<br />

and in seawater these can change due to temperature and biological fouling.<br />

Especially the biological fouling can be a problem, not only because they<br />

can cause a change in the diffusion characteristics <strong>of</strong> the membrane. But<br />

also because biological processes quite <strong>of</strong>ten involves the production or the<br />

consumption <strong>of</strong> oxygen, and as such will influence the amount measured.[11]<br />

Optodes (also know as optoelectrodes or optrodes[12]) are the optical equivalent<br />

<strong>of</strong> electrodes, and represent a theoretical solution to this. These are<br />

usually based on the quenching <strong>of</strong> a fluorescent indicator by oxygen, and<br />

described by the Stern-Volmer equation[13]:<br />

I0<br />

I = 1 + KSV [O2] (2.1)<br />

Where I and I0 are the fluorescence intensities with and without oxygen<br />

present, KSV the Stern-Volmer quenching constant and [O2] the concentration<br />

<strong>of</strong> dissolved oxygen. The sensitivity depends on the value <strong>of</strong> KSV ,<br />

which determines the change in the fluorescent intensity with various oxygen<br />

concentrations.[14]<br />

Figure 2.2: Sketch <strong>of</strong> the optical system <strong>of</strong> the optode.[10]<br />

The way the optical sensor, as shown on figure 2.2, works is:<br />

1. The blue LED emits light, at 475 nm, to an optical fiber.


2.2. ISFET (ION SELECTIVE FIELD EFFECT TRANSISTOR) 9<br />

2. The optical fiber then carries the light to the probe. The end <strong>of</strong> the<br />

probe tip consists <strong>of</strong> a thin layer <strong>of</strong> a hydrophobic material, with ruthenium<br />

inside it, protected from the liquid measured on.<br />

3. The light from the LED excites the ruthenium at the probe tip.<br />

4. The excited ruthenium then fluoresces, emitting energy at 600 nm.<br />

5. If the excited ruthenium hits an oxygen molecule, the excess energy<br />

is transferred to the oxygen molecule in a non-radiative transfer, decreasing<br />

or quenching the fluorescence). The degree <strong>of</strong> quenching corresponds<br />

to the level <strong>of</strong> oxygen concentration or to oxygen partial pressure<br />

in the film, which is in dynamic equilibrium with oxygen in the liquid.<br />

6. The energy is collected by the probe and carried through the optical<br />

fiber to the data storage/PC. Which in turn translates the data into a<br />

graph or a similar output, from which the oxygen level can be read.<br />

Some <strong>of</strong> the advantages <strong>of</strong> the optodes is that unlike traditional electrodes<br />

the optodes doesn’t consume oxygen, and is there<strong>for</strong>e not sensitive to stirring.<br />

Also the partial pressure <strong>of</strong> the oxygen in the probe is in equilibrium<br />

with the partial pressure outside, there<strong>for</strong>e it is less effected by pressure and<br />

the pressure behavior becomes easier to predict. Finally some <strong>of</strong> the commercially<br />

available optodes comes with warranties with up to two years, which<br />

allows <strong>for</strong> long term measurements.[10]<br />

Disadvantages <strong>of</strong> the optodes however is that they tend to be relatively<br />

large (the optode itself is small, but the electronics and power supply adds<br />

considerably to the size), and is there<strong>for</strong> <strong>of</strong>ten attached to ships or buoys.<br />

It also have a relatively large power consumption in comparison to a micro<br />

Clark sensor.<br />

2.2 ISFET (Ion Selective Field Effect Transistor)<br />

Electrochemical sensors are one <strong>of</strong> the older and larger groups <strong>of</strong> chemical<br />

sensor, the sensor type includes those based on electrochemical and charge<br />

transfer reactions, meaning the transfer <strong>of</strong> charge from an electrode to a<br />

solid or liquid phase, or the other way around. The chemical process takes<br />

place at the electrodes or in the probed volume, and the current from this


10 CHAPTER 2. OPTODES AND ISFET<br />

is measured. A electrochemical sensor are always made up <strong>of</strong> a minimum<br />

<strong>of</strong> two electrodes, one that allows <strong>for</strong> connection through the probed sample<br />

and the other via measuring equipment or a transducer.<br />

There are various electrochemical sensors, which can be defined by their<br />

analytical principles[15]:<br />

1. Voltammetric sensors; which measures the current-voltage relationship.<br />

The principal behind it, is having a potential applied, and measuring<br />

a current proportional to the concentration <strong>of</strong> the electro-active substance<br />

(a variant <strong>of</strong> this is amperometry[16], wherein the potential is<br />

kept constant.).<br />

2. Conductometric sensors; Which measures the conductance, by applying<br />

a small amplitude AC potential to a pair <strong>of</strong> electrodes to stop<br />

polarization. The charge carries will then determine the conductance.<br />

3. Potentiometric sensors; which measures potential <strong>of</strong> an electrode at<br />

equilibrium state, meaning no current is allowed to flow at the time <strong>of</strong><br />

the measurement.<br />

While all three can use CMOS based technology, it is especially in the<br />

case <strong>of</strong> the later, potentiometric sensors, that ISFET are used.[17]<br />

Figure 2.3: (a) diagram <strong>of</strong> a MOSFET (b) diagram <strong>of</strong> a ISFET. The metal<br />

gate <strong>of</strong> the MOSFET is replaced by the metal <strong>of</strong> a reference electrode,<br />

and a liquid to make contact with the bare gate insulator.<br />

Reproduced from [18].<br />

The ISFET is a variant <strong>of</strong> a MOSFET (see figure 2.3), where the MOS-<br />

FET is used <strong>for</strong> gas sensing, the ISFET is used <strong>for</strong> liquids. The more traditional<br />

<strong>of</strong> these are the pH-ISFET, where the basic principle is the electrolysis


2.3. SUMMARY 11<br />

<strong>of</strong> oxygen molecules, which is similar to that <strong>of</strong> a Clark sensor (more about<br />

this in Chapter 3). However the way the ISFET detects it, is by determining<br />

the surface potential at the insulator/electrolyte interface.[19] Also in<br />

the case <strong>of</strong> the pH-ISFET the surface <strong>of</strong> the gate oxide will contain OH −<br />

molecules, which can be protonated and deprotonated, so when the gate oxide<br />

contacts an aqueous solution, the change in pH will cause the silicon<br />

surface potential to change as well.<br />

The advantages <strong>of</strong> the ISFET is that the CMOS and MOSFET are tried<br />

and used technologies. However chemical sensors are still a relative new area<br />

<strong>for</strong> this to be used in, and as such suffers from ’child’-diseases.<br />

Disadvantages have included light sensitivity, lack <strong>of</strong> solid-state reference<br />

electrodes, and packaging integrity difficulties. [20]<br />

2.3 Summary<br />

While both the Optodes and the ISFET are viable ways <strong>of</strong> creating a oxygen<br />

sensor, both had certain distinct disadvantages that made the Clark type a<br />

better choice. However <strong>of</strong> the two, the optodes might in the future become a<br />

superior solution, once the problems with size and power supply have been<br />

solved.


12 CHAPTER 2. OPTODES AND ISFET


Chapter 3<br />

Theory <strong>of</strong> the Clark <strong>Sensor</strong><br />

The Clark sensor was named after its creator Leland C. Clark[21], who in<br />

the late fifties and mid sixties studied the electrochemistry <strong>of</strong> oxygen gas<br />

reduction at platinum (Pt) metal electrodes. Clark began pioneering the<br />

use <strong>of</strong> what would later be an oxygen- (and there<strong>for</strong>e chemical-) sensor. In<br />

fact, Pt electrodes used to detect oxygen electrochemically are <strong>of</strong>ten referred<br />

to generically as ’Clark electrodes’. While the sensor he originally came up<br />

with was intended <strong>for</strong> the measurement <strong>of</strong> glucose concentration in blood,<br />

the sensor itself was the start <strong>of</strong> biosensor history, and the Clark sensor is<br />

still around today with many different shapes and several applications far<br />

from the original design.<br />

Liquid measured<br />

on (Blood, water,<br />

etc.)<br />

Membrane<br />

Ag Wire<br />

Pt<br />

KCl Solution (electrolyte)<br />

Figure 3.1: Sketch <strong>of</strong> a basic 2 electrode Clark <strong>Oxygen</strong> <strong>Sensor</strong>, with a Ag<br />

reference electrode (the anode) and a Pt working electrode (the<br />

cathode).<br />

The electrodes as seen in figure 3.1 have a thin organic membrane covering<br />

a layer <strong>of</strong> electrolyte and two metallic electrodes. <strong>Oxygen</strong> diffuses through<br />

13<br />

A<br />

V


14 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

the membrane and is electrochemically reduced at the cathode. There is a<br />

carefully fixed voltage between the cathode and an anode so that only oxygen<br />

is reduced (explained further a little later in this chapter). The greater<br />

the oxygen partial pressure is, the higher the amount <strong>of</strong> oxygen that diffuses<br />

through the membrane in a given time. This results in a current that is proportional<br />

to the oxygen concentration in the sample, which can be measured<br />

with an external device, between the anode and the cathode. Temperature<br />

sensors can be built into the probe to allow compensation <strong>for</strong> the membrane<br />

and sample temperatures, as the temperature <strong>of</strong> these affect diffusion speed<br />

through the membrane and solubility in the electrolyte. The potentiostat<br />

(an instrument commonly used <strong>for</strong> measuring differences in electric potential.<br />

Essentially, this instrument balances the unknown voltage against a<br />

known, adjustable voltage. A potentionstat is frequently used in conjunction<br />

with a thermocoupler <strong>for</strong> measuring temperature. More about this in section<br />

3.2) uses cathode current, sample temperature, membrane temperature,<br />

barometric pressure and salinity in<strong>for</strong>mation to calculate the dissolved oxygen<br />

content <strong>of</strong> the sample in either concentration (ppm) or percent saturation<br />

(% Sat). The voltage <strong>for</strong> the reduction can either be supplied electronically<br />

by the meter (potentiometric oxygen electrode), or by dissimilar metals being<br />

used <strong>for</strong> the two electrodes, picked in a way so that the correct voltage<br />

is generated between them (galvanic electrode)[22].<br />

So how does this all work? In the following chapter I will cover the theory<br />

behind the Clark sensor, and explain what is happening when it is used. First<br />

a look at the chemistry in the Clark sensor, then the diffusion through the<br />

membrane, and finally the potentiostat.<br />

3.1 Chemistry <strong>of</strong> the Clark <strong>Sensor</strong><br />

When an electrode <strong>of</strong> noble metal such as platinum or gold is biased between<br />

-0.6 and -0.8 V [23] with respect to a reference electrode such as Ag/AgCl or<br />

a calomel electrode in a KCl solution, the oxygen dissolved in the liquid is<br />

reduced at the surface <strong>of</strong> the noble metal (the cathode). This phenomenon<br />

can be observed from a current-voltage diagram (called a polarogram, example<br />

