<strong>CTS</strong>-<strong>00135</strong> Technical Data Sheet<br />

Shimadzu® SPD10A, 10AVP, 10AVVP Deuterium<br />

Lamp<br />

Similar to Shimadzu® 228-34016-02<br />

4.0 Warranty<br />

<strong>CTS</strong> guarantees the lifetime of this product for 2000 hours or<br />

one year from purchase date whichever comes first. The guaranteed<br />

lifetime is defined as the time in which the radiant<br />

power of the lamp has dropped to 50% of its initial value measured<br />

at 230nm and 300mA operating current. <strong>CTS</strong> also warrants<br />

the product to be free of defects in materials and workmanship<br />

and will replace, without cost, the product which carries<br />

such defects. Please call 1-800-682-6480 for a Return<br />

Authorization Number before sending your return to <strong>CTS</strong>.<br />

5.0 Lamp Replacement<br />

5.1 General Criteria for Lamp Replacement<br />

a. Lamp will not ignite.<br />

b. Noise or drift exceeds specifications.<br />

c. Dictated by Preventive Maintenance Schedule per<br />

SOP.<br />

d. Lamp energy is below 50% of original intensity as<br />

recorded during initial installation<br />

5.2 Installation Instructions<br />

a. Power down the PDA detector and remove the power<br />

cord.<br />

b. Wait 15 minutes for the detector lamp to cool before<br />

handling the lamp.<br />

c. Access the lamp inside detector housing.<br />

d. Disconnect the lamp power connector.<br />

e. Unscrew the two mounting screws.<br />

f. Grip the metal base and pull the lamp out.<br />

g. Remove the mounting screws from the old lamp.<br />

h. Carefully unpack and hold the replacement lamp by<br />

its base only.<br />

i. Orient the lamp so the two holes in the base align with<br />

the mounting pins and screws in the optics bench.<br />

j. Insert the lamp and secure it with the two mounting<br />

screws.<br />

k. Reconnect the lamp power connector.<br />

j. Close detector housing.<br />

l. Reconnect the power cord and power up PDA<br />

detector.<br />

m. Record initial lamp intensity in the instrument log.