10.2 Load batteries<br />

30<br />

NOTE:<br />

Also note the loading instructions for batteries<br />

• in the user manual of “<strong>G40</strong>”,<br />

• in the user manual of charging unit,<br />

• on front and rear of charging unit.<br />

For care and maintenance of mechanical components<br />

of batteries see chapter 10.1.<br />

CAUTION:<br />

• Exercise care when checking / loading. Avoid<br />

short circuit at battery poles.<br />

> Danger of short circuit <<br />

Check Action<br />

• Status display at remote flashes 7<br />

times (deep discharging)?<br />

• Status display at remote flashes 8<br />

times (excess voltage )?<br />

• Measure voltage of batteries:<br />

- Voltage under 11,5V?<br />

- Voltage under 8V?<br />

(Battery deep discharged.)<br />

Different voltage<br />

(Tolerance ±0,2 V) at both batteries?<br />

Loading process<br />

CAUTION:<br />

> Measure battery voltage.<br />

> Charging unit defective.<br />

See chapter 10.3.<br />

INVACARE-Charging unit<br />

Mains cable<br />

> Load <strong>with</strong> original charging<br />

unit<br />

> Pre-load <strong>with</strong> unregulated<br />

charging unit up to 9V.<br />

(Description see below.)<br />

> Replace batteries.<br />

Charging plug<br />

NOTE:<br />

• Processing during loading of batteries <strong>with</strong> INVACARE-charging unit:<br />

Also see user manual of <strong>G40</strong>.<br />

• Pay attention to connecting order of INVACARE-charging unit:<br />

First connect charging plug, then mains plug.<br />

Charging socket<br />

• Deep discharged batteries can not be loaded <strong>with</strong> INVACARE-charging unit.<br />

• If batteries are not loaded completely they must be replaced.<br />

Charging<br />

cable<br />

Remote<br />

• During loading batteries must not lie on side or stand edgewise. The batterypoles must show<br />

upwards.<br />

• At <strong>G40</strong> <strong>with</strong> <strong>ACS</strong> use only gelbatteries (loading power maximal 7 A).<br />

Wetbatteries (loading power max. 10 A) overload cables.<br />

• Batteries and charging unit must be set out ambient temperature because the loading is also<br />

controlled <strong>with</strong> temperature sensors. Elongated cables and different locations of charging unit<br />

and batteries cause faulty loading.<br />

• During full loading the display “full loading” on charging unit must light.<br />

When a charging voltage of 28 V is reached, also the LED´s at remote flash.<br />

The voltage of a single battery shall amount during full loading and in room temperature<br />

(ca. 20 - 22°C) 14 ±0,2 V.