<strong>of</strong> this can be seen on figure 3.2) <strong>of</strong> the electrode. When the negative<br />

voltage applied to the noble metal electrode (called the cathode) is increased,<br />

the current increases initially but soon it becomes saturated. In this plateau<br />

region <strong>of</strong> the polarogram, the reaction <strong>of</strong> oxygen at the cathode is so fast that<br />

the rate <strong>of</strong> reaction is limited by the diffusion <strong>of</strong> oxygen to the cathode sur-


3.1. CHEMISTRY OF THE CLARK SENSOR 15<br />

face. When the negative bias voltage is further increased, the current output<br />

<strong>of</strong> the electrode increases rapidly due to other reactions. If a fixed voltage in<br />

the plateau region (<strong>for</strong> instance, -0.7V) is applied to the cathode, the current<br />

output <strong>of</strong> the electrode can be linearly calibrated to the dissolved oxygen<br />

(short termed DO). It should be noted that the current is proportional not<br />

to the actual concentration but to the activity or equivalent partial pressure<br />

<strong>of</strong> dissolved oxygen, which is <strong>of</strong>ten referred to as oxygen tension. A fixed<br />

voltage between -0.6 and -0.8 V (due to the cell potential, and the onset <strong>of</strong><br />

other processes) is usually selected as the polarization voltage when using<br />

Ag/AgCl as the reference electrode.<br />

Figure 3.2: Polarogram: Current to voltage at different oxygen tensions and<br />

calibration obtained at a fixed polarization voltage. Reproduced<br />

from [24]. The shape <strong>of</strong> the curves indicates the nature <strong>of</strong> the<br />

substance in the sample that is being reduced; where the height<br />

is equivalent to the concentrations <strong>of</strong> oxygen. Also note that<br />

convention usually have -V along the x-axis and not V, as is<br />

otherwise the norm.<br />

The cathode, the reference electrode, and the electrolyte are separated<br />

from the measurement medium by a membrane, which is permeable to the<br />

dissolved gas but not to most <strong>of</strong> the ions and other species. When at the<br />

same time most <strong>of</strong> the mass transfer resistance is confined in the membrane,<br />

the electrode system can measure oxygen tension in various liquids. This is<br />

the basic operating principle <strong>of</strong> the membrane covered DO probe[25].<br />

Electrode Reactions: The reaction can proceed as follows (a 4e − reaction<br />

with a Ag/AgCl electrode is used as an example)<br />

Cathodic reaction:


16 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

Anodic reaction:<br />

Overall reaction:<br />

O2 + 2H2O + 2e − → H2O2 + 2OH −<br />

H2O2 + 2e − → 2OH −<br />

O2 + H2O + 4e − → 4OH −<br />

Ag + Cl − → AgCl + e −<br />

4Ag + O2 + 2H2O + 4Cl − → 4AgCl + 4OH −<br />

Two pathways are possible <strong>for</strong> the reduction <strong>of</strong> oxygen at the noble metal<br />

surface. One is a 4-electron pathway where the oxygen in the bulk diffuses<br />

to the surface <strong>of</strong> the cathode and is converted to H2O via H2O2. The other<br />

is a 2-electron pathway where the H2O2 diffuses directly from the cathode<br />

surface into the bulk liquid.<br />

Also as can be seen above from the reaction, hydroxyl ions are constantly<br />

being substituted <strong>for</strong> chloride ions once the reaction starts, there<strong>for</strong>e KCl<br />

or NaCl have to be used as an electrolyte. However this electrolyte will<br />

slowly be depleted <strong>of</strong> Cl − , so something has to be done to keep this depletion<br />

from occurring to fast, or to replenish it somehow. Hence this is one <strong>of</strong> the<br />

problems that have to be faced when building a Clark sensor, as it can be<br />

one <strong>of</strong> the main limits <strong>for</strong> the lifetime <strong>of</strong> the sensor.<br />

3.1.1 One layer electrode model<br />

Now three assumptions will be used to look at a one dimensional diffusion<br />

equation[26]:<br />

1. The cathode is well polished and the membrane is tightly fit over the<br />

cathode surface such that the thickness <strong>of</strong> the electrolyte layer between<br />

the membrane and the cathode is negligible. There<strong>for</strong>e the membrane<br />

will define the oxygen flux to the cathode surface.<br />

2. The liquid around the sensor is well agitated so the partial pressure <strong>of</strong><br />

oxygen at the membrane surface is the same as that <strong>of</strong> the bulk liquid.<br />

3. <strong>Oxygen</strong> diffusion occurs only in one direction, perpendicular to the<br />

cathode surface.


3.1. CHEMISTRY OF THE CLARK SENSOR 17<br />

PSfrag replacements<br />

y<br />

Cathode<br />

Membrane Liquid<br />

x = 0 x = dm<br />

Figure 3.3: One layer electrode model to illustrate the three assumptions, as<br />

well as to help derive the equations based on them.<br />

Using Fick’s second law and the coordinate system from Figure 3.3, the<br />

unsteady diffusion in the membrane will be described by:<br />

∂p<br />

∂t<br />

∂<br />

= Dm<br />

2p ∂x2 p0<br />

x<br />

(3.1)<br />

Where Dm is the oxygen diffusivity in the membrane, p the partial pressure<br />

<strong>of</strong> oxygen in the membrane, and x is the distance from the cathode<br />

surface.<br />

Under the assumption that diffusion have yet to take place, the initial<br />

boundary conditions are:<br />

p = 0 at t = 0 (3.2)<br />

p = 0 at x = 0 (3.3)<br />

p = p0 at x = dm<br />

(3.4)<br />

where dm is the membrane thickness and po is the partial pressure <strong>of</strong> oxygen<br />

in the bulk liquid.


18 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

The first boundary condition assumes that we insert the membrane and cathode<br />

in the liquid, and as such no pressure have yet to build up at the cathode.<br />

The second boundary condition assumes a fast reaction at the cathode surface,<br />

which is generally achieved when the cathode is properly polarized.<br />

The third boundary condition assumes there is a state <strong>of</strong> equilibrium inside<br />

the membrane, hence no pressure is being applied from within so that the<br />

only pressure at the surface <strong>of</strong> the membrane is coming from the outside<br />

liquid.<br />

Under these boundary conditions the solution[27] <strong>of</strong> Eq. 3.1 is<br />

p<br />

p0<br />

= x<br />

+<br />

dm<br />

∞ 2<br />

nπ (−1)nsin nxπ<br />

<br />

2 2 −n π Dmt<br />

exp<br />

n=1<br />

dm<br />

d 2 m<br />

(3.5)<br />

The current output I <strong>of</strong> the electrode is proportional to the oxygen flux<br />

at the cathode surface:<br />

<br />

∂C<br />

I = NF ADm<br />

∂x<br />

x=0<br />

<br />

∂p<br />

= NF APm<br />

∂x<br />

x=0<br />

(3.6)<br />

Where N is the number <strong>of</strong> electrons per mole <strong>of</strong> reduced oxygen, F is<br />

Faraday’s constant, A the surface area <strong>of</strong> the cathode, C the concentration 1<br />

and Pm the oxygen permeability 2 <strong>of</strong> the membrane.<br />

Pm is related to diffusivity by<br />

Pm = DmSm<br />

where Sm is the oxygen solubility 3 <strong>of</strong> the membrane.<br />

From Eq. 3.5 and Eq. 3.6 the current output I <strong>of</strong> the electrode is<br />

I = NF A Pm<br />

p0<br />

dm<br />

<br />

1 + 2<br />

∞<br />

(−1) n 2 −n 2π Dmt<br />

exp<br />

<br />

n=1<br />

d 2 m<br />

(3.7)<br />

(3.8)<br />

<br />

1 ∂C<br />

J = Dm ∂x being the oxygen flux in the x direction<br />

x=0<br />

2The passage or diffusion <strong>of</strong> a gas, vapor, liquid, or solid through a barrier without<br />

physically or chemically affecting it.<br />

3Ability <strong>of</strong> a substance to dissolve in another substance


3.1. CHEMISTRY OF THE CLARK SENSOR 19<br />

The pressure pr<strong>of</strong>ile within the membrane and the current output under<br />

steady state conditions can then be obtained from Eq. 3.5 and Eq. 3.8:<br />

and<br />

p<br />

p0<br />

= x<br />

dm<br />

I = NF A Pm<br />

dm<br />

· p0<br />

(3.9)<br />

(3.10)<br />

At steady state, the pressure pr<strong>of</strong>ile is linear and the current output is<br />

proportional to the oxygen partial pressure in the bulk liquid, where Eq. 3.10<br />

<strong>for</strong>ms the basis <strong>for</strong> DO measurement by the sensor. The response time <strong>of</strong> the<br />

sensor as seen from Eq. 3.8 is:<br />

τ = d2 m<br />

Dm<br />

(3.11)<br />

Where τ determines how fast the sensor responds, due to the thickness<br />

<strong>of</strong> the membrane or a high Dm. However these conditions tend to weaken<br />

the assumption <strong>of</strong> membrane-controlled diffusion. There<strong>for</strong>e, a compromise<br />

has to be made <strong>for</strong> optimum sensor per<strong>for</strong>mance. Changing dm (rather than<br />

Dm) is more effective in adjusting τ (since it depends on the square <strong>of</strong> dm).<br />

Eq. 3.10 and Eq. 3.11 gives the indication, that the design variables <strong>for</strong> a<br />

DO sensor is Pm, dm, Dm, and A.<br />

3.1.2 Two layer model<br />

The problem is, however, that the second assumption, made earlier in this<br />

chapter, is not entirely accurate. A stagnant liquid film will almost always<br />

exists right outside the membrane even at high liquid velocities.[28] Hence<br />

a more realistic model have to be made, in order to account <strong>for</strong> this film as<br />

shown on Figure 3.4. This take into consideration the effect <strong>of</strong> the film, as<br />

it slows down the diffusion through the membrane.<br />

The effect <strong>of</strong> the liquid layer on the sensor current can be calculated by<br />

expanding the previous model, at steady state the oxygen flux J through<br />

each layer should be the same.


