<strong>CTS</strong>-<strong>00135</strong> Technical Data Sheet<br />
Shimadzu® SPD10A, 10AVP, 10AVVP Deuterium<br />
Lamp<br />
Similar to Shimadzu® 228-34016-02<br />
4.0 Warranty<br />
<strong>CTS</strong> guarantees the lifetime of this product for 2000 hours or<br />
one year from purchase date whichever comes first. The guaranteed<br />
lifetime is defined as the time in which the radiant<br />
power of the lamp has dropped to 50% of its initial value measured<br />
at 230nm and 300mA operating current. <strong>CTS</strong> also warrants<br />
the product to be free of defects in materials and workmanship<br />
and will replace, without cost, the product which carries<br />
such defects. Please call 1-800-682-6480 for a Return<br />
Authorization Number before sending your return to <strong>CTS</strong>.<br />
5.0 Lamp Replacement<br />
5.1 General Criteria for Lamp Replacement<br />
a. Lamp will not ignite.<br />
b. Noise or drift exceeds specifications.<br />
c. Dictated by Preventive Maintenance Schedule per<br />
SOP.<br />
d. Lamp energy is below 50% of original intensity as<br />
recorded during initial installation<br />
5.2 Installation Instructions<br />
a. Power down the PDA detector and remove the power<br />
cord.<br />
b. Wait 15 minutes for the detector lamp to cool before<br />
handling the lamp.<br />
c. Access the lamp inside detector housing.<br />
d. Disconnect the lamp power connector.<br />
e. Unscrew the two mounting screws.<br />
f. Grip the metal base and pull the lamp out.<br />
g. Remove the mounting screws from the old lamp.<br />
h. Carefully unpack and hold the replacement lamp by<br />
its base only.<br />
i. Orient the lamp so the two holes in the base align with<br />
the mounting pins and screws in the optics bench.<br />
j. Insert the lamp and secure it with the two mounting<br />
screws.<br />
k. Reconnect the lamp power connector.<br />
j. Close detector housing.<br />
l. Reconnect the power cord and power up PDA<br />
detector.<br />
m. Record initial lamp intensity in the instrument log.