Quadra-Chek ® <strong>5000</strong><br />

VED Calibrations<br />

The following calibrations are required before using VED with the QC-<strong>5000</strong>:<br />

• Auto focus teach (each magnification)<br />

• Edge calibration (each VED probe tool)<br />

• Pixel calibration (each magnification)<br />

• Camera calibration (each camera)<br />

• Parcentricity calibration (each magnification)<br />

Use the glass master standard shipped with the QC-<strong>5000</strong> to perform these calibrations.<br />

NOTE<br />

Perform all VED calibrations from the highest magnification to<br />

the lowest magnification. For example, if there are three<br />

magnifications (low, medium, and high) perform calibrations<br />

on high first.<br />

NOTE<br />

Ignore calibration warnings as shown during the calibration<br />

process. Click the hide warning button and continue the<br />

calibration. The QC-<strong>5000</strong> displays this message to warn users to<br />

complete calibrations before performing inspections.<br />

To perform an auto focus teach<br />

Step 1<br />

Highlight the desired magnification as shown.<br />

3-7