20 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

PSfrag replacements<br />

y<br />

Cathode<br />

Membrane Liquid<br />

High Velocity<br />

Low velocity<br />

Figure 3.4: Two layer model <strong>for</strong> DO sensor to illustrate the changed conditions.<br />

J = Kp0 = kL(p0 − pm) = kmpm<br />

p0<br />

x<br />

(3.12)<br />

Where K is the overall is the overall mass transfer coefficient and kL and<br />

km are the mass transfer coefficient <strong>for</strong> the liquid film and the membrane.<br />

The inverse <strong>of</strong> the mass transfer coefficient can be termed as the mass<br />

transfer resistance. If kL and km is thought <strong>of</strong> as two parallel resistors, we<br />

get:<br />

1<br />

K<br />

1<br />

= +<br />

kL<br />

1<br />

km<br />

(3.13)<br />

What this equation says is that the overall mass transfer resistance is the<br />

sum <strong>of</strong> mass transfer resistance and the membrane phase transfer resistance.<br />

The derivation is similar to Ohm’s law, where J is equivalent <strong>of</strong> the current,<br />

and p <strong>of</strong> the voltage.<br />

The individual resistance can be replaced by more commonly known parameters:<br />

1<br />

K<br />

dL<br />

= +<br />

PL<br />

dm<br />

Pm<br />

(3.14)


3.1. CHEMISTRY OF THE CLARK SENSOR 21<br />

Where dL is the liquid film thickness and PL the oxygen permeability <strong>of</strong><br />

the liquid film.<br />

When the individual mass transfer resistances are included, the steady<br />

state sensor output becomes:<br />

Where ¯ d is defined as:<br />

I = NF A Pm<br />

¯d p0<br />

dL<br />

¯d = dm + Pm<br />

PL<br />

The time constant τ from Eq. 3.11 can be modified to:<br />

with d defined as:<br />

τ = d2<br />

Dm<br />

d = dm + dL<br />

Dm<br />

DL<br />

(3.15)<br />

(3.16)<br />

(3.17)<br />

(3.18)<br />

The DO sensor, when placed in a stagnant liquid, produces a diffusion<br />

gradient extending outside the membrane and farther into the liquid. When<br />

the liquid is stirred, the diffusion gradient can no longer be extended beyond<br />

the liquid film around the membrane. Since the diffusion gradient becomes<br />

steeper with decreasing liquid film thickness, the current output <strong>of</strong> the sensor<br />

increases with increase in liquid velocity, also that the response time <strong>of</strong> the<br />

sensor increases as the liquid velocity decreases. This ’flow sensitivity’ is<br />

greater <strong>for</strong> a sensor with a larger cathode because the size <strong>of</strong> the stagnant<br />

diffusion field is proportionally greater with a larger cathode. Hence there<br />

is an advantage in going from macro to micro scale, as the cathode will<br />

naturally be smaller here.<br />

Eq. 3.15 can be written as:<br />

I = NF A p0<br />

dm<br />

Pm<br />

+ dL<br />

PL<br />

(3.19)


22 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

And from this the conditions <strong>for</strong> membrane controlled diffusion becomes:<br />

dm<br />

Pm<br />

≫ dL<br />

PL<br />

(3.20)<br />

In order to achieve this condition, a relatively thick membrane with a<br />

low oxygen permeability have to be used. When this condition is achieved,<br />

the oxygen sensor output depends only on membrane properties as given by<br />

Eq. 3.10 and the sensor calibrated in one liquid can be used in other liquids<br />

without recalibration. However there is always a liquid film (however thin it<br />

may be) and this causes variations in calibration in different liquids.<br />

3.2 The Potentiostat<br />

Most <strong>of</strong> us can probably recall from the early chemistry lessons in school ,<br />

that if you take a piece <strong>of</strong> iron and dips it into a sulphuric acid, it will start to<br />

dissolve, or rather corrode. Though in the 17th century Sir Humphrey Davy,<br />

following Galvani’s experiments [29] discovered that if a piece <strong>of</strong> non corrosive<br />

metal, like Platinum, was put in the acid, and then connected to a positive<br />

pole. While at the same time connecting the piece <strong>of</strong> iron to the negative pole<br />

<strong>of</strong> the same current source. The corrosion <strong>of</strong> the iron would slow down, or<br />

even come to a complete halt if a high enough amount <strong>of</strong> voltage was applied.<br />

Vice versa the corrosion would increase with the voltage, if the iron was<br />

connected to the positive pole and the platinum to the negative pole.<br />

Though above a specific current(the exact one varying with the temperature,<br />

composition, and size <strong>of</strong> the metals) the current will suddenly drop to<br />

low values and the iron would cease to corrode (A phenomenon discovered<br />

first by Michael Faraday.[30]). The understanding and explanation <strong>for</strong> this<br />

was later brought on by the invention <strong>of</strong> the potentiostat.<br />

As mentioned at the beginning <strong>of</strong> this chapter, a potentiostat is used in<br />

order to measure the current, so what does this device do, and what is the<br />

principle behind it. Basically the potentiostat, is an amplifier that sets and<br />

maintains a voltage between two electrodes at a specific value, a Working<br />

Electrode (also sometimes referred to as the Indicator Electrode[31]) and a<br />

Reference electrode.


3.2. THE POTENTIOSTAT 23<br />

Reference<br />

Electrode<br />

Constant E required<br />

Working<br />

Electrode<br />

Figure 3.5: A basic electrochemical cell with a cathode (Working Electrode)<br />

and a anode (Reference Electrode).<br />

However in order <strong>for</strong> this to be accomplished some conditions have to be<br />

fulfilled[32]. The Reference electrode maintains a constant voltage referred to<br />

as the hydrogen electrode potential, an <strong>of</strong>ten used metal <strong>for</strong> this (as is also the<br />

case in this project) is Ag covered with a layer <strong>of</strong> AgCl. The moment a current<br />

passes through this electrode it will become polarized, eg its potential will<br />

vary accordingly to the current. There<strong>for</strong>e in order <strong>for</strong> a steady potentential<br />

can be maintained, no current must be allowed to pass through it. This<br />

is accomplished by having a constant potential difference between the two<br />

electrodes. However in order to achieve this a third electrode have to be<br />

added to the cell, a Counter Electrode (also known as a Auxiliary Electrode).<br />

Hence a current is <strong>for</strong>ced between the Counter Electrode and the Working<br />

Electrode, this current is high enough and in the proper polarity to keep<br />

the potential <strong>of</strong> the Working Electrode at a set value in comparison to the<br />

Reference Electrode.<br />

There will be two focuses <strong>of</strong> the Potentiostat[33]:<br />

1. Measuring the potential difference between the Reference and Working<br />

Electrode, in a way that leaves the Reference Electrode unpolarized,<br />

while comparing this potential difference to an already preset voltage.<br />

2. Force a current between the Counter Electrode and the Working Electrode<br />

to counterbalance the difference between the Working Electrode<br />

potential and the preset voltage.<br />

This is realized electronically by using a operational amplifier also known<br />

as an OPAMP, the operational amplifier will has two inputs, an inverting and


24 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

Working<br />

Electrode<br />

Constant E required<br />

Reference<br />

Electrode<br />

Current<br />

Input<br />

Counter<br />

Electrode<br />

Figure 3.6: A basic electrochemical cell with a third electrode (Counter Electrode)<br />

added to control the potential.<br />

a non-inverting. If a current/voltage is feed into the non-inverting part, an<br />

amplified current/volage will be produced, while feeding it into the inverting<br />

part will maintain the magnitude, while switching the sign (+ to − or −<br />

to +). By closing the loop between in- and output, the voltage difference<br />

between the two inputs <strong>of</strong> the amplifier will fall. An increase in the voltage in<br />

the inverting input will <strong>for</strong>ce a complementary current on the output, which<br />

will counteract the difference in input voltage.<br />

Connecting the Reference Electrode to the inverting input, the Working<br />

Electrode to the non-inverting, and the Counter Electrode to the output will<br />

give this:<br />

Between the Working Electrode and the Reference Electrode the difference<br />

will now be amplified and inverted by the amplifier, the circuit will<br />

be closed by the electrochemical cell, where the current runs through the<br />

electrolyte from the Counter Electrode to the Working Electrode. This will<br />

polarize the Working Electrode, so that the input difference between that and<br />

the Reference Electrode will be zero. In doing so, the Working Electrode and<br />

the Reference Electrode potential is kept level. To shift the Working Electrode<br />

potential to some other reference value in regards to the Reference<br />

Electrode, a voltage in series between the input to the Reference Electrode<br />

and the Reference Electrode itself must be inserted. Also to measure the current<br />

running through the Counter Electrode, a resistor is used in the wiring<br />

<strong>of</strong> the Counter Electrode, over which a voltage that is proportional to the


3.2. THE POTENTIOSTAT 25<br />

Working<br />

Electrode<br />

Reference<br />

Electrode<br />

+<br />

OPAMP<br />

−<br />

Counter<br />

Electrode<br />

Figure 3.7: Using the OPAMP as a potentiostat.<br />

current is measured (See Figure 3.8)<br />

Working<br />

Electrode<br />

Reference<br />

Electrode<br />

+<br />

OPAMP<br />

−<br />

R<br />

Counter<br />

Electrode<br />

Figure 3.8: Resistor added to measure the current running through the<br />

Counter Electrode.<br />

Of course a more complex electronic system is still needed to protect<br />

the potential amplifier from voltage shocks, to counteract noise and other<br />

miscellaneous problems with electrical systems.<br />

V


26 CHAPTER 3. THEORY OF THE CLARK SENSOR<br />

3.3 Summary<br />

The theories upon which the Clark sensor is based, as well as the potentiostat,<br />

have been highlighted, and it is now a matter <strong>of</strong> trying to use these in order<br />

to design and build a micro oxygen sensor. While trying to find a way around<br />

the problems that electrolyte depletion and membrane fouling presents.


Chapter 4<br />

System Design<br />

The sensor layout is one <strong>of</strong> the more essential parts <strong>of</strong> the project, as the<br />

design and general setup will be the deciding factor, in how well the sensor<br />

will work. At such at least one design should be made that is reasonably<br />

safe, which based on the general theory should work.<br />

In this chapter the thoughts and ideas, concerning why the first designs <strong>of</strong><br />

the sensor(s) looks as they do, are given and explained.<br />

4.1 General Design<br />

Based on a previous project by Anders Hyldg˚ard [34], an ’O’-ring based<br />

design was decided on, this was done in part to take advantage <strong>of</strong> the already<br />

existing packaging scheme 1 . The setup uses an ’O’-ring as pressure sealing,<br />

which leaves the active part <strong>of</strong> the sensor exposed, while protecting the rest<br />

<strong>of</strong> the chip from the corrosive effects <strong>of</strong> seawater[35]. Of course care still has<br />

to be taken, to ensure that it is tightly sealed, since even a small roughness<br />

on the surface <strong>of</strong> the chip might allow seawater to seep through. Furthermore<br />

a lot <strong>of</strong> the clean room steps from that project can be duplicated to create<br />

the various layers in this design.<br />

Making the sensor circular however presents other problems, as the sensor<br />

needs a membrane covering it as mentioned in a Chapter 3, so that only<br />

oxygen diffusing through the membrane is measured, as well as an electrolyte.<br />

This in itself may not seem like a major obstacle, though given that the<br />

membrane (Siloprene) is liquid to begin with, it needs a small ’container’<br />

1 See the Packaging section <strong>for</strong> more info on this<br />

27


28 CHAPTER 4. SYSTEM DESIGN<br />

until it solidifies. Experiments have been made by M. Dawgul et. al. [36], in<br />

which a membrane has been spin coated on the wafer. The results from this<br />

however were not good, as the membrane <strong>of</strong>ten ended up sliding <strong>of</strong>f, or at<br />

best varied greatly in thickness, thereby lacking in regards to the feasibility<br />

