8 What to do in the event of a fault<br />

description of fault possible cause cure<br />

no output signal<br />

• power supply not connected<br />

• connect the correct power<br />

supply<br />

output signal remains<br />

constant in spite of<br />

changing pressure<br />

faulty output signal<br />

zero point can no longer<br />

be adjusted <strong>with</strong> U0 / I0<br />

• incorrect power supply<br />

connected<br />

• fuse blown<br />

• defective input protection diode<br />

• defective output protection diodes<br />

• pressure connections crossed<br />

over<br />

• creeping suppression volume set<br />

too high<br />

• defective output protection diode<br />

• defective pressure measurement<br />

canister<br />

• <strong>with</strong> current output load to large<br />

• <strong>with</strong> voltage output load<br />

resistance to small<br />

• defective pressure measurement<br />

canister<br />

• connect the correct power supply<br />

(see rating label)<br />

• replace fuse SI (type TR5 200 mAT<br />

from the company Wickmann)<br />

• replace diode D12( type ZPY 33)<br />

• replace D8 / D9 (type ZPY18)<br />

• connect pressure in accordance<br />

<strong>with</strong> “3. Installation“<br />

• turn the trimmer SMU anti-clock<br />

wise until the output<br />

signal ≠ 0 V / 0/4 mA<br />

• replace D8 / D9 (type ZPY18)<br />

• return instrument for repair at<br />

factory<br />

• observe maximum load of 500 Ω<br />

• minimum load resistance of 5 kΩ<br />

• return instrument for repair at<br />

factory<br />

Table 2<br />

Never perform a functional test using exhalation as this can lead to<br />

permanent damage of the pressure measuring cell.<br />

Simple functional test of the trans<strong>ducer</strong>:<br />

Connect a hose to the pressure port for positive gauge pressure + . Taking the hose<br />

between thumb and fore finger, squeeze it and then carefully move the volume of air<br />

trapped <strong>with</strong>in the hose in the direction of the trans<strong>ducer</strong>.<br />

halstrup-walcher GmbH, 79199 Kirchzarten Operating instructions P <strong>82</strong> R Page 6 of 9