Apex Pin Fixation<br />
Technical Details –<br />
Self-Drilling/Self-Tapping Apex Pin<br />
Pin Design<br />
The self-drilling tip of the Apex Pin<br />
acts like a new, sharp drill bit every<br />
time, and therefore, pre-drilling is<br />
not necessary. Combined with unique<br />
cutting geometry, this one-step<br />
procedure allows the pin to maintain a<br />
reduced insertion temperature due to<br />
decreased friction.<br />
A double helical flute creates a<br />
homogeneous thread profile that<br />
transports bone chips out of the drill<br />
hole for improved pin/bone interface.<br />
The U-shaped thread maximizes contact<br />
with the bone and controls stress<br />
distribution on the pin/bone interface<br />
by optimizing radial tension.<br />
The highly advanced cutting geometry<br />
allows for more precise pin insertion<br />
with reduced insertion time and<br />
temperature for optimal performance.<br />
The cylindrical thread is designed for<br />
good bone purchase, and pull-out<br />
resistance 1, 2, 3 , and offers the possibility<br />
to backing out the pin without<br />
compromising fixation.<br />
1 <strong>Stryker</strong> Test Report BML 10-183; Pull-out test for<br />
several Apex pins, Labor job number V10219; 2010.<br />
2 <strong>Stryker</strong> Test Report RA 05-168; Clinical<br />
Assessment - Hoffmann II; Hoffmann II Compact<br />
and Apex Pins; 2010.<br />
3 Fachhochschule Hamburg; Vergleichende biomechanische<br />
Untersuchung von “Fixateur Externe-Pins”;<br />
1997.<br />
U-shaped thread<br />
Self-tapping action<br />
Self-drilling tip<br />
31