<strong>of</strong> being able to create a preset, uni<strong>for</strong>m membrane thickness.<br />

A logical solution to this has been used by M. Wittkampf et. al [37] where<br />

a second wafer is anodic bonded to the first, with the second wafer having a<br />

hole created with a wet isotropic etch where the exposed part <strong>of</strong> the sensor is,<br />

there by creating a hole that the membrane liquid can be ’poured’ into. This<br />

is a fairly simple process, however the basic <strong>of</strong> this design makes it a little<br />

more difficult, as a circular rather than a rectangular hole have to be made.<br />

However it is a problem that can be overcome. In the initial design, the<br />

’O’-ring will be used as a ’container’ <strong>for</strong> the electrolyte and the membrane.<br />

Also it should be mentioned that a micro electrode have the advantage over<br />

a macro electrode, in that the diffusion occurs in all directions, where as with<br />

the macro it only have diffusion in one direction[38]. Hence at macro scale<br />

the oxygen will be consumed faster at the electrode, and thereby influence<br />

the measurements. (See Figure 4.1)<br />

Figure 4.1: Micro electrodes have diffusion to all directions while Macro only<br />

have to one, why at macro scale setup the oxygen is consumed<br />

faster at the electrode.<br />

4.2 Actual Design<br />

Four ’major’ designs, each with some variant sub design, have been created,<br />

taking into consideration some <strong>of</strong> the experiences that have already been<br />

made by others[37, 39, 36, 38, 40, 41]. The idea behind the main one as


4.2. ACTUAL DESIGN 29<br />

seen on figure 4.2 have been to create a layout, which as mentioned earlier,<br />

would have a high chance <strong>of</strong> working. Thereby enabling me to do some<br />

measurements with it, so that some practical experience could be gathered.<br />

Also <strong>of</strong> course to see if it did indeed work, and what response could be<br />

expected from it. The design is basically an array <strong>of</strong> Pt working electrodes<br />

located in the middle, then a torus shaped Pt counter electrode surrounding<br />

it, and finally a circular Ag/AgCl reference electrode around the two <strong>of</strong> them.<br />

Extreme care has been taken to avoid, or at the very least minimize, the<br />

chances <strong>of</strong> any short circuits. Put in other words each electrode have its own<br />

connection to the outside via a silicide ’wire’, that touches neither the other<br />

electrodes, nor the other ’wires’. Also there is a temperature sensor at the<br />

top, more about this later however.<br />

Figure 4.2: Primary/Safe Design, the white circles are where the ’O’-ring will<br />

be placed on the chip, showing both the actual inner and outer<br />

diameter <strong>of</strong> it, as well as where it minimally will touch the chip.<br />

The number refer to the design, while the roman numerals refer<br />

to the size and distance between the electrodes in the array. For<br />

a closeup <strong>of</strong> the Working Array see Figure 4.6


30 CHAPTER 4. SYSTEM DESIGN<br />

Based on this design, three others were created, a second where the a<strong>for</strong>ementioned<br />

short circuits is assumed not to take place. A third, where the<br />

counter electrode and the counter array are switched, and a fourth, where<br />

the size <strong>of</strong> the reference electrode is increased.<br />

The second (See Figure 4.3) is intended to replace the first if it works, since<br />

as can be seen on the figure, it is basically the same as the first, except that<br />

I have here assumed no short circuits will occur.<br />

Figure 4.3: Second Design, note that the electrodes are now closed rings and<br />

lies over the contact layer, albeit with a Nitride layer inbetween


4.2. ACTUAL DESIGN 31<br />

The third design (See Figure 4.4) is partly done to see if the right choice<br />

was made in placing the working array in the center, instead it have here<br />

switched place with the counter electrode. As such it is more done out <strong>of</strong><br />

curiosity to learn more about the workings <strong>of</strong> a Clark sensor, rather than a<br />

design in which great expectations is placed.<br />

Figure 4.4: Third Design, the working array is now the inner ring, rather<br />

than the center circle


32 CHAPTER 4. SYSTEM DESIGN<br />

The reference electrode is always placed furthest out in the circle, due<br />

to the silver being slowly consumed as the sensor is used. There<strong>for</strong>e it is<br />

preferable to have as much <strong>of</strong> it to consume as possible, in order to extend<br />

the lifetime <strong>of</strong> the sensor, which is the basis <strong>of</strong> the fourth design (See Figure<br />

4.5).<br />

Figure 4.5: Fourth Design, the reference electrode has been enlarged.<br />

4.2.1 Minor Design Variations<br />

In order to get a better understanding <strong>of</strong> the sensor, as well as trying to<br />

optimize it, some variations have be made on various aspects <strong>of</strong> the sensor,<br />

such as distance between each electrode in the array (See Figure 4.6), as well<br />

as the size <strong>of</strong> each electrode in the array (See Figure 4.7). Other possible<br />

variations could include the distance between the reference electrode, the<br />

counter electrode and the working array, as well as making the already used<br />

variations smaller or larger.<br />

Why these variations? Well as the reference electrode (Ag/AgCl) is slowly<br />

consumed, it will be deposited at the working array, the speed at which this<br />

occur will in part depend on the size and numbers <strong>of</strong> array electrodes. So if<br />

there is too few, or they are too small, they will be covered faster. On the


4.3. TEMPERATURE SENSOR 33<br />

Figure 4.6: Closeup <strong>of</strong> the working array, there is 150 µm between each electrode<br />

on the left array, and 100 µm between each on the right.<br />

The radius <strong>of</strong> the green circle (contact area <strong>for</strong> the working array)<br />

is in both cases 350 µm.<br />

other hand if there is too many in the array, there is a risk <strong>of</strong> the array starting<br />

to act like a macro scale electrode, where the oxygen will be consumed, hence<br />

corrupt the obtained measurements. Not to mention that the sensor itself<br />

will be used underwater, while being attached to a fish, hence no external<br />

power source is possible. There<strong>for</strong>e it is an advantage if it uses as little power<br />

as possible, this equals a smaller battery which in turn will make the overall<br />

size <strong>of</strong> the finished device as small as possible, and hence it can be attached<br />

to smaller fish. Finally the size and distance also influence the signal output<br />

<strong>of</strong> the device, as the measured current is partly dependable on these two<br />

parameters.[42, 43]<br />

So it is in large part a question <strong>of</strong> finding a balance between all the factors,<br />

to ensure the lifetime is as long as possible, without sacrificing accuracy.<br />

There<strong>for</strong>e variations on all conceivable variables should be looked into.<br />

4.3 Temperature <strong>Sensor</strong><br />

Since the diffusion through the membrane among other things is dependent<br />

on temperature, and there was room to spare on the chip, it was deemed<br />

an advantage to add a temperature sensor (platinum resistance temperature


34 CHAPTER 4. SYSTEM DESIGN<br />

Figure 4.7: Closeup <strong>of</strong> electrode in a working array, the one on the left have<br />

a radius <strong>of</strong> 5 µm, while the right have a radius <strong>of</strong> 10 µm, to get a<br />

more accurate idea <strong>of</strong> the size compare it to the distance between<br />

the brown area (the counter electrode) and the green area (the<br />

contact layer <strong>for</strong> the electrode array), which is 20 µm in both<br />

cases. The black diagonal area is the nitride contact holes.<br />

sensor). Since the relationship between temperature and resistance <strong>for</strong> platinum<br />

is approximately linear over a small temperature range, and since the<br />

properties <strong>for</strong> platinum is well known and documented this was a preferable<br />

choice. The resistance is calculated from Eq. 4.1<br />

R = ρ l<br />

wh<br />

(4.1)<br />

Where R is the resistance, ρ the resistivity <strong>of</strong> the material (ρ = 1.06 ×<br />

10 −8 Ωm <strong>for</strong> Pt), and l, w and h the length, width and height <strong>of</strong> the resistor.<br />

Figure 4.8: Close up <strong>of</strong> the l-edit design <strong>for</strong> the Pt50<br />

3 versions <strong>of</strong> this have been implemented on the various designs, a Pt50,


4.3. TEMPERATURE SENSOR 35<br />

a Pt300, and a Pt600 (where the number is the resistance <strong>of</strong> the sensor.<br />

Figure 4.9: Close up <strong>of</strong> the l-edit design <strong>for</strong> the Pt300<br />

Of these 3, the Pt50 is the safest design. The reason <strong>for</strong> this is that the<br />

The Pt300 and Pt600 have a width <strong>of</strong> 10 µm, while the Pt50 have a width<br />

<strong>of</strong> 20 µm. Hence if some error occurs during the metallization part <strong>of</strong> the<br />

process <strong>for</strong> the sensor, and you end up with a width <strong>of</strong> 9 µm instead, which is<br />

merely 1 µm <strong>of</strong>f. It would translate into a 10% shift in the resistance, while it<br />

would merely be 5% with the Pt50. Hence having a reasonably wide resistor<br />

can minimize a potential error, un<strong>for</strong>tunately however the higher the width,<br />

the lower the resistance is as well. So once again it is a matter <strong>of</strong> balancing<br />

one advantage against another disadvantage.<br />

Figure 4.10: Close up <strong>of</strong> the l-edit design <strong>for</strong> the Pt600, it is basically the<br />

Pt300 done twice.<br />

Why the 4 contacts, instead <strong>of</strong> merely 2? Well consider if there was only<br />

2, each serving as a contact <strong>for</strong> both current and voltage, the total resistance<br />

between the two can then be calculated from:<br />

R = V<br />

I = 2Rc + 2Rsp + Rs<br />

(4.2)<br />

Where Rc is the contact resistance between the chip and the equipment<br />

connected to it, Rsp the spreading resistance, and Rs the chip resistance (and


36 CHAPTER 4. SYSTEM DESIGN<br />

the resistance that I want to assign). This in itself may seem straight<strong>for</strong>ward,<br />

however neither Rc nor Rsp can be calculated precise, so as a result it isn’t<br />

possible to measure Rs accurately.<br />

The 4-contact approach solves this problem however; 2 carry the current<br />

while the other 2 are used to measure the voltage. While Rc and Rsp is still<br />

there, they are now negligible due to the very small current that flows through<br />

them, as the voltage is now measured with a potentiometer that draws no<br />

current, or with a high impedance voltmeter that draws little current.<br />

Temperature Calculated Width Effective Heigth<br />

<strong>Sensor</strong> Resistance (Ω) (µm) Length (µm) (µm)<br />

Pt50 47.8 20 18019 0.2<br />

Pt300 287 10 54170 0.2<br />

Pt600 590 10 113207 0.2<br />

4.4 Packaging<br />

Table 4.1: Overview <strong>of</strong> the temperature sensors.<br />

Shielding the interface electronic part <strong>of</strong> the chip from the corrosive effect<br />

<strong>of</strong> seawater, is as mentioned earlier the ’O’-ring and packaging. How is this<br />

built?<br />

The package or casing consists <strong>of</strong> three different layers (hereafter referred<br />

to as the lower, middle and upper layer) <strong>of</strong> laser 2 cut PMMA (PolyMethyl-<br />

