17:06 132. Imaging of Metallic Implant Using 3D Ultrashort Echo Time (3D UTE) Pulse<br />

Sequence<br />

Jiang Du 1 , Kelly Borden 1 , Eric Diaz 1 , Mark Bydder 1 , Won Bae 1 , Shantanu Patil 2 , Graeme<br />

Bydder 1 , Christine Chung 1<br />

1 Radiology, University of California, San Diego, CA, United States; 2 Shiley Center for Orthopedic Research<br />

and Education, La Jolla, CA, United States<br />

Magnetic resonance imaging (MRI) near metal implants suffers from severe artifacts due to large metal-induced field<br />

inhomogeneities. The steep field gradients near metal implants result in increased intra-voxel dephasing and a much shortened T2*.<br />

Clinical gradient echo (GE) sequences suffer from large signal loss. Spin echo (SE) type sequences only partly refocus the dephased<br />

spins, resulting in spatially dependent signal voids and pile-ups. Here we present a 3D ultrashort TE (UTE) sequence which employs<br />

short hard pulse excitation and 3D radial sampling with a nominal TE of 8 µs to image metallic implants with markedly reduced<br />

artifact.<br />

17:18 133. kf ARC Reconstruction for Improving MRI Around Metal Using MAVRIC<br />

Peng Lai 1 , Weitian Chen 1 , Christina Chen 2 , Kevin M. Koch 3 , Anja CS. Brau 1<br />

1 Applied Science Laboratory, GE Healthcare, Menlo Park, CA, United States; 2 Stanford University, Stanford,<br />

CA, United States; 3 Applied Science Laboratory, GE Healthcare, Waukesha, WI, United States<br />

This work developed a new method, kf ARC, for highly accelerated MAVRIC imaging around metal implants. The proposed method<br />

utilizes both k-space correlation and spectral correlation between adjacent spectral images to improve reconstruction. kf ARC was<br />

evaluated on 2 patients with metallic implants in comparison with conventional parallel imaging. Our results show that kf ARC can<br />

significantly improve image quality at high acceleration factors and is a promising approach to fast MAVRIC data acquisition.<br />

17:30 134. Morphological and Quantitative Evaluation of Meniscal Calcifications by Novel 2D<br />

IR and 3D UTE MR Techniques<br />

Patrick Omoumi 1,2 , Eric S. Diaz 1 , Jiang Du 1 , Sheronda S. Statum 1 , Won C. Bae 1 , Graeme<br />

Bydder 1 , Christine B. Chung 1<br />

1 University of California, San Diego, San Diego, CA, United States; 2 Cliniques Universitaire St Luc, Brussels,<br />

Belgium<br />

Meniscal calcifications are frequent and likely alter the normal biomechanics of the meniscus. Although MR imaging is the noninvasive<br />

technique of choice for the evaluation of meniscal pathology, it does not allow the facile visualization of meniscal<br />

calcifications. This is due to a lack of contrast (both calcifications and menisci have relatively short T2 relaxation times), and a lack of<br />

saptial résolution with standard clinical sequences. We describe novel MR imaging techniques based on 2D-UTE inversion recovery<br />

and 3D-UTE data acquisition to address these factors. We assessed the ability of these sequences to allow the visualization,<br />

characterization and quantitative evaluation of meniscal calcifications.<br />

17:42 135. Fiber Tracking of Dipolar Directions in the Meniscus<br />

Nikolaus M. Szeverenyi 1 , Graeme M. Bydder 1<br />

1 Radiology, University of California, San Diego, San Diego, CA, United States<br />

This study examines a method to extract and use dipolar information to characterize an ex-vivo meniscus sample. A goat meniscus<br />

was embedded in a spherical epoxy ball and the MR signal intensity examined as a function of orientation to a 3T static field.<br />

Unaveraged dipolar interactions caused dramatic signal variations in sub-structures. After correcting for coil sensitivity and coregistering<br />

all images, a principle dipolar direction was extracted for each voxel. This directional data could be analyzed and viewed<br />

as a direction map, similar to DTI brain data. The intensity fluctuations provided a FA map. Fiber tracks were generated.<br />

17:54 136. Ultrashort Echo Imaging (UTE) of Rotator Cuff Repair in an Ovine Model<br />

Matthew F. Koff 1 , Hollis G. Potter 1<br />

1 Department of Radiology and Imaging, Hospital for Special Surgery, New York, United States<br />

The rotator cuff tendons typically display low signal on standard clinical images due to the highly ordered collagen within the tissue.<br />

Ultrashort echo (UTE) imaging creates contrast for visualization and for T2* quantitation. This study used T2* mapping to evaluate<br />

rotator cuff repair in an ovine model. Reparative surgery was performed to the supraspinatus tendon in sheep. Shoulders were scanned<br />

ex-vivo 8 weeks post-operatively. T2* values of repaired tendon were significantly longer than normal tendon. The T2* values<br />

decreased in magnitude along the length of the repair, but not significantly. This pilot study highlights the use of UTE for quantitative<br />

evaluation of soft tissue repair.<br />

18:06 137. Detection of Dipolar Splitting in Rodent Tendons as a Function Axial Position with<br />

Double-Quantum Filtered Spectroscopic Imaging<br />

Henry H. Ong 1 , Joseph J. Sarver 2 , Jason E. Hsu 2 , Louis J. Soslowsky 2 , Felix W. Wehrli 1<br />

1 Laboratory for Structural NMR Imaging, Department of Radiology, University of Pennsylvania School of<br />

Medicine, Philadelphia, PA, United States; 2 McKay Orthopaedic Research Laboratory, University of<br />

Pennsylvania, Philadelphia, PA, United States<br />

Tendons are comprised of parallel collagen fibers that connect muscles to bone. Collagen-associated water has anisotropic rotational<br />

motion, which gives rise to residual dipolar splitting in 1H NMR. Double-quantum filtered (DQF) NMR and MRI can be used to<br />

observe the splitting and study the biophysical and structural properties of tendon. Here, we modified a DQF 1D spectroscopic<br />

imaging sequence to obtain 1H DQF spectra along the axis of the flexor digitorum profundus (FDP) tendons from rat hind limbs and<br />

show spectral differences in the region that wraps under the calcaneus, which experiences compressive forces.