Quadra-Chek ® <strong>5000</strong><br />

Step 3<br />

True magnification must be enabled. If the True magnification box is not checked,<br />

click on True magnification and then click Show.<br />

Step 4<br />

Click OK to close the Customize screen, and then return to the Tools/Options/<br />

VED tab.<br />

Step 5<br />

Bring a standard artifact into sharp focus in the video Window’s field of view.<br />

The Metronics calibration slide is an excellent standard to use for this procedure.<br />

Step 6<br />

Measure the distance between 2 of the<br />

standard’s marks on the computer<br />

screen using a scale or ruler. Measure<br />

as great a distance as possible.<br />

Step 7<br />

Calculate the measured magnification<br />

by dividing the measured distance<br />

(Step 6) by the actual distance shown<br />

on the standard.<br />

Measured magnification = Measured / Actual<br />

Step 8<br />

Calculate the pixel size factor by dividing the measured magnification by the<br />

magnification value currently displayed in the right corner of the status bar.<br />

Pixel size factor = Measured Magnification / Status bar value<br />

Step 9<br />

Enter the pixel size factor into the X pixel<br />

size... field.<br />

NOTE<br />

The value shown here is only an example.<br />