MethAcrylate), that each have their own structure. While there are materials<br />

that are not as porous, and have better chemical properties, PMMA<br />

have the advantage that several can be fabricated within minutes once the<br />

actual structure is in place. Which have a higher priority than long term<br />

usage as far as testing purposes goes. Also PMMA can generally be used in<br />

the temperature area from -40 o C to 70 o C, as well as being electric isolating.<br />

So what do these layers look like, and what purpose does each serve?<br />

2 Laser used is a SYNRAD laser


4.4. PACKAGING 37<br />

Figure 4.11: Close up photography <strong>of</strong> a flex print.<br />

1 mm<br />

1.7 mm<br />

11 mm<br />

6.5 mm<br />

66.5 mm<br />

Figure 4.12: Rough sketch <strong>of</strong> the middle layer <strong>of</strong> the encapsulation, sizes<br />

listed to give a clearer picture <strong>of</strong> actual size<br />

Figure 4.13: Pictures taken <strong>of</strong> the middle layer, to show what it looks like<br />

once fabricated, the layer is tilted slight on the picture to the<br />

right to compensate <strong>for</strong> the light.<br />

The middle layer as seen on figure 4.12 and 4.13 is where the chip is<br />

located, 4 areas are cut into the PMMA with a laser. The area with the


38 CHAPTER 4. SYSTEM DESIGN<br />

chip, 2 adjourning not quite as deep areas (a short and a long area) where<br />

the flex print (This provide the electronic wiring to devices outside the sensor,<br />

see Figure 4.11 <strong>for</strong> a picture <strong>of</strong> the flex print) is located, and a hole <strong>for</strong> the<br />

flex print on the short side (So that the connection to the outside system is<br />

located in one end, with the sensor at the other.)<br />

The lower layer as seen on figure 4.14 and 4.15 have been cut to provide<br />

a a grove path where the flex print sticking through the hole on the middle<br />

can run back.<br />

6 mm<br />

Figure 4.14: Rough sketch <strong>of</strong> the lower layer <strong>of</strong> the encapsulation, sizes listed<br />

to give a clearer picture <strong>of</strong> actual size<br />

1.5 mm<br />

Figure 4.15: Pictures taken <strong>of</strong> the lower layer.<br />

The upper layer as seen on figure 4.16 and 4.17 consist <strong>of</strong> 3 parts, 2<br />

square holes that provides room <strong>for</strong> the connection between then chip and<br />

the flex print since there isn’t a direct contact between the two (More on this<br />

connection and how the 3 layers are made to stick together in Chapter 5).


4.4. PACKAGING 39<br />

90 mm<br />

15 mm<br />

6 mm<br />

4.75 mm<br />

4.5 mm<br />

4.75 mm<br />

Figure 4.16: Rough sketch <strong>of</strong> the middle layer <strong>of</strong> the encapsulation, sizes<br />

listed to give a clearer picture <strong>of</strong> actual size<br />

2 mm<br />

0.5 mm<br />

Figure 4.17: Pictures taken <strong>of</strong> the upper layer.<br />

Located between the two squares are a circle with a small hole cut entirely<br />

through the PMMA. The ’O’-ring is located in the circle, see Figure 4.18,<br />

while the hole will be located directly above the center <strong>of</strong> the center on the<br />

chip. Thereby allowing the dissolved oxygen to interact with the 3 electrodes<br />

(Reference, Working and Counter) on the chip. Once the three layers are<br />

put together, the ’O’-ring should be compressed roughly 20%. The ’O’-ring<br />

itself should be chosen, so that it can resist the water, acids, oils and other<br />

biological and non-biological substances in the ocean.


40 CHAPTER 4. SYSTEM DESIGN<br />

Figure 4.18: Pictures taken <strong>of</strong> the ’O’-ring, on the picture to the right the<br />

’O’-ring is placed in the PMMA.<br />

4.5 Summary<br />

4 different sensor layouts have been designed, along with variations on electrode<br />

size and distance, as well as 3 different Pt-temperature sensors. Combing<br />

into a total <strong>of</strong> 13 different chips, opening up <strong>for</strong> the possibility to test<br />

various aspect <strong>of</strong> the overall design <strong>of</strong> the Clark sensor. Finally the encapsulation<br />

have been chosen so that it is easy and fast to produce, while still<br />

being durable. Table 4.2 shows the various chips, and which design variation<br />

that can be found on each.<br />

Chip Electrode Electrode Temperature<br />

Design Distance (µm) Radius (µm) <strong>Sensor</strong><br />

1CVA 100 5 Pt50<br />

1CVB 100 5 Pt300<br />

1CXA 100 10 Pt50<br />

1CXB 100 10 Pt300<br />

1CLV 150 5 Pt300<br />

1CLXA 150 10 Pt50<br />

1CLXB 150 10 Pt300<br />

2CV 100 5 Pt600<br />

2X 100 10 Pt600<br />

3CV 100 5 Pt300<br />

3CX 100 10 Pt300<br />

4CV 100 5 Pt300<br />

4CX 100 10 Pt300<br />

Table 4.2: Overview <strong>of</strong> the chips, listing the size and distance between the<br />

electrodes in the working electrode array, and what temperature<br />

sensor there is on it.


Chapter 5<br />

Fabrication<br />

The fabrication 1 process is a fairly straight <strong>for</strong>ward process, with 5 masks<br />

being used in total. Only one side <strong>of</strong> the wafer is used, so the potential problems<br />

with double sided wafers is avoided. Within this chapter is a description<br />

<strong>of</strong> the steps in the process, as well as some <strong>of</strong> the thoughts about each step.<br />

The starting material is single crystalline n-type silicon (100) wafers. Further,<br />

several steps in the fabrication process could be duplicated from the<br />

project which this is subpart <strong>of</strong> [34].<br />

Figure 5.1: Illustration <strong>of</strong> a bare Silicon wafer, <strong>for</strong> the complete process, and<br />

a better understanding <strong>of</strong> this illustration and those that will<br />

follow, refer to Appendix B.<br />

5.1 Processing<br />

Silicon Dioxide Layer<br />

First a Silicon Dioxide layer is <strong>of</strong> 2000 ˚A is grown with thermal oxidation<br />

using a wet oxidation, the layer will serve to protect the wafer from the later<br />

processes, as well as working as an insulator. The reason wet oxidation is<br />

1 For the step by step cleanroom process recipe, see Appendix C<br />

41


42 CHAPTER 5. FABRICATION<br />

used is mainly that it is faster, basically it have a faster growth rate, and<br />

that the chip doesn’t require the higher breakdown voltage <strong>of</strong> dry oxidation.<br />

Figure 5.2: A layer <strong>of</strong> Silicon Dioxide have been added.<br />

Silicide Layer<br />

Next comes the Silicide layer, the silicide is going to serve as the conduction<br />

layer, between the actual sensor and the contacts. Silicide 2 (or rather<br />

TiSi2), resistivity <strong>of</strong> 13-16mΩcm is well suited <strong>for</strong> this due to its low resistivity.<br />

Which is an advantage since the finished Clark <strong>Sensor</strong> have to have<br />

an internal power supply, due to the sensor being attached to a fish, where<br />

the battery cannot be exchanged as one see fit.<br />

Figure 5.3: Illustration <strong>of</strong> the wafer after the silicide step.<br />

The Silicide layer is created by first depositing a layer <strong>of</strong> undoped Polysilicon<br />

with Low Pressure Chemical Vapor Deposition (LPCVD). It is also at<br />

this point the first alignments marks are made (See Figure 5.6). After this a<br />

Titanium layer is deposited on top <strong>of</strong> the Polysilicon, which is the patterned<br />

as well. Finally by using Rapid Thermal Annealing (RTA), the desired TiSi2<br />

is made.<br />

2 The silicide recipe is in Appendix A


5.1. PROCESSING 43<br />

Figure 5.4: Alignment marks from Silicide.<br />

Silicon Nitride Layer<br />

The Si3N4 layer is used to protect the bulk chip (not the actual sensor itself <strong>of</strong><br />

course) from exposure to seawater, as seawater is highly corrosive and could<br />

otherwise quickly corrode the metals.<br />

Figure 5.5: Illustration <strong>of</strong> the wafer after the LPCVD and RIE <strong>of</strong> the Nitride.<br />

LPCVD is used to deposit the Si3N4, after which a RIE is used to make<br />

the contact holes to the silicide layer <strong>for</strong> the metals.<br />

Figure 5.6: Alignment marks after Si3N4. Since a dark mask is used in this<br />

step, the alignment mark is here made relatively big, so that<br />

there is hole to peek through to to the wafer in order to locate<br />

the alignment marks from the previous step.<br />

Metal(s) Layer<br />

Finally the metals, Pt (Working Electrode Array, Counter Electrode and


44 CHAPTER 5. FABRICATION<br />

Thermometer), Ag (Reference Electrode), and Au (Contacts) are deposited.<br />

Why these 3?<br />

As discussed previously Pt and Ag have been used almost exclusively <strong>for</strong><br />

Clark <strong>Sensor</strong>s, and referenced as the best suited metals from a dissolved oxygen<br />

sensor[37].<br />

Where as Au gold provides an excellent surface <strong>for</strong> bonding to wirebond, as<br />

well as being able resist some exposure to saltwater.<br />

Figure 5.7: Illustration <strong>of</strong> the wafer with the three metals added.<br />

Figure 5.8: Alignment marks after the metals are added. From left to right,<br />

ignoring the first one, it is Pt(brown), Ag(light brown), Au(red).<br />

Each metal have double alignment marks, two boxes and a crisscross<br />

pattern. This is mainly to have a little extra safety when<br />

aligning the masks, so that there is always another alignment<br />

mark to check with.<br />

Pt will be put on first <strong>for</strong> two reasons; It is the metal with the smallest<br />

structure, and hence the metal where the lithography step is most sensitive,<br />

and there<strong>for</strong>e where it is most like to go wrong. Also since the Au-Contacts<br />

and the Pt-Thermometer crosses, having the PT buried beneath the Au will<br />

ensure an unbroken line (Especially as the Au, 3000 ˚A, layer is over 10 times<br />

as thick as the Pt, 200 ˚A, layer).<br />

A minor problem however is that a lot <strong>of</strong> metals have poor adhesion to<br />

many <strong>for</strong>ms <strong>of</strong> Silicon (including Silicide), hence a titanium layer is added


5.2. BACK-END PROCESSING 45<br />

under each to improve adhesion.<br />

The 3 metals are added and patterned with 3 double rounds (First Ti, then<br />

the actual metal) <strong>of</strong> metalization and then lift-<strong>of</strong>f.<br />

5.2 Back-end Processing<br />

Once the cleanroom processes are done, there is still a few steps that have to<br />

be completed. These steps involves chemicals usually not found in a cleanroom,<br />

and the chip have to soak in them <strong>for</strong> a while, there<strong>for</strong>e they are done<br />

outside the cleanroom, also there is no real contamination danger <strong>for</strong> the<br />

electronics part <strong>of</strong> the wafer at this point.<br />

5.2.1 Packaging<br />

First the chip have to be packaged, this is done in three steps.<br />

The Flexprint and then Chip is glued to the middle layer <strong>of</strong> the PMMA. The<br />

Flexprint is attached first as to avoid getting glue onto the chip, however<br />

should it occur, the glue used (Super Attakgel) can be removed using<br />

ethanol, provided it is only a thin layer <strong>of</strong> glue.<br />

Figure 5.9: Flex print and Chip glued to the PMMA.<br />

Once the glue have dried (<strong>for</strong> 1-2 hours to be sure that it is completely<br />

dry), the flex print and the Au-contacts have to be connected. This can be<br />

done with either wirebonding, which will be the first choice, or by using a


46 CHAPTER 5. FABRICATION<br />

conducting glue.<br />

The final step in the packaging, will be to bond the 3 layers <strong>of</strong> PMMA<br />

together (the ’O’-ring is inserted here as well), this is accomplished by putting<br />

them in a oven at 115 o C <strong>for</strong> 1 hour, which will cause them to bond. To hold<br />

then layers in place, two pieces <strong>of</strong> metal are used as well as a screw clamp.<br />

Care should be taken not to put to much pressure on the PMMA with the<br />

screw clamp, nor to little. However applying less than needed pressure is<br />

preferable, as it can be put back into the oven, if the bonding have not taken<br />

place. While to much pressure can cause the chip to break (a closer look is<br />

taken on this problem in Chapter 6).<br />

Figure 5.10: Preparation <strong>for</strong> the oven. Top left: The screw clamp. Top<br />

right: The two pieces <strong>of</strong> metal that holds the PMMA, chip and<br />

flexprint in place and together. Bottom Left: The 3 layers <strong>of</strong><br />

PMMA have been placed in the the metal holder. Bottom right:<br />

Screw clamp have been closed tightly around the metal holding<br />

the PMMA.


5.2. BACK-END PROCESSING 47<br />

5.2.2 Ag to Ag/AgCl<br />

Ag has to be turned into AgCl. This can be done in several ways, one <strong>of</strong> the<br />

easier ways[44] however consists <strong>of</strong> using:<br />

1. 0.1 M HCl<br />

2. A flashlight battery.<br />

3. A few centimeters <strong>of</strong> pure silver.<br />

Figure 5.11: Ag turned into Ag/AgCl.<br />

The packaged chip is dipped down into the HCl along with one end <strong>of</strong> the<br />

silver wire. While the battery is connected with the negative terminal to the<br />

silver wire and the positive to the Ag output on the flexprint. Slowly the Ag<br />

electrode will turn a light tan and then a dark brown, at which point it will<br />

have a suitable thick layer <strong>of</strong> AgCl.<br />

5.2.3 Electrolyte and Membrane<br />

Finally the electrolyte and the membrane is added. The material used is<br />

Siloprene from Fluka-Chemie. It consists <strong>of</strong> 3 liquids Siloprene Crosslinking<br />

Agent K-11, Siloprene K 1000 and Hexane. All three liquids are dripped into<br />

the small hole on the device, where the chip is located, by using a syringe<br />

(The amounts used are <strong>of</strong> the order <strong>of</strong> 100:10:1 <strong>for</strong> Siloprene Crosslinking<br />

Agent K-11 : Siloprene K 1000 : Hexane). Finally the sensor is soaked in a<br />

liquid containing a large amount <strong>of</strong> Cl − ions (the time it needs to be soaked<br />

depends on the exact concentration, the higher the concentration is, the less<br />

time it needs. Care should be taken however if a strong acid is used, as it<br />

might damage the membrane) to add the electrolyte. Other membranes have<br />

been used with micro electrode oxygen sensor(such as Nafion[45]). However<br />

this was chosen, due to the recommendation and experiences <strong>of</strong> a pr<strong>of</strong>essor<br />

at MIC.


48 CHAPTER 5. FABRICATION<br />

5.3 Summary<br />

The fabrication steps have been outlined and explained, and while there are<br />

always the possibility <strong>of</strong> changes or variations in parameters in each step, the<br />

process as described in this chapter, should be duplicable by others.


Chapter 6<br />

Problems and Solutions<br />

In almost every project some unexpected problems appears, ranging from<br />

merely inconvenient problems such as cleanroom machinery being out <strong>of</strong><br />

work, to more actual problems, such as fabrications steps being more complicated<br />

than they appear on paper. In this chapter some <strong>of</strong> the problems<br />

encountered during the process <strong>of</strong> creating the Clark sensor will be highlighted,<br />

as well as how they were solved when possible.<br />

6.1 Wirebonding<br />

As mentioned in Chapter 5, the first, and at the time the only, choice <strong>for</strong><br />

connecting the chip to the flex print was wirebonding. Wirebonding is basically<br />

done by having a needle, with a metal (Al) thread through it, pressing<br />

down on the contact pad <strong>of</strong> the chip, and then through the use <strong>of</strong> ultrasound<br />

and vibrations attach the thread. The similar process is then repeated on<br />

the flex print, at which time the rest <strong>of</strong> the thread is snapped <strong>of</strong>f. However<br />

this would soon prove to be a lot more complicated than it sounds. For some<br />

unexplainable reasons the thread refused to per<strong>for</strong>m the second attachment,<br />

whether it was on the gold on the chip, or on the flex print. Albeit some<br />

were wirebondings were successful, they were few and there was far between<br />

them.<br />

Why this is so, remained a mystery <strong>for</strong> a while, however there was some<br />

problems with the machine at the time. However a later examination <strong>of</strong> the<br />

flex print revealed that the metal on it was tin covered copper. There<strong>for</strong> it<br />

was impossible to bond to the flex print, unless the tin was scraped <strong>of</strong>f, as<br />

49


50 CHAPTER 6. PROBLEMS AND SOLUTIONS<br />

Figure 6.1: Picture <strong>of</strong> the wirebonder used to attach the wire to the chip and<br />

the flex print.<br />

wiring bonding is not possible to tin. Also people in the industry have suggested<br />

that the Au layer on the chip might have been to thin, and should be<br />

increased to 10k ˚A(it was 3000 ˚A), however others at MIC have successfully<br />

bonded to this thickness be<strong>for</strong>e. However this was only discovered late in the<br />

project, so some other way <strong>of</strong> connecting the chip and the flex print had to<br />

be found at the time.<br />

6.1.1 Conducting Glue<br />

As is <strong>of</strong>ten the case the simplest solution is also the most viable, in this case<br />

a conducting glue. 1 While it may look and sound less refined than a a metal<br />

wire, it is relatively easy to apply with the tip <strong>of</strong> a needle dipped in the glue.<br />

That is if as long as some distance exists between the output electrodes.<br />

The glue used is, like most glues, somewhat viscous and hence if the output<br />

electrodes are to close to each other, it can be very hard if not impossible to<br />

1 The glue is a conductive paste (H9807) from Namics, viscosity: 20 P a ·<br />

s, V olumeresistivity : 0.7 × 10 −4 Ω · cm


6.1. WIREBONDING 51<br />

make a line between the chip and the flex print without a short circuit being<br />

created. So while the conducting glue is a possible solution, it should only<br />

be considered a temporary solution.<br />

Figure 6.2: One <strong>of</strong> the chips with the silver colored electric glue on, as can<br />

be seen by comparing the left and the right side, there is a decent<br />

amount <strong>of</strong> distance between the output electrodes (left) <strong>for</strong> the<br />

Clark <strong>Sensor</strong>. While the output electrodes (right) <strong>of</strong> the temperature<br />

sensor is closer and harder to connect to with the conducting<br />

glue.<br />

Figure 6.3: Closeup <strong>of</strong> the design, the arrows point towards the platinum line<br />

that will cause the short circuit.<br />

While the conducting glue was a very viable solution one minor problem<br />

had to be overcome first. In order to make the wafer easier to dice, the chip<br />

design (as seen in Chapter 4 had a platinum line near the edge, see picture<br />

6.3, which if the conducting glue was just put on as it were, would work as a<br />

short circuit, hence a small procedure was developed to apply the glue. This<br />

procedure was as follows:<br />

1. Put a piece <strong>of</strong> ordinary tape over the sensor, to prevent it from being<br />

covered with glue, leaving roughly half <strong>of</strong> the output electrodes free.


52 CHAPTER 6. PROBLEMS AND SOLUTIONS<br />

2. Put a thick layer <strong>of</strong> non-conducting glue over the exposed half <strong>of</strong> the<br />

output electrodes and the platinum wire, and remove the tape. Then<br />

leave it to dry <strong>for</strong> at least a few hours. Then check under a microscope<br />

if more needs to be applied.<br />

Figure 6.4: Non conducting glue have been put on the chip, and the tape<br />

have been removed.<br />

3. Connect the output electrodes and the flex print with the conducting<br />

glue, using the tip <strong>of</strong> a needle dipped in it. Check <strong>for</strong> short circuits<br />

under a microscope, and then leave it overnight to dry.<br />

Figure 6.5: Both the conducting and the non conducting glue have been<br />

added, and the non conducting is now separating the conducting<br />

glue from the platinum line.


6.2. THE FLEX PRINT 53<br />

The reason behind letting the conducting glue dry slowly in step 3, rather<br />

than putting it in the oven right away, was that tests showed a lack <strong>of</strong> connection<br />

if the glue hadn’t dried properly first.<br />

Also it should be noted that while the solution with the conducting glue<br />

does work, it is not the most optimal solution, as the yield on working sensors<br />

with this solution is discouragingly low.<br />

6.2 The flex print<br />

There was also another problem with the wirebonds, or rather the flex print,<br />

even if the wire had been successfully attached to both the output electrodes<br />

and the flex print. After taking the packaged device out <strong>of</strong> the oven, the flex<br />

print appeared to bend upwards, despite the glue. This had the side effect<br />

<strong>of</strong> making the thread snap <strong>of</strong>f at one <strong>of</strong> the ends. Hence cutting <strong>of</strong>f the connection<br />

between the chip and the flex print. The reason <strong>for</strong> this appears to<br />

be that the flex print used, isn’t suitable <strong>for</strong> use with temperatures over 80 C.<br />

Figure 6.6: Picture from a chip, where conducting glue have been used to<br />

<strong>for</strong>m the connection between chip and flex print, as can be seen<br />

the flex print have bend and caused the connection to break.<br />

So in order to solve this, a way to keep the wire attached or to prevent<br />

the flex print from bending had to be found.


54 CHAPTER 6. PROBLEMS AND SOLUTIONS<br />

6.2.1 Non Conducting Glue<br />

Once again the solution involves glue, albeit a non-conducting one this time,<br />

the idea <strong>for</strong> this came to mind when a fellow master student 2 had to encapsulate<br />

a chip fast, and hence used glue instead to protect the thin metal wire<br />

from being torn <strong>of</strong>f. This made me thinking what if a similar glue was put in<br />

the hole in the upper PMMA piece? Once it was dry it should ensure that<br />

the flex print is unable to bend, as there is no room <strong>for</strong> it to bend to. Also<br />

as an added bonus it would help protect the more sensitive electronics from<br />

any seawater that might seep past the ’O’-ring.<br />

Figure 6.7: Glue covering both the flex print, wirebonds, and chip, thereby<br />

making the flex print and wirebond durable.<br />

6.3 The ’O’-ring<br />

Another problem with the design was the ’O’-ring, or rather making the<br />

’O’-ring stay where it was supposed to in the PMMA. The hole which the<br />

’O’-ring rest in had to have the right depth, since if it was to deep, the ’O’ring<br />

wouldn’t fully protect the rest <strong>of</strong> the chip from the seawater. While<br />

making it to shallow would cause the ’O’-ring to be squeezed out <strong>of</strong> the hole,<br />

or outside the area it is supposed to cover, as seen on picture 6.8.<br />

2 Sune Duun


6.3. THE ’O’-RING 55<br />

Figure 6.8: The ’O’-ring is being squeezed out <strong>of</strong> where it is supposed to be.<br />

The picture to the left shows it sticking out <strong>of</strong> the hole, while the<br />

picture to the right shows it being squeezed over the chip.<br />

6.3.1 Changing the resting area<br />

Now the original design <strong>for</strong> the packaging had the ’O’-ring resting on a flat<br />

area as show on picture 6.9. Also while the ’O’- was supposed to be squeezed<br />

together it was only supposed to contract about 20 % percent, in order to<br />

have a tight fit. So how was this solved? First <strong>of</strong>f the resting area <strong>for</strong> the<br />

’O’-ring was changed, so that it became a grove instead <strong>of</strong> a flat, as show on<br />

picture 6.9.<br />

Figure 6.9: The sketch to the left shows the original flat area, which the ’O’ring<br />

rested upon, while the sketch to the right shows the changed<br />

design, where a grove is added <strong>for</strong> the ’O’-ring.<br />

This, along with a optimization <strong>of</strong> the process where the laser cuts the<br />

grove, ensured that the the ’O’-ring no longer did this in most cases. Also it<br />

was better to have the screw clamp give a smaller amount <strong>of</strong> pressure, and<br />

then have the packing inside the oven a little longer if the PMMA layers<br />

hadn’t bonded properly, rather than applying much pressure and cause the<br />

’O’-ring to slip out. Since everything in the design should be able to survive<br />

the extended duration inside the oven.


56 CHAPTER 6. PROBLEMS AND SOLUTIONS<br />

6.4 Summary<br />

While many problems can occur in creating a sensor, it is not safe to assume<br />

that merely because the cleanroom processing is done that no further problems<br />

will be encountered. The problems shown and solved in this chapter,<br />

illustrates this very well.


Chapter 7<br />

Evaluation <strong>of</strong> Results and<br />

Measurements<br />

Shown in this chapter is the results gained during this project, as well as how<br />

they might be improved upon. While the process itself was successful, the<br />

problem that showed up with the packaging and the wirebonder, as detailed<br />

in Chapter refsec:problems, there is not as many measurements on the Clark<br />

<strong>Sensor</strong> as hoped, and there<strong>for</strong>e these will have to be obtained after this thesis<br />

is handed in.<br />

Figure 7.1: Pictures showing the missing Pt. On the picture on the left,<br />

some residues <strong>of</strong> platinum can be seen, where as the middle one<br />

shows a hole in the temperature sensor, while on the picture to<br />

the right the temperature sensor is missing entirely.<br />

7.1 Fabrication results<br />

The fabrication <strong>of</strong> the chips was successful, and one wafer made it through<br />

the whole clean room process, while a few others were postponed at an early<br />

57


58 CHAPTER 7. EVALUATION OF RESULTS AND MEASUREMENTS<br />

process step, in case problems with the rest <strong>of</strong> the process should arise.<br />

There was one small problem with the Pt, the exposure time in the photolithography<br />

was originally set to 7 seconds instead <strong>of</strong> 4. As a result this,<br />

both the working array and the temperature sensor had huge gaps, or were<br />

missing entire as seen on Figure 7.1.<br />

However once the exposure time was lowered to 4, the problem was solved,<br />

and the fine structure and small dots <strong>of</strong> Pt was clearly evident, as seen on<br />

Figure 7.2.<br />

Figure 7.2: Pt are no longer missing and both the dots in the working and<br />

the lines in the temperature sensor looks fine.<br />

Beyond this however there was no real problem in fabricating the chip,<br />

except <strong>for</strong> the problems already mentioned, also while the recipe <strong>for</strong> the<br />

titanium silicide was new, it <strong>for</strong>med successfully.<br />

7.2 Temperature <strong>Sensor</strong><br />

The temperature sensor was tested to see if it had the required linear dependency,<br />

and to see if the TCR (Temperature Coefficient <strong>of</strong> Resistance) varied<br />

from the standard and if so, how much it did. This is done in order to know<br />

how much that needs to be compensated, since there is a titanium strip beneath<br />

the platinum, and the gold contacts can also be expected to change it<br />

a bit.<br />

The experiment was done using 2 multimeter’s, a probestation, and a<br />

heating plate. The chip measured on was placed in the probestation, and<br />

the 2 multimeter’s was connected, one measuring the temperature <strong>of</strong> the chip,<br />

and the other measuring the resistance through the temperature sensor. The<br />

heating plate was then used to raise the temperature to around 50 o C, after<br />

which it was cooled down. Measurements was made both while heating and<br />

cooling.


7.2. TEMPERATURE SENSOR 59<br />

Figure 7.3: Graphs showing the relative temperature against the resistance,<br />

TCR is equal to the slope <strong>of</strong> the lines.<br />

As can be seen from Figure 7.3 the measured TCR’s (3.33×10 −3 , 4.09×10 −3 ,<br />

4.09×10 −3 , 4.11×10 −3 )are very close to the standard[46] 3.8×10 −3 <strong>for</strong> platinum.


60 CHAPTER 7. EVALUATION OF RESULTS AND MEASUREMENTS<br />

7.3 Clark <strong>Sensor</strong><br />

While there as mentioned was problems with the wire bonding and packaging,<br />

there was one sensor that was both bonded and packaged, and there<strong>for</strong>e<br />

could be measured upon. The measurements was done using a Gamry FAS2<br />

Femtostat, and a beaker <strong>of</strong> water with the sensor in it, finally grounded metal<br />

plates was put up around the beaker, in order to create the equivalent <strong>of</strong> a<br />

Faraday cage. This was done in order to minimize any outside noise (See<br />

Figure 7.4 <strong>for</strong> a sketch <strong>of</strong> the setup.<br />

PC<br />

Ground WE CE RE<br />

Potentiostat<br />

Constructed Faraday cage<br />

Figure 7.4: Sketch <strong>of</strong> the setup used to measure on the Clark sensor, 3 wires<br />

connects the working, reference and counter electrodes respectively<br />

to the potentiostat. The potentiostat collects the data and<br />

sends it to a nearby pc, which translates the signal into a graph.<br />

Un<strong>for</strong>tunately due to the faulty wirebonding one <strong>of</strong> the wires broke after<br />

a few test measurements, so no detailed measurements with varying amounts<br />

<strong>of</strong> dissolved oxygen could be made.<br />

However, as can be seen on Figure 7.5, there is the plateau shaped curve,<br />

which indicates that the sensor itself works. Of course since the a wire<br />

broke during the test measurements, no conclusion can be made in regards<br />

to whether it is measuring the right amount <strong>of</strong> oxygen or not. Hence further<br />

tests have to be made to confirm this. That the right shape <strong>of</strong> curve was<br />

obtained, is taken as a good sign, especially as there seems to be no noticeable<br />

internal noise generated.<br />

Beaker<br />

with<br />

Clark<br />

sensor<br />

and<br />

water


7.4. SUMMARY 61<br />

Figure 7.5: Polarogram from the test measurement with the designed Clark<br />

<strong>Sensor</strong> (Design 1CLX).<br />

7.4 Summary<br />

While there was a lack <strong>of</strong> measurements on the Clark sensor itself, the results<br />

from the measurements that have been obtained, as well as pictures <strong>of</strong> the<br />

chips, indicates that once the packaging problems have been solved, the Clark<br />

sensor should in all likelihood work as well.


62 CHAPTER 7. EVALUATION OF RESULTS AND MEASUREMENTS


Chapter 8<br />

Conclusion<br />

The primary goal <strong>for</strong> this project was to create a first generation oxygen<br />

sensor to compliment the Fish ’n Chip sensor. While a sensor have been<br />

created, it have not been tested, due to problems encountered during the<br />

projects, and as such the primary goal have not been fully fulfilled. However<br />

important steps towards a functioning dissolved oxygen sensor have been<br />

taken.<br />

Possible ways <strong>of</strong> measuring dissolved oxygen have been investigated, and<br />

a Clark type chosen.<br />

The theory behind the Clark sensor have been detailed, a further study<br />

into membrane materials and electrolytes could allow <strong>for</strong> improvements, as<br />

some <strong>of</strong> the potential optimization lies hidden here. Albeit this fall outside<br />

what knowledge <strong>of</strong> chemistry that I have.<br />

Various designs and the considerations behind them have been described,<br />

and with future results these should provide the basis <strong>for</strong> improving the<br />

sensor. This should also give a basic idea <strong>of</strong> what design outlay that is the<br />

most advantageous.<br />

The packaging concept, which proved to be the main problem despite not<br />

being within the focus <strong>of</strong> this project, have several flaws and will have to be<br />

given some thorough investigation.<br />

The temperature sensor seems to be working within the expected norms,<br />

and while not being revolutionary or even new, it is needed <strong>for</strong> the Clark<br />

sensor, due to the influence <strong>of</strong> temperature. An interesting aspect yet to<br />

explore however, is whether the Clark sensor influence it or vice versa.<br />

63


64 CHAPTER 8. CONCLUSION<br />

In summary this project still have a lot <strong>of</strong> potential, once the results from<br />

the Clark sensors have been obtained. Though from the gathered results and<br />

the design theory behind it, I feel confident that it does work, and plans have<br />

been made to measure on the chips to verify this and gather date, once the<br />

packaging problem is solved, even if it un<strong>for</strong>tunately falls outside the time<br />

frame <strong>of</strong> this project.


Bibliography<br />

[1] http://earthobservatory.nasa.gov/study/weighingwater/.<br />

[2] http://www.fisheriesmanagement.co.uk/fish%20studies/bloodflow countercurrent.htm.<br />

[3] Fortunat Joos, Gian-Kasper Plattner, Thomas F. Stocker, Arne Kortzinger,<br />

and Douglas W.R. Wallace. Trends in marine dissolved oxygen:<br />

Implications <strong>for</strong> ocean circulation changes and the carbon budget. EOS,<br />

84:197–204, May 2003.<br />

[4] E. Hunter, J.N. Aldridge, J.D. Metcalfe, and G.P. Arnold. Geolocation<br />

<strong>of</strong> free-ranging fish on the european continental shelf as determined from<br />

enviromental variables. <strong>Marine</strong> Biology, 142:601–609, 2003.<br />

[5] http://www.sk.lung.ca/content.cfm.<br />

[6] Chen Yu-Quan and Li Guang. An auto-calbrated miniature microhole<br />

cathode array sensor system <strong>for</strong> measuring dissolved oxygen. <strong>Sensor</strong>s<br />

and Actuators B, 10:219–222, 1993.<br />

[7] http://scienceworld.wolfram.com/chemistry/fugacity.html.<br />

[8] http://www.state.ky.us/nrepc/water/wcpdo.htm.<br />

[9] http://waterontheweb.org/under/waterquality/oxygen.html.<br />

[10] Td 218 operating manual oxygen optode 3830 and 3930.<br />

[11] S. Gatti, T. Brey, W.E.G. Muller, O. Heilmayer, and G. Holst. <strong>Oxygen</strong><br />

microoptodes: a new tool <strong>for</strong> oxygen measurement in aquatic animal<br />

ecology. <strong>Marine</strong> Biology, 140:1075–1085, 2002.<br />

[12] I. Klimant, M. Kuhl, R.N. Glud, and G. Holst. Optical measurements<br />

<strong>of</strong> oxygen and temperature in microscale: strategies and biological applications.<br />

<strong>Sensor</strong>s and Actuators B, 38-39:29–37, 1997.<br />

65


66 BIBLIOGRAPHY<br />

[13] S. McCulloch and D. Uttamchandani. Fibre optic micro-optrode <strong>for</strong><br />

dissolved oxygen measurements. IEE Proc.-Sci Meas. Technol., 146:123–<br />

127, May 1993.<br />

[14] S.McCulloch and D.Uttamchandi. Fibre optic micro-optrode <strong>for</strong> dissolved<br />

oxygen measurements. IEEE Proc.-Sci. Meas. Tecnol, 146:123–<br />

127, May 1999.<br />

[15] P. Bergveld. Isfet, theory and practise. IEEE <strong>Sensor</strong> Conference Toronto,<br />

October 2003.<br />

[16] R. Ramamoorthy, P.K. Dutta, and S.A. Akbar. <strong>Oxygen</strong> sensors: Materials,<br />

methods, designs and applications. Journal <strong>of</strong> Materials Science,<br />

38:4271–4282, 2003.<br />

[17] Andreas Hierleman and Henry Baltes. Cmos-based chemical microsensors.<br />

The Analyst, pages 15–28, November 2002.<br />

[18] P. Bergveld. Isfet, theory and practice. IEEE <strong>Sensor</strong> Conference<br />

Toronto, October 2003.<br />

[19] Byung-Ki Sohn and Chang-Soo Kim. A new ph-isfet based dissolved oxygen<br />

sensor by employing electrolysis <strong>of</strong> oxygen. <strong>Sensor</strong>s and Actuators<br />

B, 34:435–440, 1996.<br />

[20] W. Oelssner, J. Zosel, U. Guth, T. Pechstein, W. Babel, J.G. Connery,<br />

C. Demuth, M. Grote Gansey, and J.B. Verburg. Encapsulation <strong>of</strong> isfet<br />

sensor chips. <strong>Sensor</strong>s and Actuators B, pages 1–13, 2004.<br />

[21] L.C. Clark. Monitor and control <strong>of</strong> blood and tissue tensions. Trans.<br />

Am. Soc. Artif. Intern. Organs., 2:41–48, 1956.<br />

[22] Cynthia G. Zoski and Nafeesa Simjee. Addressable microelectrode arrays:<br />

Characterization by imaging with scanning electrochemical microscopy.<br />

Analytical Chemistry, 76:62–72, January 2004.<br />

[23] Fumio Hine. Electrode Processes and Electrochemical Engineering.<br />

Plenum Press, 1985.<br />

[24] http://www.eidusa.com/theory do.htm.<br />

[25] R. Mark Wightman. Voltammetry with microscopic electrodes in new<br />

domains. Science, 240:415–420, April 1988.


BIBLIOGRAPHY 67<br />

[26] W.E. Morf. The Principles <strong>of</strong> Ion-Selective Electrodes and <strong>of</strong> Membrane<br />

Transport. Elsevier, 1981.<br />

[27] T. R. Yu and G. L. Ji. Electrochemical Methods in Soil and Water<br />

Research. Pergamon Press, 1993.<br />

[28] A.F. Albantov and A.L. Levin. New functional possibilities <strong>for</strong> amperometric<br />

dissolved oxygen sensors. Biosensors & Bioelectronics, 9:515–526,<br />

1994.<br />

[29] http://www.corrosion-doctors.org/biographies/davybio.htm.<br />

[30] http://www.rigb.org/rimain/heritage/faradaypage.jsp.<br />

[31] D.L. Short and G.S.G. Shell. Fundamentals <strong>of</strong> clark membrane configuration<br />

oxygen sensors: some confussion clarified. J. Phys. E. Sci.<br />

Instrum., 17:1085–1092, 1984.<br />

[32] Alexander Frey, Martin Jenker, Meinrad Schienle, Christian Paulus, Birgit<br />

Holzapfl, Petra Schindler-Bauer, Franz H<strong>of</strong>mann, Dirk Kuhlmeier,<br />

Jurgen Krause, Jorg Albers, Walter Gumbrecht, Doris Schmitt-<br />

Lansiedel, and Roland Thewes. Design <strong>of</strong> an integrated potentiostat<br />

circuit <strong>for</strong> cmos bio sensor chips. IEEE, 2003.<br />

[33] Slawomir Kalinowski and Zbigniew Figaszewski. A four-electrode<br />

potentiostat-galvanostat <strong>for</strong> studies <strong>of</strong> bilayer lipid membranes. Meas.<br />

Sci. Technol, 6:1050–1055, 1995.<br />

[34] Anders Hyldg˚ard. Developement <strong>of</strong> a multi-sensor <strong>for</strong> marine environment<br />

studies. Master’s thesis, MIC, April 2004.<br />

[35] U. Guth, W. Oelssner, and W. Vonau. Investigation <strong>of</strong> corrosion phenomena<br />

on chemical microsensors. Electrochimica Acta, 47:201–210,<br />

2001.<br />

[36] Marek Dawgul, Dorota G. Pijanowska, Alfred Krzyskow, Jerzy Kruk,<br />

and Wladyslaw Torbicz. An influence <strong>of</strong> polyHEMA gate layer on properties<br />

<strong>of</strong> chemFETs. <strong>Sensor</strong>s, 3:146–159, 2003.<br />

[37] M. Wittkampf, K. Cammann, M. Amrein, and R. Reichelt. Characterization<br />

<strong>of</strong> microelectrode arrays by means <strong>of</strong> electrochemical and surface<br />

analysis methods. <strong>Sensor</strong>s and Actuators B, 40:79–84, 1997.<br />

[38] Yuzuru Iwasaki and Masao Morita. Electrochemical measurements with<br />

interdigitated array microelectrodes. Current Seperations, 14, 1995.


68 BIBLIOGRAPHY<br />

[39] Mairi E. Sandison, Natalie Anicet, Andrew Glidle, and Jonathan M.<br />

Cooper. Optimization <strong>of</strong> the geometry and porosity <strong>of</strong> microelectrode<br />

arrays <strong>for</strong> sensor design. Analytical Chemistry, 74:5717–5725, November<br />

2002.<br />

[40] Hiroaki Suzuki. Advances in the micr<strong>of</strong>abrication <strong>of</strong> electrochemical<br />

sensors and systems. Electroanalysis, 12:703–715, 2000.<br />

[41] B. Ross, K. Cammann, W. Mokwa, and M. Rospert. Ultramicroelectrode<br />

arrays as tranducers <strong>for</strong> new amperometric oxygen sensors. <strong>Sensor</strong>s<br />

and Actuators B, 7:758–762, 1992.<br />

[42] Glen W. McLaughlin, Katie Braden, Benjamin Franc, and Gregory T.A.<br />

Kovacs. Micr<strong>of</strong>abricated solid-state dissolved oxygen sensor. <strong>Sensor</strong>s and<br />

Actuators B, 83:138–148, 2002.<br />

[43] Chen Yu-Quan and Li Guang. A mathematical model with finiteelement<br />

analysis <strong>of</strong> recessed dissolved-oxygen cathode array. <strong>Sensor</strong>s<br />

and Actuators B, 10:223–228, 1993.<br />

[44] http://www.sablesys.com/oxrechlo.htm.<br />

[45] Sotiris Sotiropoulos and Kirsi Wallgren. Solid-state microelectrode oxygen<br />

sensors. Analytica Chimica Acta, 388:51–62, 1999.<br />

[46] http://www.microwaves101.com/encyclopedia/temperature.cfm.


Appendix A<br />

Silicide Recipe<br />

The following two recipes are <strong>for</strong> the Jipelec RTP machine at MIC.<br />

Step Initial Temperature Temperature Duration Thermocoupler Open Vent N2<br />

or Power or Power Control Valve (sccm)<br />

1 20 0 30 Y Y 0<br />

2 0 0 30 Y Y 200<br />

3 0 0 30 Y Y 0<br />

4 0 0 100 Y Y 200<br />

5 0 0 100 Y Y 0<br />

6 0 Power 20 30 N Y 0<br />

7 Power 20 Power 20 2400 N Y 0<br />

8 Power 20 0 10 N Y 0<br />

Table A.1: Prebake Settings to remove any oxygen from the ceramic wafer<br />

holder and prepare it (Recipe name Bake Clark).<br />

69


70 APPENDIX A. SILICIDE RECIPE<br />

Step Initial Temperature Temperature Duration TC Open Vent N2<br />

or Power or Power Control Valve (sccm)<br />

1 20 20 40 Y N 0<br />

2 20 20 30 N Y 400<br />

3 20 20 40 Y N 0<br />

4 20 20 30 N Y 400<br />

5 20 20 40 Y N 0<br />

6 20 200 20 Y N 0<br />

7 200 400 20 Y N 0<br />

8 400 600 20 Y N 0<br />

9 600 800 20 Y N 0<br />

10 800 850 10 Y N 0<br />

11 850 850 30 Y N 0<br />

12 850 20 600 Y N 0<br />

13 20 20 120 N Y 400<br />

Table A.2: Silicide Settings to make the actual Silicide(Recipe name PKH<br />

1).


Appendix B<br />

Fabrication Process<br />

71


72 APPENDIX B. FABRICATION PROCESS


74 APPENDIX B. FABRICATION PROCESS


Appendix C<br />

Process Sequence<br />

75


76 APPENDIX C. PROCESS SEQUENCE


78 APPENDIX C. PROCESS SEQUENCE

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!